Innovative procedure for measuring left ventricular ejection fraction from (18)F-FDG first-pass ultra-sensitive digital PET/CT images: evaluation with an anthropomorphic heart phantom
BACKGROUND: Left ventricular ejection fraction (LVEF) is usually measured by cine-cardiac magnetic resonance imaging (MRI), planar and single-photon emission-computerized tomography (SPECT) equilibrium radionuclide angiocardiography (ERNA), and echocardiography. It would be clinically useful to meas...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8137784/ https://www.ncbi.nlm.nih.gov/pubmed/34018067 http://dx.doi.org/10.1186/s40658-021-00387-2 |
_version_ | 1783695671413440512 |
---|---|
author | Verrecchia-Ramos, Emilie Morel, Olivier Retif, Paul Ben Mahmoud, Sinan |
author_facet | Verrecchia-Ramos, Emilie Morel, Olivier Retif, Paul Ben Mahmoud, Sinan |
author_sort | Verrecchia-Ramos, Emilie |
collection | PubMed |
description | BACKGROUND: Left ventricular ejection fraction (LVEF) is usually measured by cine-cardiac magnetic resonance imaging (MRI), planar and single-photon emission-computerized tomography (SPECT) equilibrium radionuclide angiocardiography (ERNA), and echocardiography. It would be clinically useful to measure LVEF from first-pass positron-emission tomography/computed tomography (PET/CT) radionuclide angiography, but this approach has been limited by fast radiotracer diffusion. Ultra-sensitive digital PET systems can produce high-quality images within 3-s acquisition times. This study determined whether digital PET/CT accurately measured LVEF in an anthropomorphic heart phantom under conditions mimicking radiotracer first-pass into the cardiac cavities. METHODS: Heart phantoms in end-diastole and end-systole were 3D-printed from a patient’s MRI dataset. Reference left ventricle end-diastole volume (EDV), end-systole volume (ESV), and LVEF were determined by phantom weights before/after water filling. PET/CT (3-s acquisitions), MRI, and planar and SPECT ERNA were performed. EDV, ESV, and/or LVEF were measured by manual and automated cardiac cavity delineation, using clinical segmentation softwares. LVEF was also measured from PET images converted to 2D “pseudo-planar” images along the short axis and horizontal long axis. LVEF was also calculated for planar ERNA images. All LVEF, ESV and EDV values were compared to the reference values assessed by weighing. RESULTS: Manually calculated 3D-PET-CT-based EDV, ESV, and LVEF were close to MRI and reference values. Automated calculations on the 3D-PET-CT dataset were unreliable, suggesting that the SPECT-based tool used for this calculation is not well adapted for PET acquisitions. Manual and automated LVEF estimations from “pseudo-planar” PET images were very close/identical to MRI and reference values. CONCLUSIONS: First-pass “pseudo-planar” PET may be a promising method for estimating LVEF, easy to use in clinical practice. Processing 3D PET images is also a valid method but to date suffers from a lack of well-suited software for automated LV segmentation. |
format | Online Article Text |
id | pubmed-8137784 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-81377842021-06-03 Innovative procedure for measuring left ventricular ejection fraction from (18)F-FDG first-pass ultra-sensitive digital PET/CT images: evaluation with an anthropomorphic heart phantom Verrecchia-Ramos, Emilie Morel, Olivier Retif, Paul Ben Mahmoud, Sinan EJNMMI Phys Original Research BACKGROUND: Left ventricular ejection fraction (LVEF) is usually measured by cine-cardiac magnetic resonance imaging (MRI), planar and single-photon emission-computerized tomography (SPECT) equilibrium radionuclide angiocardiography (ERNA), and echocardiography. It would be clinically useful to measure LVEF from first-pass positron-emission tomography/computed tomography (PET/CT) radionuclide angiography, but this approach has been limited by fast radiotracer diffusion. Ultra-sensitive digital PET systems can produce high-quality images within 3-s acquisition times. This study determined whether digital