Smart Eye Camera: A Validation Study for Evaluating the Tear Film Breakup Time in Human Subjects

PURPOSE: This study aimed to demonstrate the efficacy of a “Smart Eye Camera (SEC)” in comparison with the efficacy of the conventional slit-lamp microscope by evaluating their diagnostic functionality for dry eye disease (DED) in clinical cases. METHODS: This retrospective study included 106 eyes f...

Descripción completa

Detalles Bibliográficos
Autores principales: Shimizu, Eisuke, Yazu, Hiroyuki, Aketa, Naohiko, Yokoiwa, Ryota, Sato, Shinri, Katayama, Taiichiro, Hanyuda, Akiko, Sato, Yasunori, Ogawa, Yoko, Tsubota, Kazuo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Association for Research in Vision and Ophthalmology 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8083120/
https://www.ncbi.nlm.nih.gov/pubmed/34004005
http://dx.doi.org/10.1167/tvst.10.4.28
_version_ 1783685970624774144
author Shimizu, Eisuke
Yazu, Hiroyuki
Aketa, Naohiko
Yokoiwa, Ryota
Sato, Shinri
Katayama, Taiichiro
Hanyuda, Akiko
Sato, Yasunori
Ogawa, Yoko
Tsubota, Kazuo
author_facet Shimizu, Eisuke
Yazu, Hiroyuki
Aketa, Naohiko
Yokoiwa, Ryota
Sato, Shinri
Katayama, Taiichiro
Hanyuda, Akiko
Sato, Yasunori
Ogawa, Yoko
Tsubota, Kazuo
author_sort Shimizu, Eisuke
collection PubMed
description PURPOSE: This study aimed to demonstrate the efficacy of a “Smart Eye Camera (SEC)” in comparison with the efficacy of the conventional slit-lamp microscope by evaluating their diagnostic functionality for dry eye disease (DED) in clinical cases. METHODS: This retrospective study included 106 eyes from 53 adult Japanese patients who visited the Ophthalmology outpatient clinics in Keio University Hospital from June 2019 to March 2020. Tear film breakup time (TFBUT) and corneal fluorescence score (CFS) measurements for the diagnosis of DED were compared between the conventional slit-lamp microscope and SEC. RESULTS: The objective findings of DED showed that there was a strong correlation between the conventional slit-lamp microscope and SEC with respect to TFBUT and CFS results (Spearman's r = 0.887, 95% confidence interval [CI] = 0.838–0.922, and r = 0.920, 95% CI = 0.884–0.945, respectively). The interobserver reliability between the conventional slit-lamp microscope and SEC showed a moderate agreement (weighted Kappa κ = 0.527, 95% CI = 0.517–0.537 and κ = 0.550, 95% CI = 0.539–0.561 for TFBUT and CFS, respectively). The diagnostic performance of the SEC for Asia Dry Eye Society diagnostic criteria showed a sensitivity of 0.957 (95% CI = 0.841–0.992), specificity of 0.900 (95% CI = 0.811–0.927), positive predictive value of 0.880 (95% CI = 0.774–0.912), and negative predictive value of 0.964 (95% CI = 0.869–0.993). Moreover, the area under the receiver operating characteristic curve was 0.928 (95% CI = 0.849–1.000). CONCLUSIONS: Compared with the conventional slit-lamp microscope, SEC has sufficient validity and reliability for diagnosing DED in the clinical setting. TRANSLATIONAL RELEVANCE: The SEC can portably evaluate TFBUT in both basic research and clinical care.
format Online
Article
Text
id pubmed-8083120
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher The Association for Research in Vision and Ophthalmology
record_format MEDLINE/PubMed
