ClearSight™ finger cuff versus invasive arterial pressure measurement in patients with body mass index above 45 kg/m(2)
BACKGROUND: Measuring blood pressure in patients with obesity is challenging. The ClearSight™ finger cuff (FC) uses the vascular unloading technique to provide continuous non-invasive blood pressure measurements. We aimed to test the agreement of the FC with invasive radial arterial monitoring (INV)...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8130355/ https://www.ncbi.nlm.nih.gov/pubmed/34006231 http://dx.doi.org/10.1186/s12871-021-01374-x |
Sumario: | BACKGROUND: Measuring blood pressure in patients with obesity is challenging. The ClearSight™ finger cuff (FC) uses the vascular unloading technique to provide continuous non-invasive blood pressure measurements. We aimed to test the agreement of the FC with invasive radial arterial monitoring (INV) in patients with obesity. METHODS: Participants had a body mass index (BMI) ≥45 kg/m(2) and underwent laparoscopic bariatric surgery. FC and INV measurements were obtained simultaneously every 5 min on each patient, following induction of anesthesia. Agreement over time was assessed using modified Bland-Altman plots and error grid analysis permitted clinical interpretation of the results. Four-quadrant plots allowed assessment of concordance in blood pressure changes. RESULTS: The 30 participants had a median (IQR) BMI of 50.2 kg/m(2) (IQR 48.3–55.3). The observed bias (SD, 95% limits of agreement) for systolic blood pressure (SBP) was 14.3 mmHg (14.1, -13.4 – 42.0), 5.2 mmHg (10.9, -16.0 – 26.5) for mean arterial pressure (MAP) and 2.6 mmHg (10.8, -18.6 – 23.8) for diastolic blood pressure (DBP). Error grid analysis showed that the proportion of readings in risk zones A-E were 90.8, 6.5, 2.7, 0 and 0% for SBP and 91.4, 4.3, 4.3, 0 and 0% for MAP, respectively. Discordance occurred in ≤8% of pairs for consecutive change in SBP, MAP and DBP. CONCLUSIONS: The vascular unloading technique was not adequately in agreement with radial arterial monitoring. Evaluation in a larger sample is required before recommending this technique for intraoperative monitoring of patients with BMI ≥45 kg/m(2). |
---|