Cargando…

Microfluidic Device for the Identification of Biological Sex by Analysis of Latent Fingermark Deposits

To date, most research regarding amino acid detection and quantification in fingermarks relies on spectrometric methods. Herein, the Sakaguchi colorimetric test was adapted to a rotationally-driven microfluidic platform and used to detect and quantify arginine in fingermarks deposited by male and fe...

Descripción completa

Detalles Bibliográficos
Autores principales: Marshall, Jamila S., Sita, Madelyn L., Landers, James P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8071498/
https://www.ncbi.nlm.nih.gov/pubmed/33921163
http://dx.doi.org/10.3390/mi12040442
Descripción
Sumario:To date, most research regarding amino acid detection and quantification in fingermarks relies on spectrometric methods. Herein, the Sakaguchi colorimetric test was adapted to a rotationally-driven microfluidic platform and used to detect and quantify arginine in fingermarks deposited by male and female donors. A red color indicates the presence of arginine in a given sample following the reaction, and the intensity of this color is linearly proportional to the concentration. Objective detection and quantification of arginine were accomplished using image analysis software (freeware) based on this colorimetric result. The mean concentrations obtained in a blind study were 96.4 ± 5.1 µM for samples from female donors and 55.3 ± 5.3 µM for samples from males. These were not statistically different from the literature values of 94.8 µM ± 12.9 µM for females (p = 0.908) and 54.0 ± 12.6 µM for males (p = 0.914), respectively (± SEM in all cases). Conversely, the experimental means from males and female samples were statistically different from each other (p < 0.001). Objective differentiation between male and female fingermark deposits was achieved in a blind study with 93% accuracy. Additionally, the method was compatible both with samples lifted from common surfaces and with magnetically-powdered samples.