Cargando…

Validating and implementing cardiac telemetry for continuous QTc monitoring: A novel approach to increase healthcare personnel safety during the COVID-19 pandemic

BACKGROUND: Minimizing direct patient contact among healthcare personnel is crucial for mitigating infectious risk during the coronavirus disease 2019 (COVID-19) pandemic. The use of remote cardiac telemetry as an alternative to 12‑lead electrocardiography (ECG) for continuous QTc monitoring may fac...

Descripción completa

Detalles Bibliográficos
Autores principales: Kassis, Nicholas, Tanaka-Esposito, Christine, Chung, Roy, Kalra, Ankur, Shao, Mingyuan, Kumar, Ashish, Alzubi, Jafar, Chung, Mina K., Khot, Umesh N.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Author(s). Published by Elsevier Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8076730/
https://www.ncbi.nlm.nih.gov/pubmed/33975077
http://dx.doi.org/10.1016/j.jelectrocard.2021.04.014
_version_ 1783684743476281344
author Kassis, Nicholas
Tanaka-Esposito, Christine
Chung, Roy
Kalra, Ankur
Shao, Mingyuan
Kumar, Ashish
Alzubi, Jafar
Chung, Mina K.
Khot, Umesh N.
author_facet Kassis, Nicholas
Tanaka-Esposito, Christine
Chung, Roy
Kalra, Ankur
Shao, Mingyuan
Kumar, Ashish
Alzubi, Jafar
Chung, Mina K.
Khot, Umesh N.
author_sort Kassis, Nicholas
collection PubMed
description BACKGROUND: Minimizing direct patient contact among healthcare personnel is crucial for mitigating infectious risk during the coronavirus disease 2019 (COVID-19) pandemic. The use of remote cardiac telemetry as an alternative to 12‑lead electrocardiography (ECG) for continuous QTc monitoring may facilitate this strategy, but its application has not yet been validated or implemented. METHODS: In the validation component of this two-part prospective cohort study, a total of 65 hospitalized patients with simultaneous ECG and telemetry were identified. QTc obtained via remote telemetry as measured by 3 independent, blinded operators were compared with ECG as assessed by 2 board-certified electrophysiologists as the gold-standard. Pearson correlation coefficients were calculated to measure the strength of linear correlation between the two methods. In a separate cohort comprised of 68 COVID-19 patients treated with combined hydroxychloroquine and azithromycin, telemetry-based QTc values were compared at serial time points after medication administration using Friedman rank-sum test of repeated measures. RESULTS: Telemetry-based QTc measurements highly correlated with QTc values derived from ECG, with correlation coefficients of 0.74, 0.79, 0.85 (individual operators), and 0.84 (mean of all operators). Among the COVID-19 cohort, treatment led to a median QTc increase of 15 milliseconds between baseline and following the 9th dose (p = 0.002), with 8 (12%) patients exhibiting an increase in QTc ≥ 60 milliseconds and 4 (6%) developing QTc ≥ 500 milliseconds. CONCLUSIONS: Cardiac telemetry is a validated clinical tool for QTc monitoring that may serve an expanding role during the COVID-19 pandemic strengthened by its remote and continuous monitoring capability and ubiquitous presence throughout hospitals.
format Online
Article
Text
id pubmed-8076730
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher The Author(s). Published by Elsevier Inc.
record_format MEDLINE/PubMed
spelling pubmed-80767302021-04-27 Validating and implementing cardiac telemetry for continuous QTc monitoring: A novel approach to increase healthcare personnel safety during the COVID-19 pandemic Kassis, Nicholas Tanaka-Esposito, Christine Chung, Roy Kalra, Ankur Shao, Mingyuan Kumar, Ashish Alzubi, Jafar Chung, Mina K. Khot, Umesh N. J Electrocardiol Article BACKGROUND: Minimizing direct patient contact among healthcare personnel is crucial for mitigating infectious risk during the coronavirus disease 2019 (COVID-19) pandemic. The use of remote cardiac telemetry as an alternative to 12‑lead electrocardiography (ECG) for continuous QTc monitoring may facilitate this strategy, but its application has not yet been validated or implemented. METHODS: In the validation component of this two-part prospective cohort study, a total of 65 hospitalized patients with simultaneous ECG and telemetry were identified. QTc obtained via remote telemetry as measured by 3 independent, blinded operators were compared with ECG as assessed by 2 board-certified electrophysiologists as the gold-standard. Pearson correlation coefficients were calculated to measure the strength of linear correlation between the two methods. In a separate cohort comprised of 68 COVID-19 patients treated with combined hydroxychloroquine and azithromycin, telemetry-based QTc values were compared at serial time points