Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci
Success of total knee replacement (TKR) depends on the prosthetic design and materials. The use of metal components is well established with the disadvantage of allergic reactions. Ceramics have been recently proposed because of high wear resistance, excellent biocompatibility, wettability, and suit...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8122674/ https://www.ncbi.nlm.nih.gov/pubmed/33921972 http://dx.doi.org/10.3390/ma14092112 |
_version_ | 1783692682082648064 |
---|---|
author | Affatato, Saverio Erani, Paolo Fersini, Maurizio Contaldi, Vincenzo Terrizzi, Anna Rita Licciulli, Antonio |
author_facet | Affatato, Saverio Erani, Paolo Fersini, Maurizio Contaldi, Vincenzo Terrizzi, Anna Rita Licciulli, Antonio |
author_sort | Affatato, Saverio |
collection | PubMed |
description | Success of total knee replacement (TKR) depends on the prosthetic design and materials. The use of metal components is well established with the disadvantage of allergic reactions. Ceramics have been recently proposed because of high wear resistance, excellent biocompatibility, wettability, and suitable mechanical properties. This study was aimed at investigating in vitro wear resistance of Zirconia Toughened Alumina (ZTA)/Ultra-high-molecular-weight polyethylene (UHMWPE) of TKR femoral components. An in vitro protocol was designed with the application of relevant load profile, 6-degrees-of-freedom knee simulator, and 8 × 10(5) cycles on the ZTA/UHMWPE configuration under bovine calf serum. Before and after wear test, the femoral components were investigated by using the Scanning Electron Microscope (SEM) and the X-Ray Diffraction (XRD) analyses, and stylus surface roughness measurements. The proposed pre-clinical test yielded repeatable results. In particular, gravimetric results showed that, after 8 × 10(5) cycles, the mean weight loss of the polyethylene mobile components is 5.3 ± 1.1 mg. The surface roughness measurements (Ra(max)) performed after the wear test showed no significant variation on the UHMWPE menisci. A slight increase of roughness has been found on the ZTA (0.02 µm before wear test, 0.28 µm after the test). SEM observations did not show significant modification of the surface morphology. Tetragonal to monoclinic phase ratio was measured by XRD before and after wear test to evaluate stability of tetragonal ZrO(2) phase. Minimal conversion of tetragonal to monoclinic phase was found from 5.4 to 8%. Although this study is a preliminary evaluation limited to in vitro tests, it provides novel pre-clinical indications about the potential of ceramic TKR femoral components. |
format | Online Article Text |
id | pubmed-8122674 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-81226742021-05-16 Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci Affatato, Saverio Erani, Paolo Fersini, Maurizio Contaldi, Vincenzo Terrizzi, Anna Rita Licciulli, Antonio Materials (Basel) Article Success of total knee replacement (TKR) depends on the prosthetic design and materials. The use of metal components is well established with the disadvantage of allergic reactions. Ceramics have been recently proposed because of high wear resistance, excellent biocompatibility, wettability, and suitable mechanical properties. This study was aimed at investigating in vitro wear resistance of Zirconia Toughened Alumina (ZTA)/Ultra-high-molecular-weight polyethylene (UHMWPE) of TKR femoral components. An in vitro protocol was designed with the application of relevant load profile, 6-degrees-of-freedom knee simulator, and 8 × 10(5) cycles on the ZTA/UHMWPE configuration under bovine calf serum. Before and after wear test, the femoral components were investigated by using the Scanning Electron Microscope (SEM) and the X-Ray Diffraction (XRD) analyses, and stylus surface roughness measurements. The proposed pre-clinical test yielded repeatable results. In particular, gravimetric results showed that, after 8 × 10(5) cycles, the mean weight loss of the polyethylene mobile components is 5.3 ± 1.1 mg. The surface roughness measurements (Ra(max)) performed after the wear test showed no significant variation on the UHMWPE menisci. A slight increase of roughness has been found on the ZTA (0.02 µm before wear test, 0.28 µm after the test). SEM observations did not show significant modification of the surface morphology. Tetragonal to monoclinic phase ratio was measured by XRD before and after wear test to evaluate stability of tetragonal ZrO(2) phase. Minimal conversion of tetragonal to monoclinic phase was found from 5.4 to 8%. Although this study is a preliminary evaluation limited to in vitro tests, it provides novel pre-clinical indications about the potential of ceramic TKR femoral components. MDPI 2021-04-22 /pmc/articles/PMC8122674/ /pubmed/33921972 http://dx.doi.org/10.3390/ma14092112 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Affatato, Saverio Erani, Paolo Fersini, Maurizio Contaldi, Vincenzo Terrizzi, Anna Rita Licciulli, Antonio Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci |
title | Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci |
title_full | Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci |
title_fullStr | Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci |
title_full_unstemmed | Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci |
title_short | Preliminary In Vitro Wear Assessment of Ceramic Cemented Femoral Components Coupled with Polyethylene Menisci |
title_sort | preliminary in vitro wear assessment of ceramic cemented femoral components coupled with polyethylene menisci |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8122674/ https://www.ncbi.nlm.nih.gov/pubmed/33921972 http://dx.doi.org/10.3390/ma14092112 |
work_keys_str_mv | AT affatatosaverio preliminaryinvitrowearassessmentofceramiccementedfemoralcomponentscoupledwithpolyethylenemenisci AT eranipaolo preliminaryinvitrowearassessmentofceramiccementedfemoralcomponentscoupledwithpolyethylenemenisci AT fersinimaurizio preliminaryinvitrowearassessmentofceramiccementedfemoralcomponentscoupledwithpolyethylenemenisci AT contaldivincenzo preliminaryinvitrowearassessmentofceramiccementedfemoralcomponentscoupledwithpolyethylenemenisci AT terrizziannarita preliminaryinvitrowearassessmentofceramiccementedfemoralcomponentscoupledwithpolyethylenemenisci AT licciulliantonio preliminaryinvitrowearassessmentofceramiccementedfemoralcomponentscoupledwithpolyethylenemenisci |