Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study
BACKGROUND: Accurate, continuous heart rate measurements are important for health assessment, physical activity, and sporting performance, and the integration of heart rate measurements into wearable devices has extended its accessibility. Although the use of photoplethysmography technology is not n...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
JMIR Publications
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8088863/ https://www.ncbi.nlm.nih.gov/pubmed/33764310 http://dx.doi.org/10.2196/25313 |
_version_ | 1783686925884850176 |
---|---|
author | Muggeridge, David Joseph Hickson, Kirsty Davies, Aimie Victoria Giggins, Oonagh M Megson, Ian L Gorely, Trish Crabtree, Daniel R |
author_facet | Muggeridge, David Joseph Hickson, Kirsty Davies, Aimie Victoria Giggins, Oonagh M Megson, Ian L Gorely, Trish Crabtree, Daniel R |
author_sort | Muggeridge, David Joseph |
collection | PubMed |
description | BACKGROUND: Accurate, continuous heart rate measurements are important for health assessment, physical activity, and sporting performance, and the integration of heart rate measurements into wearable devices has extended its accessibility. Although the use of photoplethysmography technology is not new, the available data relating to the validity of measurement are limited, and the range of activities being performed is often restricted to one exercise domain and/or limited intensities. OBJECTIVE: The primary objective of this study was to assess the validity of the Polar OH1 and Fitbit Charge 3 devices for measuring heart rate during rest, light, moderate, vigorous, and sprint-type exercise. METHODS: A total of 20 healthy adults (9 female; height: mean 1.73 [SD 0.1] m; body mass: mean 71.6 [SD 11.0] kg; and age: mean 40 [SD 10] years) volunteered and provided written informed consent to participate in the study consisting of 2 trials. Trial 1 was split into 3 components: 15-minute sedentary activities, 10-minute cycling on a bicycle ergometer, and incremental exercise test to exhaustion on a motorized treadmill (18-42 minutes). Trial 2 was split into 2 components: 4 × 15-second maximal sprints on a cycle ergometer and 4 × 30- to 50-m sprints on a nonmotorized resistance treadmill. Data from the 3 devices were time-aligned, and the validity of Polar OH1 and Fitbit Charge 3 was assessed against Polar H10 (criterion device). Validity was evaluated using the Bland and Altman analysis, Pearson moment correlation coefficient, and mean absolute percentage error. RESULTS: Overall, there was a very good correlation between the Polar OH1 and Polar H10 devices (r=0.95), with a mean bias of −1 beats·min(-1) and limits of agreement of −20 to 19 beats·min(-1). The Fitbit Charge 3 device underestimated heart rate by 7 beats·min(-1) compared with Polar H10, with a limit of agreement of −46 to 33 beats·min(-1) and poor correlation (r=0.8). The mean absolute percentage error for both devices was deemed acceptable (<5%). Polar OH1 performed well across each phase of trial 1; however, validity was worse for trial 2 activities. Fitbit Charge 3 performed well only during rest and nonsprint-based treadmill activities. CONCLUSIONS: Compared with our criterion device, Polar OH1 was accurate at assessing heart rate, but the accuracy of Fitbit Charge 3 was generally poor. Polar OH1 performed worse during trial 2 compared with the activities in trial 1, and the validity of the Fitbit Charge 3 device was particularly poor during our cycling exercises. |
format | Online Article Text |
id | pubmed-8088863 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | JMIR Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-80888632021-05-07 Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study Muggeridge, David Joseph Hickson, Kirsty Davies, Aimie Victoria Giggins, Oonagh M Megson, Ian L Gorely, Trish Crabtree, Daniel R JMIR Mhealth Uhealth Original Paper BACKGROUND: Accurate, continuous heart rate measurements are important for health assessment, physical activity, and sporting performance, and the integration of heart rate measurements into wearable devices has extended its accessibility. Although the use of photoplethysmography technology is not new, the available data relating to the validity of measurement are limited, and the range of activities being performed is often restricted to one exercise domain and/or limited intensities. OBJECTIVE: The primary objective of this study was to assess the validity of the Polar OH1 and Fitbit Charge 3 devices for measuring heart rate during