Biomechanical comparison of pullout strengths of six pedicle screws with different thread designs
OBJECTIVES: This study aims to assess the pullout strength of six different pedicle screw thread patterns. MATERIALS AND METHODS: A total of 36 sheep spines were divided into six groups including six spines in each group: fully threaded cortical (Type A), fully threaded spongeous (Type B), fully cor...
Autores principales: | Karakaşlı, Ahmet, Acar, Nihat, Hüsemoğlu, R. Buğra |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Bayçınar Medical Publishing
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8073434/ https://www.ncbi.nlm.nih.gov/pubmed/33463436 http://dx.doi.org/10.5606/ehc.2021.77004 |
Ejemplares similares
-
The Influence of Thread Tap Mismatch on Pedicle Screw Pullout Strength
por: Pinheiro, Rômulo Pedroza, et al.
Publicado: (2022) -
The evaluation of a de novo biplanar distal humerus plate: A biomechanical study
por: Acar, Nihat, et al.
Publicado: (2022) -
DIRECT PEDICLE SCREW INSERTION PULLOUT STRENGTH
por: PINHEIRO, RÔMULO PEDROZA, et al.
Publicado: (2021) -
Comparison of the Pullout Strength of Different Pedicle Screw Designs and Augmentation Techniques in an Osteoporotic Bone Model
por: Kiyak, Gorkem, et al.
Publicado: (2018) -
Axial pullout strength comparison of different screw designs: fenestrated screw, dual outer diameter screw and standard pedicle screw
por: Christodoulou, Evangelos, et al.
Publicado: (2015)