spelling pubmed-80831202021-05-05 Smart Eye Camera: A Validation Study for Evaluating the Tear Film Breakup Time in Human Subjects Shimizu, Eisuke Yazu, Hiroyuki Aketa, Naohiko Yokoiwa, Ryota Sato, Shinri Katayama, Taiichiro Hanyuda, Akiko Sato, Yasunori Ogawa, Yoko Tsubota, Kazuo Transl Vis Sci Technol Article PURPOSE: This study aimed to demonstrate the efficacy of a “Smart Eye Camera (SEC)” in comparison with the efficacy of the conventional slit-lamp microscope by evaluating their diagnostic functionality for dry eye disease (DED) in clinical cases. METHODS: This retrospective study included 106 eyes from 53 adult Japanese patients who visited the Ophthalmology outpatient clinics in Keio University Hospital from June 2019 to March 2020. Tear film breakup time (TFBUT) and corneal fluorescence score (CFS) measurements for the diagnosis of DED were compared between the conventional slit-lamp microscope and SEC. RESULTS: The objective findings of DED showed that there was a strong correlation between the conventional slit-lamp microscope and SEC with respect to TFBUT and CFS results (Spearman's r = 0.887, 95% confidence interval [CI] = 0.838–0.922, and r = 0.920, 95% CI = 0.884–0.945, respectively). The interobserver reliability between the conventional slit-lamp microscope and SEC showed a moderate agreement (weighted Kappa κ = 0.527, 95% CI = 0.517–0.537 and κ = 0.550, 95% CI = 0.539–0.561 for TFBUT and CFS, respectively). The diagnostic performance of the SEC for Asia Dry Eye Society diagnostic criteria showed a sensitivity of 0.957 (95% CI = 0.841–0.992), specificity of 0.900 (95% CI = 0.811–0.927), positive predictive value of 0.880 (95% CI = 0.774–0.912), and negative predictive value of 0.964 (95% CI = 0.869–0.993). Moreover, the area under the receiver operating characteristic curve was 0.928 (95% CI = 0.849–1.000). CONCLUSIONS: Compared with the conventional slit-lamp microscope, SEC has sufficient validity and reliability for diagnosing DED in the clinical setting. TRANSLATIONAL RELEVANCE: The SEC can portably evaluate TFBUT in both basic research and clinical care. The Association for Research in Vision and Ophthalmology 2021-04-26 /pmc/articles/PMC8083120/ /pubmed/34004005 http://dx.doi.org/10.1167/tvst.10.4.28 Text en Copyright 2021 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
spellingShingle Article
Shimizu, Eisuke
Yazu, Hiroyuki
Aketa, Naohiko
Yokoiwa, Ryota
Sato, Shinri
Katayama, Taiichiro
Hanyuda, Akiko
Sato, Yasunori
Ogawa, Yoko
Tsubota, Kazuo
Smart Eye Camera: A Validation Study for Evaluating the Tear Film Breakup Time in Human Subjects
title Smart Eye Camera: A Validation Study for Evaluating the Tear Film Breakup Time in Human Subjects
title_full Smart Eye Camera: A Validation Study for Evaluating the Tear Film Breakup Time in Human Subjects
title_fullStr Smart Eye Camera: A Validation Study for Evaluating the Tear Film Breakup Time in Human Subjects
title_full_unstemmed Smart Eye Camera: A Validation Study for Evaluating the Tear Film Breakup Time in Human Subjects
title_short Smart Eye Camera: A Validation Study for Evaluating the Tear Film Breakup Time in Human Subjects
title_sort smart eye camera: a validation study for evaluating the tear film breakup time in human subjects
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8083120/
https://www.ncbi.nlm.nih.gov/pubmed/34004005
http://dx.doi.org/10.1167/tvst.10.4.28
work_keys_str_mv AT shimizueisuke smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects
AT yazuhiroyuki smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects
AT aketanaohiko smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects
AT yokoiwaryota smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects
AT satoshinri smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects
AT katayamataiichiro smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects
AT hanyudaakiko smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects
AT satoyasunori smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects
AT ogawayoko smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects
AT tsubotakazuo smarteyecameraavalidationstudyforevaluatingthetearfilmbreakuptimeinhumansubjects