Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing
BACKGROUND: With regard to biomechanical testing of orthopaedic implants, there is no consensus on whether artificial creation of standardized bone fractures or their simulation by means of osteotomies result in more realistic outcomes. Therefore, the aim of this study was to artificially create and...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8058974/ https://www.ncbi.nlm.nih.gov/pubmed/33879133 http://dx.doi.org/10.1186/s12891-021-04234-4 |
_version_ | 1783681116773810176 |
---|---|
author | Schader, J.F. Zderic, I. Gehweiler, D. Dauwe, J. Mys, K. Danker, C. Acklin, Y. P. Sommer, C. Gueorguiev, B. Stoffel, K. |
author_facet | Schader, J.F. Zderic, I. Gehweiler, D. Dauwe, J. Mys, K. Danker, C. Acklin, Y. P. Sommer, C. Gueorguiev, B. Stoffel, K. |
author_sort | Schader, J.F. |
collection | PubMed |
description | BACKGROUND: With regard to biomechanical testing of orthopaedic implants, there is no consensus on whether artificial creation of standardized bone fractures or their simulation by means of osteotomies result in more realistic outcomes. Therefore, the aim of this study was to artificially create and analyze in an appropriate setting the biomechanical behavior of standardized stable pertrochanteric fractures versus their simulation via osteotomizing. METHODS: Eight pairs of fresh-frozen human cadaveric femora aged 72.7 ± 14.9 years (range 48–89 years) were assigned in paired fashion to two study groups. In Group 1, stable pertrochanteric fractures AO/OTA 31-A1 were artificially created via constant force application on the anterior cortex of the femur through a blunt guillotine blade. The same fracture type was simulated in Group 2 by means of osteotomies. All femora were implanted with a dynamic hip screw and biomechanically tested in 20° adduction under progressively increasing physiologic cyclic axial loading at 2 Hz, starting at 500 N and increasing at a rate of 0.1 N/cycle. Femoral head fragment movements with respect to the shaft were monitored by means of optical motion tracking. RESULTS: Cycles/failure load at 15° varus deformation, 10 mm leg shortening and 15° femoral head rotation around neck axis were 11324 ± 848/1632.4 ± 584.8 N, 11052 ± 1573/1605.2 ± 657.3 N and 11849 ± 1120/1684.9 ± 612.0 N in Group 1, and 10971 ± 2019/1597.1 ± 701.9 N, 10681 ± 1868/1568.1 ± 686.8 N and 10017 ± 4081/1501.7 ± 908.1 N in Group 2, respectively, with no significant differences between the two groups, p ≥ 0.233. CONCLUSION: From a biomechanical perspective, by resulting in more consistent outcomes under dynamic loading, standardized artificial stable pertrochanteric femur fracture creation may be more suitable for orthopaedic implant testing compared to osteotomizing the bone. |
format | Online Article Text |
id | pubmed-8058974 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-80589742021-04-21 Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing Schader, J.F. Zderic, I. Gehweiler, D. Dauwe, J. Mys, K. Danker, C. Acklin, Y. P. Sommer, C. Gueorguiev, B. Stoffel, K. BMC Musculoskelet Disord Research Article BACKGROUND: With regard to biomechanical testing of orthopaedic implants, there is no consensus on whether artificial creation of standardized bone fractures or their simulation by means of osteotomies result in more realistic outcomes. Therefore, the aim of this study was to artificially create and analyze in an appropriate setting the biomechanical behavior of standardized stable pertrochanteric fractures versus their simulation via osteotomizing. METHODS: Eight pairs of fresh-frozen human cadaveric femora aged 72.7 ± 14.9 years (range 48–89 years) were assigned in paired fashion to two study groups. In Group 1, stable pertrochanteric fractures AO/OTA 31-A1 were artificially created via constant force application on the anterior cortex of the femur through a blunt guillotine blade. The same fracture type was simulated in Group 2 by means of osteotomies. All femora were implanted with a dynamic hip screw and biomechanically tested in 20° adduction under progressively increasing physiologic cyclic axial loading at 2 Hz, starting at 500 N and increasing at a rate of 0.1 N/cycle. Femoral head fragment movements with respect to the shaft were monitored by means of optical motion tracking. RESULTS: Cycles/failure load at 15° varus deformation, 10 mm leg shortening and 15° femoral head rotation around neck axis were 11324 ± 848/1632.4 ± 584.8 N, 11052 ± 1573/1605.2 ± 657.3 N and 11849 ± 1120/1684.9 ± 612.0 N in Group 1, and 10971 ± 2019/1597.1 ± 701.9 N, 10681 ± 1868/1568.1 ± 686.8 N and 10017 ± 4081/1501.7 ± 908.1 N in Group 2, respectively, with no significant differences between the two groups, p ≥ 0.233. CONCLUSION: From a biomechanical perspective, by resulting in more consistent outcomes under dynamic loading, standardized artificial stable pertrochanteric femur fracture creation may be more suitable for orthopaedic implant testing compared to osteotomizing the bone. BioMed Central 2021-04-20 /pmc/articles/PMC8058974/ /pubmed/33879133 http://dx.doi.org/10.1186/s12891-021-04234-4 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Article Schader, J.F. Zderic, I. Gehweiler, D. Dauwe, J. Mys, K. Danker, C. Acklin, Y. P. Sommer, C. Gueorguiev, B. Stoffel, K. Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing |
title | Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing |
title_full | Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing |
title_fullStr | Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing |
title_full_unstemmed | Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing |
title_short | Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing |
title_sort | standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8058974/ https://www.ncbi.nlm.nih.gov/pubmed/33879133 http://dx.doi.org/10.1186/s12891-021-04234-4 |
work_keys_str_mv | AT schaderjf standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting AT zderici standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting AT gehweilerd standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting AT dauwej standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting AT mysk standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting AT dankerc standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting AT acklinyp standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting AT sommerc standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting AT gueorguievb standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting AT stoffelk standardizedartificiallycreatedstablepertrochantericfemurfracturespresentmorehomogenousresultscomparedtoosteotomiesfororthopaedicimplanttesting |