Dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy

PURPOSE: 3D printing has become a popular and widely available technique of rapid prototyping. The impact of used materials on the dose distribution has been studied for high energy sources. However, brachytherapy sources emit lower energy photons, and materials used in 3D printing may differ. This...

Descripción completa

Detalles Bibliográficos
Autores principales: Bielęda, Grzegorz, Zwierzchowski, Grzegorz, Rosłan, Katarzyna, Adamus, Agnieszka, Malicki, Julian
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Termedia Publishing House 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8060963/
https://www.ncbi.nlm.nih.gov/pubmed/33897793
http://dx.doi.org/10.5114/jcb.2021.105287
_version_ 1783681468438937600
author Bielęda, Grzegorz
Zwierzchowski, Grzegorz
Rosłan, Katarzyna
Adamus, Agnieszka
Malicki, Julian
author_facet Bielęda, Grzegorz
Zwierzchowski, Grzegorz
Rosłan, Katarzyna
Adamus, Agnieszka
Malicki, Julian
author_sort Bielęda, Grzegorz
collection PubMed
description PURPOSE: 3D printing has become a popular and widely available technique of rapid prototyping. The impact of used materials on the dose distribution has been studied for high energy sources. However, brachytherapy sources emit lower energy photons, and materials used in 3D printing may differ. This study was conducted to analyze the influence of common materials (polylactic acid – PLA and acrylonitrile butadiene styrene – ABS) used in stereolithography. MATERIAL AND METHODS: A 3D-printed phantom was designed, printed, and used to calibrate Gafchromic films. In the next step, a range of 1 mm thick plates of PLA and ABS (from zero to thirty) were inserted between source and detector to measure the impact of studied materials on delivered dose. Measurements were performed using a calibrated radiochromic film and Farmer ionization chamber in water. RESULTS: No statistically significant correlation (p = 0.4854) between the thickness of inserted PLA and the dose delivered to the film was obtained. With ionization chamber, Spearman’s rank order test showed a significant correlation (p = 0.00004); however, the correlation was found weak. In case of ABS measurement, a statistically significant (p = 0.0159), yet weak negative correlation was found between the thickness of used material and the dose delivered to the film. On the other hand, a weak statistically significant (p = 0.0212) but positive correlation was found when the dose was measured with Farmer ionization chamber. We find these correlations false, as all measured doses were within the measurement uncertainty range (film ±8.0%, Farmer ±8.8%) from 100% of the prescribed dose. CONCLUSIONS: According to obtained results, with the accuracy of measurement under clinical conditions, the impact of highly filled PLA and ABS printed objects on the dose distribution from an (192)Ir source in water can be omitted.
format Online
Article
Text
id pubmed-8060963
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Termedia Publishing House
record_format MEDLINE/PubMed
spelling pubmed-80609632021-04-23 Dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy Bielęda, Grzegorz Zwierzchowski, Grzegorz Rosłan, Katarzyna Adamus, Agnieszka Malicki, Julian J Contemp Brachytherapy Original Paper PURPOSE: 3D printing has become a popular and widely available technique of rapid prototyping. The impact of used materials on the dose distribution has been studied for high energy sources. However, brachytherapy sources emit lower energy photons, and materials used in 3D printing may differ. This study was conducted to analyze the influence of common materials (polylactic acid – PLA and acrylonitrile butadiene styrene – ABS) used in stereolithography. MATERIAL AND METHODS: A 3D-printed phantom was designed, printed, and used to calibrate Gafchromic films. In the next step, a range of 1 mm thick plates of PLA and ABS (from zero to thirty) were inserted between source and detector to measure the impact of studied materials on delivered dose. Measurements were performed using a calibrated radiochromic film and Farmer ionization chamber in water. RESULTS: No statistically significant correlation (p = 0.4854) between the thickness of inserted PLA and the dose delivered to the film was obtained. With ionization chamber, Spearman’s rank order test showed a significant correlation (p = 0.00004); however, the correlation was found weak. In case of ABS measurement, a statistically significant (p = 0.0159), yet weak negative correlation was found between the thickness of used material and the dose delivered to the film. On the other hand, a weak statistically significant (p = 0.0212) but positive correlation was found when the dose was measured with Farmer ionization chamber. We find these correlations false, as all measured doses were within the measurement uncertainty range (film ±8.0%, Farmer ±8.8%) from 100% of the prescribed dose. CONCLUSIONS: According to obtained results, with the accuracy of measurement under clinical conditions, the impact of highly filled PLA and ABS printed objects on the dose distribution from an (192)Ir source in water can be omitted. Termedia Publishing House 2021-04-14 2021-04 /pmc/articles/PMC8060963/ /pubmed/33897793 http://dx.doi.org/10.5114/jcb.2021.105287 Text en Copyright © 2021 Termedia https://creativecommons.org/licenses/by-nc-sa/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/ (https://creativecommons.org/licenses/by-nc-sa/4.0/) )
spellingShingle Original Paper
Bielęda, Grzegorz
Zwierzchowski, Grzegorz
Rosłan, Katarzyna
Adamus, Agnieszka
Malicki, Julian
Dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy
title Dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy
title_full Dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy
title_fullStr Dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy
title_full_unstemmed Dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy
title_short Dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy
title_sort dosimetric assessment of the impact of low-cost materials used in stereolithography in high-dose-rate brachytherapy
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8060963/
https://www.ncbi.nlm.nih.gov/pubmed/33897793
http://dx.doi.org/10.5114/jcb.2021.105287
work_keys_str_mv AT bieledagrzegorz dosimetricassessmentoftheimpactoflowcostmaterialsusedinstereolithographyinhighdoseratebrachytherapy
AT zwierzchowskigrzegorz dosimetricassessmentoftheimpactoflowcostmaterialsusedinstereolithographyinhighdoseratebrachytherapy
AT rosłankatarzyna dosimetricassessmentoftheimpactoflowcostmaterialsusedinstereolithographyinhighdoseratebrachytherapy
AT adamusagnieszka dosimetricassessmentoftheimpactoflowcostmaterialsusedinstereolithographyinhighdoseratebrachytherapy
AT malickijulian dosimetricassessmentoftheimpactoflowcostmaterialsusedinstereolithographyinhighdoseratebrachytherapy