Dose accuracy improvement on head and neck VMAT treatments by using the Acuros algorithm and accurate FFF beam calibration

BACKGROUND: The purpose of this study was to assess dose accuracy improvement and dosimetric impact of switching from the anisotropic analytical algorithm (AA) to the Acuros XB algorithm (AXB) when performing an accurate beam calibration in head and neck (H&N) FFF-VMAT treatments. MATERIALS AND...

Descripción completa

Detalles Bibliográficos
Autores principales: Martin-Martin, Guadalupe, Walter, Stefan, Guibelalde, Eduardo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Via Medica 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8086705/
https://www.ncbi.nlm.nih.gov/pubmed/33948305
http://dx.doi.org/10.5603/RPOR.a2021.0014
_version_ 1783686559593136128
author Martin-Martin, Guadalupe
Walter, Stefan
Guibelalde, Eduardo
author_facet Martin-Martin, Guadalupe
Walter, Stefan
Guibelalde, Eduardo
author_sort Martin-Martin, Guadalupe
collection PubMed
description BACKGROUND: The purpose of this study was to assess dose accuracy improvement and dosimetric impact of switching from the anisotropic analytical algorithm (AA) to the Acuros XB algorithm (AXB) when performing an accurate beam calibration in head and neck (H&N) FFF-VMAT treatments. MATERIALS AND METHODS: Twenty H&N cancer patients treated with FFF-VMAT techniques were included. Calculations were performed with the AA and AXB algorithm (dose-to-water — AXB(w)– and dose-to-medium — AXB(m)–). An accurate beam calibration was used for AXB calculations. Dose prescription to the tumour (PTV70) and at-risk-nodal region (PTV58.1) were 70 Gy and 58.1 Gy, respectively. A PTV70(_bone) including bony structures in PTV70 was contoured. Dose-volume parameters were compared between the algorithms. Statistical tests were used to analyze the differences in mean values and the correlation between compliance with the D95 > 95% requirement and occurrence of local recurrence. RESULTS: AA systematically overestimated the dose compared to AXB algorithm with mean dose differences within 1.3 Gy/2%, except for the PTV70(_bone) (2.2 Gy/3.2%). Dose differences were significantly higher for AXB(m) calculations when including accurate beam calibration (maximum dose differences up to 2.8 Gy/4.1% and 4.2 Gy/6.3% for PTV70 and PTV70(_bone), respectively). 80% of AA-calculated plans did not meet the D95 > 95% requirement after recalculation with AXB(m) and accurate beam calibration. The reduction in D95 coverage in the tumour was not clinically relevant. CONCLUSIONS: Using the AXB(m) algorithm and carefully reviewing the beam calibration procedure in H&N FFF-VMAT treatments ensures (1) dose accuracy increase by approximately 3%; (2) a consequent dose increase in targets; and (3) a dose reporting mode that is consistent with the trend of current algorithms.
format Online
Article
Text
id pubmed-8086705
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Via Medica
record_format MEDLINE/PubMed
spelling pubmed-80867052021-05-03 Dose accuracy improvement on head and neck VMAT treatments by using the Acuros algorithm and accurate FFF beam calibration Martin-Martin, Guadalupe Walter, Stefan Guibelalde, Eduardo Rep Pract Oncol Radiother Research Paper BACKGROUND: The purpose of this study was to assess dose accuracy improvement and dosimetric impact of switching from the anisotropic analytical algorithm (AA) to the Acuros XB algorithm (AXB) when performing an accurate beam calibration in head and neck (H&N) FFF-VMAT treatments. MATERIALS AND METHODS: Twenty H&N cancer patients treated with FFF-VMAT techniques were included. Calculations were performed with the AA and AXB algorithm (dose-to-water — AXB(w)– and dose-to-medium — AXB(m)–). An accurate beam calibration was used for AXB calculations. Dose prescription to the tumour (PTV70) and at-risk-nodal region (PTV58.1) were 70 Gy and 58.1 Gy, respectively. A PTV70(_bone) including bony structures in PTV70 was contoured. Dose-volume parameters were compared between the algorithms. Statistical tests were used to analyze the differences in mean values and the correlation between compliance with the D95 > 95% requirement and occurrence of local recurrence. RESULTS: AA systematically overestimated the dose compared to AXB algorithm with mean dose differences within 1.3 Gy/2%, except for the PTV70(_bone) (2.2 Gy/3.2%). Dose differences were significantly higher for AXB(m) calculations when including accurate beam calibration (maximum dose differences up to 2.8 Gy/4.1% and 4.2 Gy/6.3% for PTV70 and PTV70(_bone), respectively). 80% of AA-calculated plans did not meet the D95 > 95% requirement after recalculation with AXB(m) and accurate beam calibration. The reduction in D95 coverage in the tumour was not clinically relevant. CONCLUSIONS: Using the AXB(m) algorithm and carefully reviewing the beam calibration procedure in H&N FFF-VMAT treatments ensures (1) dose accuracy increase by approximately 3%; (2) a consequent dose increase in targets; and (3) a dose reporting mode that is consistent with the trend of current algorithms. Via Medica 2021-02-25 /pmc/articles/PMC8086705/ /pubmed/33948305 http://dx.doi.org/10.5603/RPOR.a2021.0014 Text en © 2021 Greater Poland Cancer Centre https://creativecommons.org/licenses/by-nc-nd/4.0/This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially
spellingShingle Research Paper
Martin-Martin, Guadalupe
Walter, Stefan
Guibelalde, Eduardo
Dose accuracy improvement on head and neck VMAT treatments by using the Acuros algorithm and accurate FFF beam calibration
title Dose accuracy improvement on head and neck VMAT treatments by using the Acuros algorithm and accurate FFF beam calibration
title_full Dose accuracy improvement on head and neck VMAT treatments by using the Acuros algorithm and accurate FFF beam calibration
title_fullStr Dose accuracy improvement on head and neck VMAT treatments by using the Acuros algorithm and accurate FFF beam calibration
title_full_unstemmed Dose accuracy improvement on head and neck VMAT treatments by using the Acuros algorithm and accurate FFF beam calibration
title_short Dose accuracy improvement on head and neck VMAT treatments by using the Acuros algorithm and accurate FFF beam calibration
title_sort dose accuracy improvement on head and neck vmat treatments by using the acuros algorithm and accurate fff beam calibration
topic Research Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8086705/
https://www.ncbi.nlm.nih.gov/pubmed/33948305
http://dx.doi.org/10.5603/RPOR.a2021.0014
work_keys_str_mv AT martinmartinguadalupe doseaccuracyimprovementonheadandneckvmattreatmentsbyusingtheacurosalgorithmandaccuratefffbeamcalibration
AT walterstefan doseaccuracyimprovementonheadandneckvmattreatmentsbyusingtheacurosalgorithmandaccuratefffbeamcalibration
AT guibelaldeeduardo doseaccuracyimprovementonheadandneckvmattreatmentsbyusingtheacurosalgorithmandaccuratefffbeamcalibration