Dosimetric Evaluation of the QFix kVue(TM) Calypso Couch Top

PURPOSE: To evaluate the dosimetric accuracy of the default couch model of the QFix kVue(TM) Calypso couch top in the treatment planning system. METHODS: With the gantry 180°, field size 20 × 20 cm, 6 MV, we measured the depth dose, off-axis dose, and dose plane of different depths in the phantom wi...

Descripción completa

Detalles Bibliográficos
Autores principales: Hou, Lingtong, Zhang, Huiqin, Sun, Xiaomei, Liu, Qianqian, Chen, Tingfeng, Liu, Yong, Jiang, Xiaodong, Yao, Shengyu
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8107663/
https://www.ncbi.nlm.nih.gov/pubmed/33910440
http://dx.doi.org/10.1177/15330338211011964
_version_ 1783689990108086272
author Hou, Lingtong
Zhang, Huiqin
Sun, Xiaomei
Liu, Qianqian
Chen, Tingfeng
Liu, Yong
Jiang, Xiaodong
Yao, Shengyu
author_facet Hou, Lingtong
Zhang, Huiqin
Sun, Xiaomei
Liu, Qianqian
Chen, Tingfeng
Liu, Yong
Jiang, Xiaodong
Yao, Shengyu
author_sort Hou, Lingtong
collection PubMed
description PURPOSE: To evaluate the dosimetric accuracy of the default couch model of the QFix kVue(TM) Calypso couch top in the treatment planning system. METHODS: With the gantry 180°, field size 20 × 20 cm, 6 MV, we measured the depth dose, off-axis dose, and dose plane of different depths in the phantom with the couch rails in and out, respectively. Isocenter doses at different angles were also obtained. The results were compared to the doses calculated using the default couch top model and the real scanned couch top model. Then we revised the default model according to the measured results. RESULTS: With “Rails In,” the depth dose, off-axis dose, and dose plane of the default couch top model had a big difference with the dose of the real scanned couch top model and the measured result. The dose of the real scanned couch top model was much closer to the measured result, but in the region of the rail edge, the difference was still significant. With “Rails Out,” there was a minor difference between the measured result, the dose of the default couch top model and the real scanned couch top model. The difference between the measurement and the default couch top model became very small after being revised. CONCLUSIONS: It is better to avoid the beam angle passing through the couch rails in treatment plans, or you should revise the parameter of the QFix kVue(TM) Calypso couch top model based on the measured results, and verify the treatment plan before clinical practice.
format Online
Article
Text
id pubmed-8107663
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-81076632021-05-14 Dosimetric Evaluation of the QFix kVue(TM) Calypso Couch Top Hou, Lingtong Zhang, Huiqin Sun, Xiaomei Liu, Qianqian Chen, Tingfeng Liu, Yong Jiang, Xiaodong Yao, Shengyu Technol Cancer Res Treat Original Article PURPOSE: To evaluate the dosimetric accuracy of the default couch model of the QFix kVue(TM) Calypso couch top in the treatment planning system. METHODS: With the gantry 180°, field size 20 × 20 cm, 6 MV, we measured the depth dose, off-axis dose, and dose plane of different depths in the phantom with the couch rails in and out, respectively. Isocenter doses at different angles were also obtained. The results were compared to the doses calculated using the default couch top model and the real scanned couch top model. Then we revised the default model according to the measured results. RESULTS: With “Rails In,” the depth dose, off-axis dose, and dose plane of the default couch top model had a big difference with the dose of the real scanned couch top model and the measured result. The dose of the real scanned couch top model was much closer to the measured result, but in the region of the rail edge, the difference was still significant. With “Rails Out,” there was a minor difference between the measured result, the dose of the default couch top model and the real scanned couch top model. The difference between the measurement and the default couch top model became very small after being revised. CONCLUSIONS: It is better to avoid the beam angle passing through the couch rails in treatment plans, or you should revise the parameter of the QFix kVue(TM) Calypso couch top model based on the measured results, and verify the treatment plan before clinical practice. SAGE Publications 2021-04-29 /pmc/articles/PMC8107663/ /pubmed/33910440 http://dx.doi.org/10.1177/15330338211011964 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Article
Hou, Lingtong
Zhang, Huiqin
Sun, Xiaomei
Liu, Qianqian
Chen, Tingfeng
Liu, Yong
Jiang, Xiaodong
Yao, Shengyu
Dosimetric Evaluation of the QFix kVue(TM) Calypso Couch Top
title Dosimetric Evaluation of the QFix kVue(TM) Calypso Couch Top
title_full Dosimetric Evaluation of the QFix kVue(TM) Calypso Couch Top
title_fullStr Dosimetric Evaluation of the QFix kVue(TM) Calypso Couch Top
title_full_unstemmed Dosimetric Evaluation of the QFix kVue(TM) Calypso Couch Top
title_short Dosimetric Evaluation of the QFix kVue(TM) Calypso Couch Top
title_sort dosimetric evaluation of the qfix kvue(tm) calypso couch top
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8107663/
https://www.ncbi.nlm.nih.gov/pubmed/33910440
http://dx.doi.org/10.1177/15330338211011964
work_keys_str_mv AT houlingtong dosimetricevaluationoftheqfixkvuetmcalypsocouchtop
AT zhanghuiqin dosimetricevaluationoftheqfixkvuetmcalypsocouchtop
AT sunxiaomei dosimetricevaluationoftheqfixkvuetmcalypsocouchtop
AT liuqianqian dosimetricevaluationoftheqfixkvuetmcalypsocouchtop
AT chentingfeng dosimetricevaluationoftheqfixkvuetmcalypsocouchtop
AT liuyong dosimetricevaluationoftheqfixkvuetmcalypsocouchtop
AT jiangxiaodong dosimetricevaluationoftheqfixkvuetmcalypsocouchtop
AT yaoshengyu dosimetricevaluationoftheqfixkvuetmcalypsocouchtop