Comparison of Actual Performance in the Flow and Fraction of Inspired O(2) among Different High-Flow Nasal Cannula Devices: A Bench Study

BACKGROUND: High-flow nasal cannula (HFNC) oxygen therapy has been recommended for use in coronavirus disease 2019 (COVID-19) patients with acute respiratory failure and many other clinical conditions. HFNC devices produced by different manufacturers may have varied performance. Whether there is a d...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhou, Yuyan, Ni, Zhong, Ni, Yuenan, Liang, Binmiao, Liang, Zongan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8112956/
https://www.ncbi.nlm.nih.gov/pubmed/34055113
http://dx.doi.org/10.1155/2021/6638048
_version_ 1783690772552351744
author Zhou, Yuyan
Ni, Zhong
Ni, Yuenan
Liang, Binmiao
Liang, Zongan
author_facet Zhou, Yuyan
Ni, Zhong
Ni, Yuenan
Liang, Binmiao
Liang, Zongan
author_sort Zhou, Yuyan
collection PubMed
description BACKGROUND: High-flow nasal cannula (HFNC) oxygen therapy has been recommended for use in coronavirus disease 2019 (COVID-19) patients with acute respiratory failure and many other clinical conditions. HFNC devices produced by different manufacturers may have varied performance. Whether there is a difference in these devices and the extent of the differences in performance remain unknown. METHODS: Four HFNC devices (AIRVO 2, TNI softFlow 50, HUMID-BH, and OH-70C) and a ventilator with an HFNC module (bellavista 1000) were evaluated. The flow was set at 20, 25, 30, 35, 40, 45, 50, 60, 70, and 80 L/min, and the FiO(2) was set at 21%, 26%, 30%, 35%, 40%, 45%, 50%, 60%, 70%, 80%, and 90%. Then, one side of the cannulas was clipped to simulate the compression, bending, or blocking of the nasal cannulas. The flow and FiO(2) of the delivered gas were recorded and compared among settings and devices. RESULTS: The actual-flow and actual-FiO(2) delivered by different settings and devices varied. AIRVO 2 had superior performance in flow and FiO(2) accuracy. bellavista 1000 and OH-70C had good performance in the accuracy of actual-flows and actual-FiO(2,) respectively. bellavista 1000 and HUMID-BH had a larger flow range from 10 to 80 L/min, but only bellavista 1000 could provide a stable flow with an excessive resistance up to 60 L/min. TNI softFlow 50 had the best flow compensation and could provide sufficient flow with excessive resistance at 20–50 L/min. CONCLUSIONS: The variation in flow, FiO(2) settings, and devices could influence the actual-flow and actual-FiO(2) delivered. AIRVO 2 and OH-70C showed better FiO(2) accuracy. TNI softFlow 50, bellavista 1000, and HUMID-BH could lower the risk of insufficient flow support due to accidental compression or blocking of the cannulas. In addition, ventilators with HFNC modules provided comparable flow and FiO(2) and could be an alternative to standalone HFNC devices.
format Online
Article
Text
id pubmed-8112956
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-81129562021-05-27 Comparison of Actual Performance in the Flow and Fraction of Inspired O(2) among Different High-Flow Nasal Cannula Devices: A Bench Study Zhou, Yuyan Ni, Zhong Ni, Yuenan Liang, Binmiao Liang, Zongan Can Respir J Research Article BACKGROUND: High-flow nasal cannula (HFNC) oxygen therapy has been recommended for use in coronavirus disease 2019 (COVID-19) patients with acute respiratory failure and many other clinical conditions. HFNC devices produced by different manufacturers may have varied performance. Whether there is a difference in these devices and the extent of the differences in performance remain unknown. METHODS: Four HFNC devices (AIRVO 2, TNI softFlow 50, HUMID-BH, and OH-70C) and a ventilator with an HFNC module (bellavista 1000) were evaluated. The flow was set at 20, 25, 30, 35, 40, 45, 50, 60, 70, and 80 L/min, and the FiO(2) was set at 21%, 26%, 30%, 35%, 40%, 45%, 50%, 60%, 70%, 80%, and 90%. Then, one side of the cannulas was clipped to simulate the compression, bending, or blocking of the nasal cannulas. The flow and FiO(2) of the delivered gas were recorded and compared among settings and devices. RESULTS: The actual-flow and actual-FiO(2) delivered by different settings and devices varied. AIRVO 2 had superior performance in flow and FiO(2) accuracy. bellavista 1000 and OH-70C had good performance in the accuracy of actual-flows and actual-FiO(2,) respectively. bellavista 1000 and HUMID-BH had a larger flow range from 10 to 80 L/min, but only bellavista 1000 could provide a stable flow with an excessive resistance up to 60 L/min. TNI softFlow 50 had the best flow compensation and could provide sufficient flow with excessive resistance at 20–50 L/min. CONCLUSIONS: The variation in flow, FiO(2) settings, and devices could influence the actual-flow and actual-FiO(2) delivered. AIRVO 2 and OH-70C showed better FiO(2) accuracy. TNI softFlow 50, bellavista 1000, and HUMID-BH could lower the risk of insufficient flow support due to accidental compression or blocking of the cannulas. In addition, ventilators with HFNC modules provided comparable flow and FiO(2) and could be an alternative to standalone HFNC devices. Hindawi 2021-05-04 /pmc/articles/PMC8112956/ /pubmed/34055113 http://dx.doi.org/10.1155/2021/6638048 Text en Copyright © 2021 Yuyan Zhou et al. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Zhou, Yuyan
Ni, Zhong
Ni, Yuenan
Liang, Binmiao
Liang, Zongan
Comparison of Actual Performance in the Flow and Fraction of Inspired O(2) among Different High-Flow Nasal Cannula Devices: A Bench Study
title Comparison of Actual Performance in the Flow and Fraction of Inspired O(2) among Different High-Flow Nasal Cannula Devices: A Bench Study
title_full Comparison of Actual Performance in the Flow and Fraction of Inspired O(2) among Different High-Flow Nasal Cannula Devices: A Bench Study
title_fullStr Comparison of Actual Performance in the Flow and Fraction of Inspired O(2) among Different High-Flow Nasal Cannula Devices: A Bench Study
title_full_unstemmed Comparison of Actual Performance in the Flow and Fraction of Inspired O(2) among Different High-Flow Nasal Cannula Devices: A Bench Study
title_short Comparison of Actual Performance in the Flow and Fraction of Inspired O(2) among Different High-Flow Nasal Cannula Devices: A Bench Study
title_sort comparison of actual performance in the flow and fraction of inspired o(2) among different high-flow nasal cannula devices: a bench study
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8112956/
https://www.ncbi.nlm.nih.gov/pubmed/34055113
http://dx.doi.org/10.1155/2021/6638048
work_keys_str_mv AT zhouyuyan comparisonofactualperformanceintheflowandfractionofinspiredo2amongdifferenthighflownasalcannuladevicesabenchstudy
AT nizhong comparisonofactualperformanceintheflowandfractionofinspiredo2amongdifferenthighflownasalcannuladevicesabenchstudy
AT niyuenan comparisonofactualperformanceintheflowandfractionofinspiredo2amongdifferenthighflownasalcannuladevicesabenchstudy
AT liangbinmiao comparisonofactualperformanceintheflowandfractionofinspiredo2amongdifferenthighflownasalcannuladevicesabenchstudy
AT liangzongan comparisonofactualperformanceintheflowandfractionofinspiredo2amongdifferenthighflownasalcannuladevicesabenchstudy