Cargando…

Accuracy and reliability of measurements obtained with a noncontact tono-pachymeter for clinical use in mass screening

We evaluated the reliability and accuracy of the noncontact CT-1P tonopachymeter (Topcon, Japan) in terms of intraocular pressure (IOP) and central corneal thickness (CCT). One hundred sixty-three healthy participants and 33 patients with open angle glaucoma were enrolled. IOPs were measured by CT-1...

Descripción completa

Detalles Bibliográficos
Autores principales: Lee, Jinho, Choi, Hyuk Jin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group UK 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8076298/
https://www.ncbi.nlm.nih.gov/pubmed/33903678
http://dx.doi.org/10.1038/s41598-021-88364-8
Descripción
Sumario:We evaluated the reliability and accuracy of the noncontact CT-1P tonopachymeter (Topcon, Japan) in terms of intraocular pressure (IOP) and central corneal thickness (CCT). One hundred sixty-three healthy participants and 33 patients with open angle glaucoma were enrolled. IOPs were measured by CT-1P (T-IOP) and Goldmann applanation tonometer (G-IOP), and CCTs were measured by the CT-1P (T-CCT) and an ultrasound pachymeter (US-CCT). Both CCT instrument-adjusted (T-IOP-C) and unadjusted T-IOPs (T-IOP-NC) were included. Pearson correlation coefficients and biases assessed with Bland–Altman analysis with 95% confidence interval (CI) were calculated for reliability evaluation. Intrasession repeatability was excellent for both T-IOP (intraclass correlation coefficient [ICC] 0.91) and T-CCT (ICC 0.98). Intersession reproducibility was also excellent for T-CCT (ICC 0.94). T-IOP-NC and T-IOP-C both showed significant correlations with G-IOP (r = 0.801, P  <  0.001 and r = 0.658, P  < 0.001, respectively). T-CCT was also strongly correlated with US-CCT (r = 0.958; P  < 0.001). T-IOP-NC and T-IOP-C both showed a positive bias (1.37 mmHg, 95% CI [1.14, 1.61] and 2.77 mmHg, 95% CI [2.49, 3.05], respectively). T-CCT showed a negative bias of − 17.3 µm (95% CI [−18.8, − 15.8]). With cautious interpretation, the CT-1P may offer good feasibility for IOP and CCT measurement in screening centers.