rest, light, moderate, vigorous, and sprint-type exercise. METHODS: A total of 20 healthy adults (9 female; height: mean 1.73 [SD 0.1] m; body mass: mean 71.6 [SD 11.0] kg; and age: mean 40 [SD 10] years) volunteered and provided written informed consent to participate in the study consisting of 2 trials. Trial 1 was split into 3 components: 15-minute sedentary activities, 10-minute cycling on a bicycle ergometer, and incremental exercise test to exhaustion on a motorized treadmill (18-42 minutes). Trial 2 was split into 2 components: 4 × 15-second maximal sprints on a cycle ergometer and 4 × 30- to 50-m sprints on a nonmotorized resistance treadmill. Data from the 3 devices were time-aligned, and the validity of Polar OH1 and Fitbit Charge 3 was assessed against Polar H10 (criterion device). Validity was evaluated using the Bland and Altman analysis, Pearson moment correlation coefficient, and mean absolute percentage error. RESULTS: Overall, there was a very good correlation between the Polar OH1 and Polar H10 devices (r=0.95), with a mean bias of −1 beats·min(-1) and limits of agreement of −20 to 19 beats·min(-1). The Fitbit Charge 3 device underestimated heart rate by 7 beats·min(-1) compared with Polar H10, with a limit of agreement of −46 to 33 beats·min(-1) and poor correlation (r=0.8). The mean absolute percentage error for both devices was deemed acceptable (<5%). Polar OH1 performed well across each phase of trial 1; however, validity was worse for trial 2 activities. Fitbit Charge 3 performed well only during rest and nonsprint-based treadmill activities. CONCLUSIONS: Compared with our criterion device, Polar OH1 was accurate at assessing heart rate, but the accuracy of Fitbit Charge 3 was generally poor. Polar OH1 performed worse during trial 2 compared with the activities in trial 1, and the validity of the Fitbit Charge 3 device was particularly poor during our cycling exercises. JMIR Publications 2021-03-25 /pmc/articles/PMC8088863/ /pubmed/33764310 http://dx.doi.org/10.2196/25313 Text en ©David Joseph Muggeridge, Kirsty Hickson, Aimie Victoria Davies, Oonagh M Giggins, Ian L Megson, Trish Gorely, Daniel R Crabtree. Originally published in JMIR mHealth and uHealth (http://mhealth.jmir.org), 25.03.2021. https://creativecommons.org/licenses/by/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mHealth and uHealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/, as well as this copyright and license information must be included. |
spellingShingle | Original Paper Muggeridge, David Joseph Hickson, Kirsty Davies, Aimie Victoria Giggins, Oonagh M Megson, Ian L Gorely, Trish Crabtree, Daniel R Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study |
title | Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study |
title_full | Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study |
title_fullStr | Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study |
title_full_unstemmed | Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study |
title_short | Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study |
title_sort | measurement of heart rate using the polar oh1 and fitbit charge 3 wearable devices in healthy adults during light, moderate, vigorous, and sprint-based exercise: validation study |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8088863/ https://www.ncbi.nlm.nih.gov/pubmed/33764310 http://dx.doi.org/10.2196/25313 |
work_keys_str_mv | AT muggeridgedavidjoseph measurementofheartrateusingthepolaroh1andfitbitcharge3wearabledevicesinhealthyadultsduringlightmoderatevigorousandsprintbasedexercisevalidationstudy AT hicksonkirsty measurementofheartrateusingthepolaroh1andfitbitcharge3wearabledevicesinhealthyadultsduringlightmoderatevigorousandsprintbasedexercisevalidationstudy AT daviesaimievictoria measurementofheartrateusingthepolaroh1andfitbitcharge3wearabledevicesinhealthyadultsduringlightmoderatevigorousandsprintbasedexercisevalidationstudy AT gigginsoonaghm measurementofheartrateusingthepolaroh1andfitbitcharge3wearabledevicesinhealthyadultsduringlightmoderatevigorousandsprintbasedexercisevalidationstudy AT megsonianl measurementofheartrateusingthepolaroh1andfitbitcharge3wearabledevicesinhealthyadultsduringlightmoderatevigorousandsprintbasedexercisevalidationstudy AT gorelytrish measurementofheartrateusingthepolaroh1andfitbitcharge3wearabledevicesinhealthyadultsduringlightmoderatevigorousandsprintbasedexercisevalidationstudy AT crabtreedanielr measurementofheartrateusingthepolaroh1andfitbitcharge3wearabledevicesinhealthyadultsduringlightmoderatevigorousandsprintbasedexercisevalidationstudy |