Prospective blinded evaluation of smartphone-based ECG for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia

INTRODUCTION: Supraventricular tachycardias (SVT) are often difficult to document due to their intermittent, short-lasting nature. Smartphone-based one-lead ECG monitors (sECG) were initially developed for the diagnosis of atrial fibrillation. No data have been published regarding their potential ro...

Descripción completa

Detalles Bibliográficos
Autores principales: Wegner, Felix K., Kochhäuser, Simon, Frommeyer, Gerrit, Lange, Philipp S., Ellermann, Christian, Leitz, Patrick, Müller, Patrick, Köbe, Julia, Eckardt, Lars, Dechering, Dirk G.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8103426/
https://www.ncbi.nlm.nih.gov/pubmed/33961097
http://dx.doi.org/10.1007/s00392-021-01856-5
_version_ 1783689308635398144
author Wegner, Felix K.
Kochhäuser, Simon
Frommeyer, Gerrit
Lange, Philipp S.
Ellermann, Christian
Leitz, Patrick
Müller, Patrick
Köbe, Julia
Eckardt, Lars
Dechering, Dirk G.
author_facet Wegner, Felix K.
Kochhäuser, Simon
Frommeyer, Gerrit
Lange, Philipp S.
Ellermann, Christian
Leitz, Patrick
Müller, Patrick
Köbe, Julia
Eckardt, Lars
Dechering, Dirk G.
author_sort Wegner, Felix K.
collection PubMed
description INTRODUCTION: Supraventricular tachycardias (SVT) are often difficult to document due to their intermittent, short-lasting nature. Smartphone-based one-lead ECG monitors (sECG) were initially developed for the diagnosis of atrial fibrillation. No data have been published regarding their potential role in differentiating inappropiate sinus tachycardia (IST) from regular SVT. If cardiologists could distinguish IST from SVT in sECG, economic health care burden might be significantly reduced. METHODS: We prospectively recruited 75 consecutive patients with known SVT undergoing an EP study. In all patients, four ECG were recorded: a sECG during SVT and during sinus tachycardia and respective 12-lead ECG. Two experienced electrophysiologists were blinded to the diagnoses and separately evaluated all ECG. RESULTS: Three hundred individual ECG were recorded in 75 patients (47 female, age 50 ± 18 years, BMI 26 ± 5 kg/m(2), 60 AVNRT, 15 AVRT). The electrophysiologists’ blinded interpretation of sECG recordings showed a sensitivity of 89% and a specificity of 91% for the detection of SVT (interobserver agreement κ = 0.76). In high-quality sECG recordings (68%), sensitivity rose to 95% with a specificity of 92% (interobserver agreement of κ = 0.91). Specificity increased to 96% when both electrophysiologists agreed on the diagnosis. Respective 12-lead ECG had a sensitivity of 100% and specificity of 98% for the detection of SVT. CONCLUSION: A smartphone-based one-lead ECG monitor allows for differentiation of SVT from IST in about 90% of cases. These results should encourage cardiologists to integrate wearables into clinical practice, possibly reducing time to definitive diagnosis of an arrhythmia and unnecessary EP procedures. GRAPHICAL ABSTRACT: A smartphone-based one lead ECG device (panel A) can be used reliably to differentiate supraventricular tachycardia (panel B) from inappropriate sinus tachycardia when compared to a simultaneously conducted gold-standard electrophysiology study (panels C, D). [Image: see text]
format Online
Article
Text
id pubmed-8103426
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-81034262021-05-07 Prospective blinded evaluation of smartphone-based ECG for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia Wegner, Felix K. Kochhäuser, Simon Frommeyer, Gerrit Lange, Philipp S. Ellermann, Christian Leitz, Patrick Müller, Patrick Köbe, Julia Eckardt, Lars Dechering, Dirk G. Clin Res Cardiol Original Paper INTRODUCTION: Supraventricular tachycardias (SVT) are often difficult to document due to their intermittent, short-lasting nature. Smartphone-based one-lead ECG monitors (sECG) were initially developed for the diagnosis of atrial fibrillation. No data have been published regarding their potential role in differentiating inappropiate sinus tachycardia (IST) from regular SVT. If cardiologists could distinguish IST from SVT in sECG, economic health care burden might be significantly reduced. METHODS: We prospectively recruited 75 consecutive patients with known SVT undergoing an EP study. In all patients, four ECG were recorded: a sECG during SVT and during sinus tachycardia and respective 12-lead ECG. Two experienced electrophysiologists were blinded to the diagnoses and separately evaluated all ECG. RESULTS: Three hundred individual ECG were recorded in 75 patients (47 female, age 50 ± 18 years, BMI 26 ± 5 kg/m(2), 60 AVNRT, 15 AVRT). The electrophysiologists’ blinded interpretation of sECG recordings showed a sensitivity of 89% and a specificity of 91% for the detection of SVT (interobserver agreement κ = 0.76). In high-quality sECG recordings (68%), sensitivity rose to 95% with a specificity of 92% (interobserver agreement of κ = 0.91). Specificity increased to 96% when both electrophysiologists agreed on the diagnosis. Respective 12-lead ECG had a sensitivity of 100% and specificity of 98% for the detection of SVT. CONCLUSION: A smartphone-based one-lead ECG monitor allows for differentiation of SVT from IST in about 90% of cases. These results should encourage cardiologists to integrate wearables into clinical practice, possibly reducing time to definitive diagnosis of an arrhythmia and unnecessary EP procedures. GRAPHICAL ABSTRACT: A smartphone-based one lead ECG device (panel A) can be used reliably to differentiate supraventricular tachycardia (panel B) from inappropriate sinus tachycardia when compared to a simultaneously conducted gold-standard electrophysiology study (panels C, D). [Image: see text] Springer Berlin Heidelberg 2021-05-07 2021 /pmc/articles/PMC8103426/ /pubmed/33961097 http://dx.doi.org/10.1007/s00392-021-01856-5 Text en © The Author(s) 2021, corrected publication 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Paper
Wegner, Felix K.
Kochhäuser, Simon
Frommeyer, Gerrit
Lange, Philipp S.
Ellermann, Christian
Leitz, Patrick
Müller, Patrick
Köbe, Julia
Eckardt, Lars
Dechering, Dirk G.
Prospective blinded evaluation of smartphone-based ECG for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia
title Prospective blinded evaluation of smartphone-based ECG for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia
title_full Prospective blinded evaluation of smartphone-based ECG for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia
title_fullStr Prospective blinded evaluation of smartphone-based ECG for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia
title_full_unstemmed Prospective blinded evaluation of smartphone-based ECG for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia
title_short Prospective blinded evaluation of smartphone-based ECG for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia
title_sort prospective blinded evaluation of smartphone-based ecg for differentiation of supraventricular tachycardia from inappropriate sinus tachycardia
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8103426/
https://www.ncbi.nlm.nih.gov/pubmed/33961097
http://dx.doi.org/10.1007/s00392-021-01856-5
work_keys_str_mv AT wegnerfelixk prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia
AT kochhausersimon prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia
AT frommeyergerrit prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia
AT langephilipps prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia
AT ellermannchristian prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia
AT leitzpatrick prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia
AT mullerpatrick prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia
AT kobejulia prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia
AT eckardtlars prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia
AT decheringdirkg prospectiveblindedevaluationofsmartphonebasedecgfordifferentiationofsupraventriculartachycardiafrominappropriatesinustachycardia