after medication administration using Friedman rank-sum test of repeated measures. RESULTS: Telemetry-based QTc measurements highly correlated with QTc values derived from ECG, with correlation coefficients of 0.74, 0.79, 0.85 (individual operators), and 0.84 (mean of all operators). Among the COVID-19 cohort, treatment led to a median QTc increase of 15 milliseconds between baseline and following the 9th dose (p = 0.002), with 8 (12%) patients exhibiting an increase in QTc ≥ 60 milliseconds and 4 (6%) developing QTc ≥ 500 milliseconds. CONCLUSIONS: Cardiac telemetry is a validated clinical tool for QTc monitoring that may serve an expanding role during the COVID-19 pandemic strengthened by its remote and continuous monitoring capability and ubiquitous presence throughout hospitals. The Author(s). Published by Elsevier Inc. 2021 2021-04-27 /pmc/articles/PMC8076730/ /pubmed/33975077 http://dx.doi.org/10.1016/j.jelectrocard.2021.04.014 Text en © 2021 The Author(s) Since January 2020 Elsevier has created a COVID-19 resource centre with free information in English and Mandarin on the novel coronavirus COVID-19. The COVID-19 resource centre is hosted on Elsevier Connect, the company's public news and information website. Elsevier hereby grants permission to make all its COVID-19-related research that is available on the COVID-19 resource centre - including this research content - immediately available in PubMed Central and other publicly funded repositories, such as the WHO COVID database with rights for unrestricted research re-use and analyses in any form or by any means with acknowledgement of the original source. These permissions are granted for free by Elsevier for as long as the COVID-19 resource centre remains active.
spellingShingle Article
Kassis, Nicholas
Tanaka-Esposito, Christine
Chung, Roy
Kalra, Ankur
Shao, Mingyuan
Kumar, Ashish
Alzubi, Jafar
Chung, Mina K.
Khot, Umesh N.
Validating and implementing cardiac telemetry for continuous QTc monitoring: A novel approach to increase healthcare personnel safety during the COVID-19 pandemic
title Validating and implementing cardiac telemetry for continuous QTc monitoring: A novel approach to increase healthcare personnel safety during the COVID-19 pandemic
title_full Validating and implementing cardiac telemetry for continuous QTc monitoring: A novel approach to increase healthcare personnel safety during the COVID-19 pandemic
title_fullStr Validating and implementing cardiac telemetry for continuous QTc monitoring: A novel approach to increase healthcare personnel safety during the COVID-19 pandemic
title_full_unstemmed Validating and implementing cardiac telemetry for continuous QTc monitoring: A novel approach to increase healthcare personnel safety during the COVID-19 pandemic
title_short Validating and implementing cardiac telemetry for continuous QTc monitoring: A novel approach to increase healthcare personnel safety during the COVID-19 pandemic
title_sort validating and implementing cardiac telemetry for continuous qtc monitoring: a novel approach to increase healthcare personnel safety during the covid-19 pandemic
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8076730/
https://www.ncbi.nlm.nih.gov/pubmed/33975077
http://dx.doi.org/10.1016/j.jelectrocard.2021.04.014
work_keys_str_mv AT kassisnicholas validatingandimplementingcardiactelemetryforcontinuousqtcmonitoringanovelapproachtoincreasehealthcarepersonnelsafetyduringthecovid19pandemic
AT tanakaespositochristine validatingandimplementingcardiactelemetryforcontinuousqtcmonitoringanovelapproachtoincreasehealthcarepersonnelsafetyduringthecovid19pandemic
AT chungroy validatingandimplementingcardiactelemetryforcontinuousqtcmonitoringanovelapproachtoincreasehealthcarepersonnelsafetyduringthecovid19pandemic
AT kalraankur validatingandimplementingcardiactelemetryforcontinuousqtcmonitoringanovelapproachtoincreasehealthcarepersonnelsafetyduringthecovid19pandemic
AT shaomingyuan validatingandimplementingcardiactelemetryforcontinuousqtcmonitoringanovelapproachtoincreasehealthcarepersonnelsafetyduringthecovid19pandemic
AT kumarashish validatingandimplementingcardiactelemetryforcontinuousqtcmonitoringanovelapproachtoincreasehealthcarepersonnelsafetyduringthecovid19pandemic
AT alzubijafar validatingandimplementingcardiactelemetryforcontinuousqtcmonitoringanovelapproachtoincreasehealthcarepersonnelsafetyduringthecovid19pandemic
AT chungminak validatingandimplementingcardiactelemetryforcontinuousqtcmonitoringanovelapproachtoincreasehealthcarepersonnelsafetyduringthecovid19pandemic
AT khotumeshn validatingandimplementingcardiactelemetryforcontinuousqtcmonitoringanovelapproachtoincreasehealthcarepersonnelsafetyduringthecovid19pandemic