Mechanotendography: description and evaluation of a novel method for investigating the physiological mechanical oscillations of tendons using a piezo-based measurement system

The mechanotendography (MTG) is a method for analyzing the mechanical oscillations of tendons during muscular actions. The aim of this investigation was to evaluate the technical reliability of a piezo-based measurement system used for MTG. The reliability measurements were performed by using audio...

Descripción completa

Detalles Bibliográficos
Autores principales: Schaefer, Laura V., Bittmann, Frank N.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: PAGEPress Publications, Pavia, Italy 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8056155/
https://www.ncbi.nlm.nih.gov/pubmed/33709652
http://dx.doi.org/10.4081/ejtm.2021.9553
Descripción
Sumario:The mechanotendography (MTG) is a method for analyzing the mechanical oscillations of tendons during muscular actions. The aim of this investigation was to evaluate the technical reliability of a piezo-based measurement system used for MTG. The reliability measurements were performed by using audio samples played by a subwoofer. The thereby generated pressure waves were recorded by a piezo-based measurement system. An audio of 40 Hz sine oscillations and four different formerly in vivo recorded MTG-signals were converted into audio files and were used as test signals. Five trials with each audio were performed and one audio was used for repetition trials on another day. The signals’ correlation was estimated by Spearman (MCC) and intraclass correlation coefficients (ICC(3,1)), Cronbach’s alpha (CA) and by mean distances (MD). All parameters were compared between repetition and randomized matched signals. The repetition trials show high correlations (MCC: 0.86 ± 0.13, ICC: 0.89 ± 0.12, CA: 0.98 ± 0.03), low MD (0.03 ± 0.03V) and differ significantly from the randomized matched signals (MCC: 0.15 ± 0.10, ICC: 0.17 ± 0.09, CA: 0.37 ± 0.16, MD: 0.19 ± 0.01V) (p = 0.001 – 0.043). This speaks for an excellent reliability of the measurement system. Presuming the skin above superficial tendons oscillates adequately, we estimate this tool as valid for the application in musculoskeletal system.