In-Depth Metallurgical and Microstructural Analysis of Oneshape and Heat Treated Onecurve Instruments

OBJECTIVE: To define surface, mechanical, microstructural and metallurgical features of conventional OneShape (OShape) and heat-treated OneCurve (OCurve) nickel-titanium instruments. METHODS: Instruments were analysed by scanning electron microscopy (SEM) on new instruments and after simulated clini...

Descripción completa

Detalles Bibliográficos
Autores principales: Azizi, Arash, Prati, Carlo, Schiavon, Riccardo, Fitzgibbon, Raquel, Pirani, Chiara, Iacono, Francesco, Pelliccioni, Gian Andrea, Spinelli, Andrea, Zamparini, Fausto, Puddu, Pietro, Bolelli, Giovanni, Generali, Luigi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Kare Publishing 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8056813/
https://www.ncbi.nlm.nih.gov/pubmed/33762534
http://dx.doi.org/10.14744/eej.2020.63634
_version_ 1783680726210707456
author Azizi, Arash
Prati, Carlo
Schiavon, Riccardo
Fitzgibbon, Raquel
Pirani, Chiara
Iacono, Francesco
Pelliccioni, Gian Andrea
Spinelli, Andrea
Zamparini, Fausto
Puddu, Pietro
Bolelli, Giovanni
Generali, Luigi
author_facet Azizi, Arash
Prati, Carlo
Schiavon, Riccardo
Fitzgibbon, Raquel
Pirani, Chiara
Iacono, Francesco
Pelliccioni, Gian Andrea
Spinelli, Andrea
Zamparini, Fausto
Puddu, Pietro
Bolelli, Giovanni
Generali, Luigi
author_sort Azizi, Arash
collection PubMed
description OBJECTIVE: To define surface, mechanical, microstructural and metallurgical features of conventional OneShape (OShape) and heat-treated OneCurve (OCurve) nickel-titanium instruments. METHODS: Instruments were analysed by scanning electron microscopy (SEM) on new instruments and after simulated clinical use (SCU). Cyclic fatigue testing was performed and the number of cycles to fracture (NCF) and the length of the fractured instruments were measured (Mann-Whitney test). Fractured instruments during cyclic fatigue testing were then inspected by SEM fractographic analysis. Field emission gun scanning scanning electron microscopy (FEG-SEM), energy-dispersive X-ray spectroscopy (EDX) and micro-Raman spectroscopy were used to assess alloy surface chemistry. Focused ion beam (FIB) was performed to analyse the oxide layer on the surface of OCurve before and after SCU. X-Ray diffraction (XRD), metallographic evaluation and differential scanning calorimetry (DSC) were used to determine martensitic/austenitic phase transformation temperatures. RESULTS: SEM observations on new instruments revealed a smooth regular surface with flattened milling grooves. No wear features were detected after SCU. OCurve exhibited a higher cyclic fatigue resistance (P<0.05), slower crack propagation and a surface layer of TiO(2). Metallographic analysis and XRD showed the prevalence of martensitic grains on OCurve instruments that were stable at body temperature as confirmed by DSC analysis. Furthermore, DSC demonstrated a shift in the temperature transformation ranges suggesting an increase of martensite phase in autoclaved OCurve instruments. CONCLUSION: Heat treatment processes were confirmed as a valid enhancement of the properties of the new generation NiTi instruments. OCurve presented a significant improvement over OShape regarding both mechanical and metallurgical characteristics.
format Online
Article
Text
id pubmed-8056813
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Kare Publishing
record_format MEDLINE/PubMed
