Evaluation of a new automated immunoassay for the quantification of anti-Müllerian hormone
OBJECTIVES: A newly developed fully automated Lumipulse G AMH method (Fujirebio Diagnostics) was recently introduced in clinical laboratories for quantitative determination of anti-Müllerian hormone (AMH) level in human serum or plasma. AMH has emerged as value-added biomarker in the assessment of o...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8145750/ https://www.ncbi.nlm.nih.gov/pubmed/34095410 http://dx.doi.org/10.1016/j.plabm.2021.e00220 |
Sumario: | OBJECTIVES: A newly developed fully automated Lumipulse G AMH method (Fujirebio Diagnostics) was recently introduced in clinical laboratories for quantitative determination of anti-Müllerian hormone (AMH) level in human serum or plasma. AMH has emerged as value-added biomarker in the assessment of ovarian reserve, in diagnosis of granulosa cells cancer and in the investigation of gonadal disorders. We compared Lumipulse G AMH assay performances with other methods largely applied for AMH measurements. DESIGN AND METHODS: The Lumipulse G AMH method based on two-step sandwich chemiluminescence enzyme immunoassay was assessed on Lumipulse G600II analyzer. The evaluation study included imprecisions, sensitivity and linearity whereas a comparison study was performed on a heterogeneous population of 114 patients by using the Elecsys AMH Plus assay on COBAS 8000 e602 module (Roche Diagnostics). RESULTS: Lumipulse G AMH system showed good repeatability (within-run imprecision) with CV values below 1% (0.5% and 0.9% for high and low serum pools). Similarly within-laboratory imprecision was assessed with CV values of 2.5% and 1.6% for high and low level controls respectively. A linearity regression formula of 1.0119x-0.067 with a coefficient of determination (r(2)) equal to 0.999 was obtained in a range from 0.044 to 22.42 ng/ml. Passing-Bablok regression analysis was performed for assay comparability of AMH measurements. Results were closely correlated (correlation coefficient = 0.997) with a regression equation (y = 1.230x-0.025) showing a positive slope. Also, Bland-Altman analysis confirmed a good agreement between Lumipulse G AMH and Roche Elecsys AMH Plus assays with a bias of 17.76% in a large measurement range. CONCLUSIONS: The performance of Lumipulse G AMH system was highly comparable with that of Roche Elecsys AMH Plus assay although approximately 10% higher values of AMH levels were observed for Lumipulse AMH system at all range of concentrations. Nevertheless the Lumipulse G system seems to be largely suitable for quantitative determination of AMH level in small-scale laboratory because of the reduced size and the use of single cartridge per test assuring flexibility and easy handling. |
---|