Biofilm reduction potential of 0.02% polyhexanide irrigation solution in several types of urethral catheters
BACKGROUND: Long-term use of urethral catheters is associated with high risk of urinary tract infection (UTI) and blockage. Microbial biofilms are a common cause of catheter blockage, reducing their lifetime and significantly increasing morbidity of UTIs. A 0.02% polyhexanide irrigation solution dev...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8034122/ https://www.ncbi.nlm.nih.gov/pubmed/33836738 http://dx.doi.org/10.1186/s12894-021-00826-3 |
Sumario: | BACKGROUND: Long-term use of urethral catheters is associated with high risk of urinary tract infection (UTI) and blockage. Microbial biofilms are a common cause of catheter blockage, reducing their lifetime and significantly increasing morbidity of UTIs. A 0.02% polyhexanide irrigation solution developed for routine mechanical rinsing shows potential for bacterial decolonization of urethral catheters and has the potential to reduce or prevent biofilm formation. METHODS: Using an in vitro assay with standard market-leading types of catheters artificially contaminated with clinically relevant bacteria, assays were carried out to evaluate the biofilm reduction and prevention potential of a 0.02% polyhexanide solution versus no intervention (standard approach) and irrigation with saline solution (NaCl 0.9%). The efficiency of decolonization was measured through microbial plate count and membrane filtration. RESULTS: Irrigation using a 0.02% polyhexanide solution is suitable for the decolonization of a variety of transurethral catheters. The effect observed is significant compared to irrigation with 0.9% saline solution (p = 0.002) or no treatment (p = 0.011). No significant difference was found between irrigation with 0.9% saline solution and no treatment (p = 0.74). CONCLUSIONS: A 0.02% polyhexanide solution is able to reduce bacterial biofilm from catheters artificially contaminated with clinically relevant bacteria in vitro. The data shows a reduction of the viability of thick bacterial biofilms in a variety of commercially available urinary catheters made from silicone, latex-free silicone, hydrogel-coated silicone and PVC. Further research is required to evaluate the long-term tolerability and efficacy of polyhexanide in clinical practice. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1186/s12894-021-00826-3. |
---|