spelling pubmed-80568132021-04-21 In-Depth Metallurgical and Microstructural Analysis of Oneshape and Heat Treated Onecurve Instruments Azizi, Arash Prati, Carlo Schiavon, Riccardo Fitzgibbon, Raquel Pirani, Chiara Iacono, Francesco Pelliccioni, Gian Andrea Spinelli, Andrea Zamparini, Fausto Puddu, Pietro Bolelli, Giovanni Generali, Luigi Eur Endod J Original Article OBJECTIVE: To define surface, mechanical, microstructural and metallurgical features of conventional OneShape (OShape) and heat-treated OneCurve (OCurve) nickel-titanium instruments. METHODS: Instruments were analysed by scanning electron microscopy (SEM) on new instruments and after simulated clinical use (SCU). Cyclic fatigue testing was performed and the number of cycles to fracture (NCF) and the length of the fractured instruments were measured (Mann-Whitney test). Fractured instruments during cyclic fatigue testing were then inspected by SEM fractographic analysis. Field emission gun scanning scanning electron microscopy (FEG-SEM), energy-dispersive X-ray spectroscopy (EDX) and micro-Raman spectroscopy were used to assess alloy surface chemistry. Focused ion beam (FIB) was performed to analyse the oxide layer on the surface of OCurve before and after SCU. X-Ray diffraction (XRD), metallographic evaluation and differential scanning calorimetry (DSC) were used to determine martensitic/austenitic phase transformation temperatures. RESULTS: SEM observations on new instruments revealed a smooth regular surface with flattened milling grooves. No wear features were detected after SCU. OCurve exhibited a higher cyclic fatigue resistance (P<0.05), slower crack propagation and a surface layer of TiO(2). Metallographic analysis and XRD showed the prevalence of martensitic grains on OCurve instruments that were stable at body temperature as confirmed by DSC analysis. Furthermore, DSC demonstrated a shift in the temperature transformation ranges suggesting an increase of martensite phase in autoclaved OCurve instruments. CONCLUSION: Heat treatment processes were confirmed as a valid enhancement of the properties of the new generation NiTi instruments. OCurve presented a significant improvement over OShape regarding both mechanical and metallurgical characteristics. Kare Publishing 2021-03-23 /pmc/articles/PMC8056813/ /pubmed/33762534 http://dx.doi.org/10.14744/eej.2020.63634 Text en Copyright: © 2021 European Endodontic Journal https://creativecommons.org/licenses/by-nc/4.0/This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License
spellingShingle Original Article
Azizi, Arash
Prati, Carlo
Schiavon, Riccardo
Fitzgibbon, Raquel
Pirani, Chiara
Iacono, Francesco
Pelliccioni, Gian Andrea
Spinelli, Andrea
Zamparini, Fausto
Puddu, Pietro
Bolelli, Giovanni
Generali, Luigi
In-Depth Metallurgical and Microstructural Analysis of Oneshape and Heat Treated Onecurve Instruments
title In-Depth Metallurgical and Microstructural Analysis of Oneshape and Heat Treated Onecurve Instruments
title_full In-Depth Metallurgical and Microstructural Analysis of Oneshape and Heat Treated Onecurve Instruments
title_fullStr In-Depth Metallurgical and Microstructural Analysis of Oneshape and Heat Treated Onecurve Instruments
title_full_unstemmed In-Depth Metallurgical and Microstructural Analysis of Oneshape and Heat Treated Onecurve Instruments
title_short In-Depth Metallurgical and Microstructural Analysis of Oneshape and Heat Treated Onecurve Instruments
title_sort in-depth metallurgical and microstructural analysis of oneshape and heat treated onecurve instruments
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8056813/
https://www.ncbi.nlm.nih.gov/pubmed/33762534
http://dx.doi.org/10.14744/eej.2020.63634
work_keys_str_mv AT aziziarash indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT praticarlo indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT schiavonriccardo indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT fitzgibbonraquel indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT piranichiara indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT iaconofrancesco indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT pelliccionigianandrea indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT spinelliandrea indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT zamparinifausto indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT puddupietro indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT bolelligiovanni indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments
AT generaliluigi indepthmetallurgicalandmicrostructuralanalysisofoneshapeandheattreatedonecurveinstruments