Cargando…

Unexpected inhibition of bradycardia pacing due to oversensing in ICD lead fracture associated with spurious tachyarrhythmia detection and discharges

Ιn a 76-year old man with a dual-chamber ICD implanted five years ago, dizzy spells and significant bradycardia on Holter were not initially recognized as inhibition of bradycardia pacing, due to oversensing. Hospital admission was deemed necessary only after repetitive ICD shocks attributed to righ...

Descripción completa

Detalles Bibliográficos
Autores principales: Zagkli, Fani, Chronopoulos, Panagiotis, Chiladakis, John
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8116807/
https://www.ncbi.nlm.nih.gov/pubmed/33667573
http://dx.doi.org/10.1016/j.ipej.2021.02.010
Descripción
Sumario:Ιn a 76-year old man with a dual-chamber ICD implanted five years ago, dizzy spells and significant bradycardia on Holter were not initially recognized as inhibition of bradycardia pacing, due to oversensing. Hospital admission was deemed necessary only after repetitive ICD shocks attributed to right ventricular pace-sense lead fracture. The need to ensure adequate ICD antibradycardia backup pacing in pacing-dependent patients when deleterious sensing errors occur, cannot be overemphasized.