Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers

OBJECTIVES: Whole blood bilirubin measured on blood gas analyzers is accepted by physicians in neonatal hyperbilirubinemia management since it requires a small sample volume. The accuracy of bilirubin measurement on blood gas analyzers is instrument dependent and remains controversial. DESIGN AND ME...

Descripción completa

Detalles Bibliográficos
Autores principales: Huang, Yun, Dean, Robert, Dubbelman, Yvonne, Vincent, Anne, Khurshid, Faiza
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8145737/
https://www.ncbi.nlm.nih.gov/pubmed/34095416
http://dx.doi.org/10.1016/j.plabm.2021.e00231
_version_ 1783697244330917888
author Huang, Yun
Dean, Robert
Dubbelman, Yvonne
Vincent, Anne
Khurshid, Faiza
author_facet Huang, Yun
Dean, Robert
Dubbelman, Yvonne
Vincent, Anne
Khurshid, Faiza
author_sort Huang, Yun
collection PubMed
description OBJECTIVES: Whole blood bilirubin measured on blood gas analyzers is accepted by physicians in neonatal hyperbilirubinemia management since it requires a small sample volume. The accuracy of bilirubin measurement on blood gas analyzers is instrument dependent and remains controversial. DESIGN AND METHODS: Bilirubin in adult and umbilical cord whole blood samples, spiked with an unconjugated bilirubin standard, and non-spiked adult plasma samples was measured on a blood gas analyzer (GEM 4000) and a Core Laboratory Chemistry analyzer (Architect c16000) respectively. We also investigated the linear regression for neonatal and adult hemoglobin measured on the blood gas analyzer and the Core Laboratory hematology analyzer (Alinity h-Series). RESULTS: Plasma bilirubin measured on the blood gas analyzer and the chemistry analyzer was statistically identical. Adult whole blood bilirubin showed slightly increased proportional bias. When umbilical cord whole blood samples were used, the Deming regression showed GEM bilirubin =1.233(Architect) (95% CI 1.199 ~ 1.266)-44.43 ​μmol/L (95% CI -53.6 ​~ ​−35.2). The regression was significantly different from that in plasma (p ​< ​0.001) or adult whole blood (p ​< ​0.001) samples. 36.1% neonatal samples with bilirubin levels >50 ​μmol/L showed that the bias% was above laboratory standards. In addition, the regression of neonatal hemoglobin measurement between the GEM and the Alinity was significantly different from adult hemoglobin (p ​< ​0.01). CONCLUSIONS: Neonatal whole blood bilirubin measurement on blood gas analyzers may be affected by neonatal hemoglobin. The method should be validated using neonatal whole blood samples or samples with a similar matrix before the analyzers are implemented into neonatal hyperbilirubinemia management.
format Online
Article
Text
id pubmed-8145737
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-81457372021-06-03 Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers Huang, Yun Dean, Robert Dubbelman, Yvonne Vincent, Anne Khurshid, Faiza Pract Lab Med Full Length Article OBJECTIVES: Whole blood bilirubin measured on blood gas analyzers is accepted by physicians in neonatal hyperbilirubinemia management since it requires a small sample volume. The accuracy of bilirubin measurement on blood gas analyzers is instrument dependent and remains controversial. DESIGN AND METHODS: Bilirubin in adult and umbilical cord whole blood samples, spiked with an unconjugated bilirubin standard, and non-spiked adult plasma samples was measured on a blood gas analyzer (GEM 4000) and a Core Laboratory Chemistry analyzer (Architect c16000) respectively. We also investigated the linear regression for neonatal and adult hemoglobin measured on the blood gas analyzer and the Core Laboratory hematology analyzer (Alinity h-Series). RESULTS: Plasma bilirubin measured on the blood gas analyzer and the chemistry analyzer was statistically identical. Adult whole blood bilirubin showed slightly increased proportional bias. When umbilical cord whole blood samples were used, the Deming regression showed GEM bilirubin =1.233(Architect) (95% CI 1.199 ~ 1.266)-44.43 ​μmol/L (95% CI -53.6 ​~ ​−35.2). The regression was significantly different from that in plasma (p ​< ​0.001) or adult whole blood (p ​< ​0.001) samples. 36.1% neonatal samples with bilirubin levels >50 ​μmol/L showed that the bias% was above laboratory standards. In addition, the regression of neonatal hemoglobin measurement between the GEM and the Alinity was significantly different from adult hemoglobin (p ​< ​0.01). CONCLUSIONS: Neonatal whole blood bilirubin measurement on blood gas analyzers may be affected by neonatal hemoglobin. The method should be validated using neonatal whole blood samples or samples with a similar matrix before the analyzers are implemented into neonatal hyperbilirubinemia management. Elsevier 2021-04-27 /pmc/articles/PMC8145737/ /pubmed/34095416 http://dx.doi.org/10.1016/j.plabm.2021.e00231 Text en © 2021 The Authors https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Full Length Article
Huang, Yun
Dean, Robert
Dubbelman, Yvonne
Vincent, Anne
Khurshid, Faiza
Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers
title Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers
title_full Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers
title_fullStr Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers
title_full_unstemmed Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers
title_short Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers
title_sort neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on gem premier 4000 blood gas analyzers
topic Full Length Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8145737/
https://www.ncbi.nlm.nih.gov/pubmed/34095416
http://dx.doi.org/10.1016/j.plabm.2021.e00231
work_keys_str_mv AT huangyun neonatalhemoglobinaffectstheaccuracyofwholebloodbilirubinmeasurementongempremier4000bloodgasanalyzers
AT deanrobert neonatalhemoglobinaffectstheaccuracyofwholebloodbilirubinmeasurementongempremier4000bloodgasanalyzers
AT dubbelmanyvonne neonatalhemoglobinaffectstheaccuracyofwholebloodbilirubinmeasurementongempremier4000bloodgasanalyzers
AT vincentanne neonatalhemoglobinaffectstheaccuracyofwholebloodbilirubinmeasurementongempremier4000bloodgasanalyzers
AT khurshidfaiza neonatalhemoglobinaffectstheaccuracyofwholebloodbilirubinmeasurementongempremier4000bloodgasanalyzers