PET/CT accurately measured LVEF in an anthropomorphic heart phantom under conditions mimicking radiotracer first-pass into the cardiac cavities. METHODS: Heart phantoms in end-diastole and end-systole were 3D-printed from a patient’s MRI dataset. Reference left ventricle end-diastole volume (EDV), end-systole volume (ESV), and LVEF were determined by phantom weights before/after water filling. PET/CT (3-s acquisitions), MRI, and planar and SPECT ERNA were performed. EDV, ESV, and/or LVEF were measured by manual and automated cardiac cavity delineation, using clinical segmentation softwares. LVEF was also measured from PET images converted to 2D “pseudo-planar” images along the short axis and horizontal long axis. LVEF was also calculated for planar ERNA images. All LVEF, ESV and EDV values were compared to the reference values assessed by weighing. RESULTS: Manually calculated 3D-PET-CT-based EDV, ESV, and LVEF were close to MRI and reference values. Automated calculations on the 3D-PET-CT dataset were unreliable, suggesting that the SPECT-based tool used for this calculation is not well adapted for PET acquisitions. Manual and automated LVEF estimations from “pseudo-planar” PET images were very close/identical to MRI and reference values. CONCLUSIONS: First-pass “pseudo-planar” PET may be a promising method for estimating LVEF, easy to use in clinical practice. Processing 3D PET images is also a valid method but to date suffers from a lack of well-suited software for automated LV segmentation. Springer International Publishing 2021-05-20 /pmc/articles/PMC8137784/ /pubmed/34018067 http://dx.doi.org/10.1186/s40658-021-00387-2 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . |
spellingShingle | Original Research Verrecchia-Ramos, Emilie Morel, Olivier Retif, Paul Ben Mahmoud, Sinan Innovative procedure for measuring left ventricular ejection fraction from (18)F-FDG first-pass ultra-sensitive digital PET/CT images: evaluation with an anthropomorphic heart phantom |
title | Innovative procedure for measuring left ventricular ejection fraction from (18)F-FDG first-pass ultra-sensitive digital PET/CT images: evaluation with an anthropomorphic heart phantom |
title_full | Innovative procedure for measuring left ventricular ejection fraction from (18)F-FDG first-pass ultra-sensitive digital PET/CT images: evaluation with an anthropomorphic heart phantom |
title_fullStr | Innovative procedure for measuring left ventricular ejection fraction from (18)F-FDG first-pass ultra-sensitive digital PET/CT images: evaluation with an anthropomorphic heart phantom |
title_full_unstemmed | Innovative procedure for measuring left ventricular ejection fraction from (18)F-FDG first-pass ultra-sensitive digital PET/CT images: evaluation with an anthropomorphic heart phantom |
title_short | Innovative procedure for measuring left ventricular ejection fraction from (18)F-FDG first-pass ultra-sensitive digital PET/CT images: evaluation with an anthropomorphic heart phantom |
title_sort | innovative procedure for measuring left ventricular ejection fraction from (18)f-fdg first-pass ultra-sensitive digital pet/ct images: evaluation with an anthropomorphic heart phantom |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8137784/ https://www.ncbi.nlm.nih.gov/pubmed/34018067 http://dx.doi.org/10.1186/s40658-021-00387-2 |
work_keys_str_mv | AT verrecchiaramosemilie innovativeprocedureformeasuringleftventricularejectionfractionfrom18ffdgfirstpassultrasensitivedigitalpetctimagesevaluationwithananthropomorphicheartphantom AT morelolivier innovativeprocedureformeasuringleftventricularejectionfractionfrom18ffdgfirstpassultrasensitivedigitalpetctimagesevaluationwithananthropomorphicheartphantom AT retifpaul innovativeprocedureformeasuringleftventricularejectionfractionfrom18ffdgfirstpassultrasensitivedigitalpetctimagesevaluationwithananthropomorphicheartphantom AT benmahmoudsinan innovativeprocedureformeasuringleftventricularejectionfractionfrom18ffdgfirstpassultrasensitivedigitalpetctimagesevaluationwithananthropomorphicheartphantom |