Validation of the STEP deflation algorithm of the Midmark IQvitals Zone Vital Signs Monitor: part of a novel clinical ecosystem
OBJECTIVES: Assess the accuracy of the Midmark IQvitals Zone Vital Signs Monitor STEP deflation algorithm according to the ANSI/AAMI/ISO 81060-2 Standard. METHODS: A total of 85 subjects completed the testing protocol. All standard requirements for gender, blood pressure (BP) values, and arm circumf...
Autor principal: | |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Lippincott Williams & Wilkins
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8081442/ https://www.ncbi.nlm.nih.gov/pubmed/33590995 http://dx.doi.org/10.1097/MBP.0000000000000524 |
_version_ | 1783685640502640640 |
---|---|
author | Alpert, Bruce S. |
author_facet | Alpert, Bruce S. |
author_sort | Alpert, Bruce S. |
collection | PubMed |
description | OBJECTIVES: Assess the accuracy of the Midmark IQvitals Zone Vital Signs Monitor STEP deflation algorithm according to the ANSI/AAMI/ISO 81060-2 Standard. METHODS: A total of 85 subjects completed the testing protocol. All standard requirements for gender, blood pressure (BP) values, and arm circumferences were met. Manual auscultation was performed by testers blinded to the device; the manual BP values were compared to the device readings. RESULTS: The Standard Criterion 1 data analyses showed mean ± SD device minus manual BP values of 1.22 ± 6.3 mmHg for SBP and −1.67 ± 6.09 mmHg for DBP. The SD values for criterion 2 were 5.06 mmHg (SBP) and 4.98 mmHg (DBP). CONCLUSIONS: The device passed all Standard requirements. The Midmark IQvitals Zone device has features to improve accuracy and reduce or eliminate transcription errors and inaccuracy from improper patient positioning. |
format | Online Article Text |
id | pubmed-8081442 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Lippincott Williams & Wilkins |
record_format | MEDLINE/PubMed |
spelling | pubmed-80814422021-05-04 Validation of the STEP deflation algorithm of the Midmark IQvitals Zone Vital Signs Monitor: part of a novel clinical ecosystem Alpert, Bruce S. Blood Press Monit Devices and Technology OBJECTIVES: Assess the accuracy of the Midmark IQvitals Zone Vital Signs Monitor STEP deflation algorithm according to the ANSI/AAMI/ISO 81060-2 Standard. METHODS: A total of 85 subjects completed the testing protocol. All standard requirements for gender, blood pressure (BP) values, and arm circumferences were met. Manual auscultation was performed by testers blinded to the device; the manual BP values were compared to the device readings. RESULTS: The Standard Criterion 1 data analyses showed mean ± SD device minus manual BP values of 1.22 ± 6.3 mmHg for SBP and −1.67 ± 6.09 mmHg for DBP. The SD values for criterion 2 were 5.06 mmHg (SBP) and 4.98 mmHg (DBP). CONCLUSIONS: The device passed all Standard requirements. The Midmark IQvitals Zone device has features to improve accuracy and reduce or eliminate transcription errors and inaccuracy from improper patient positioning. Lippincott Williams & Wilkins 2021-02-12 2021-06 /pmc/articles/PMC8081442/ /pubmed/33590995 http://dx.doi.org/10.1097/MBP.0000000000000524 Text en Copyright © 2021 The Author(s). Published by Wolters Kluwer Health, Inc. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (https://creativecommons.org/licenses/by-nc-nd/4.0/) (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal. |
spellingShingle | Devices and Technology Alpert, Bruce S. Validation of the STEP deflation algorithm of the Midmark IQvitals Zone Vital Signs Monitor: part of a novel clinical ecosystem |
title | Validation of the STEP deflation algorithm of the Midmark IQvitals Zone Vital Signs Monitor: part of a novel clinical ecosystem |
title_full | Validation of the STEP deflation algorithm of the Midmark IQvitals Zone Vital Signs Monitor: part of a novel clinical ecosystem |
title_fullStr | Validation of the STEP deflation algorithm of the Midmark IQvitals Zone Vital Signs Monitor: part of a novel clinical ecosystem |
title_full_unstemmed | Validation of the STEP deflation algorithm of the Midmark IQvitals Zone Vital Signs Monitor: part of a novel clinical ecosystem |
title_short | Validation of the STEP deflation algorithm of the Midmark IQvitals Zone Vital Signs Monitor: part of a novel clinical ecosystem |
title_sort | validation of the step deflation algorithm of the midmark iqvitals zone vital signs monitor: part of a novel clinical ecosystem |
topic | Devices and Technology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8081442/ https://www.ncbi.nlm.nih.gov/pubmed/33590995 http://dx.doi.org/10.1097/MBP.0000000000000524 |
work_keys_str_mv | AT alpertbruces validationofthestepdeflationalgorithmofthemidmarkiqvitalszonevitalsignsmonitorpartofanovelclinicalecosystem |