Synchronized Automatic Gain Control in Bilateral Cochlear Implant Recipients Yields Significant Benefit in Static and Dynamic Listening Conditions

Individuals with bilateral cochlear implants (BiCIs) rely mostly on interaural level difference (ILD) cues to localize stationary sounds in the horizontal plane. Independent automatic gain control (AGC) in each device can distort this cue, resulting in poorer localization of stationary sound sources...

Descripción completa

Detalles Bibliográficos
Autores principales: Dwyer, Robert T., Chen, Chen, Hehrmann, Phillipp, Dwyer, Nichole C., Gifford, René H.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: SAGE Publications 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8150445/
https://www.ncbi.nlm.nih.gov/pubmed/34027718
http://dx.doi.org/10.1177/23312165211014139
_version_ 1783698153994715136
author Dwyer, Robert T.
Chen, Chen
Hehrmann, Phillipp
Dwyer, Nichole C.
Gifford, René H.
author_facet Dwyer, Robert T.
Chen, Chen
Hehrmann, Phillipp
Dwyer, Nichole C.
Gifford, René H.
author_sort Dwyer, Robert T.
collection PubMed
description Individuals with bilateral cochlear implants (BiCIs) rely mostly on interaural level difference (ILD) cues to localize stationary sounds in the horizontal plane. Independent automatic gain control (AGC) in each device can distort this cue, resulting in poorer localization of stationary sound sources. However, little is known about how BiCI listeners perceive sound in motion. In this study, 12 BiCI listeners’ spatial hearing abilities were assessed for both static and dynamic listening conditions when the sound processors were synchronized by applying the same compression gain to both devices as a means to better preserve the original ILD cues. Stimuli consisted of band-pass filtered (100–8000 Hz) Gaussian noise presented at various locations or panned over an array of loudspeakers. In the static listening condition, the distance between two sequentially presented stimuli was adaptively varied to arrive at the minimum audible angle, the smallest spatial separation at which the listener can correctly determine whether the second sound was to the left or right of the first. In the dynamic listening condition, participants identified if a single stimulus moved to the left or to the right. Velocity was held constant and the distance the stimulus traveled was adjusted using an adaptive procedure to determine the minimum audible movement angle. Median minimum audible angle decreased from 17.1° to 15.3° with the AGC synchronized. Median minimum audible movement angle decreased from 100° to 25.5°. These findings were statistically significant and support the hypothesis that synchronizing the AGC better preserves ILD cues and results in improved spatial hearing abilities. However, restoration of the ILD cue alone was not enough to bridge the large performance gap between BiCI listeners and normal-hearing listeners on these static and dynamic spatial hearing measures.
format Online
Article
Text
id pubmed-8150445
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher SAGE Publications
record_format MEDLINE/PubMed
spelling pubmed-81504452021-06-07 Synchronized Automatic Gain Control in Bilateral Cochlear Implant Recipients Yields Significant Benefit in Static and Dynamic Listening Conditions Dwyer, Robert T. Chen, Chen Hehrmann, Phillipp Dwyer, Nichole C. Gifford, René H. Trends Hear Original Article Individuals with bilateral cochlear implants (BiCIs) rely mostly on interaural level difference (ILD) cues to localize stationary sounds in the horizontal plane. Independent automatic gain control (AGC) in each device can distort this cue, resulting in poorer localization of stationary sound sources. However, little is known about how BiCI listeners perceive sound in motion. In this study, 12 BiCI listeners’ spatial hearing abilities were assessed for both static and dynamic listening conditions when the sound processors were synchronized by applying the same compression gain to both devices as a means to better preserve the original ILD cues. Stimuli consisted of band-pass filtered (100–8000 Hz) Gaussian noise presented at various locations or panned over an array of loudspeakers. In the static listening condition, the distance between two sequentially presented stimuli was adaptively varied to arrive at the minimum audible angle, the smallest spatial separation at which the listener can correctly determine whether the second sound was to the left or right of the first. In the dynamic listening condition, participants identified if a single stimulus moved to the left or to the right. Velocity was held constant and the distance the stimulus traveled was adjusted using an adaptive procedure to determine the minimum audible movement angle. Median minimum audible angle decreased from 17.1° to 15.3° with the AGC synchronized. Median minimum audible movement angle decreased from 100° to 25.5°. These findings were statistically significant and support the hypothesis that synchronizing the AGC better preserves ILD cues and results in improved spatial hearing abilities. However, restoration of the ILD cue alone was not enough to bridge the large performance gap between BiCI listeners and normal-hearing listeners on these static and dynamic spatial hearing measures. SAGE Publications 2021-05-24 /pmc/articles/PMC8150445/ /pubmed/34027718 http://dx.doi.org/10.1177/23312165211014139 Text en © The Author(s) 2021 https://creativecommons.org/licenses/by-nc/4.0/Creative Commons Non Commercial CC BY-NC: This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
spellingShingle Original Article
Dwyer, Robert T.
Chen, Chen
Hehrmann, Phillipp
Dwyer, Nichole C.
Gifford, René H.
Synchronized Automatic Gain Control in Bilateral Cochlear Implant Recipients Yields Significant Benefit in Static and Dynamic Listening Conditions
title Synchronized Automatic Gain Control in Bilateral Cochlear Implant Recipients Yields Significant Benefit in Static and Dynamic Listening Conditions
title_full Synchronized Automatic Gain Control in Bilateral Cochlear Implant Recipients Yields Significant Benefit in Static and Dynamic Listening Conditions
title_fullStr Synchronized Automatic Gain Control in Bilateral Cochlear Implant Recipients Yields Significant Benefit in Static and Dynamic Listening Conditions
title_full_unstemmed Synchronized Automatic Gain Control in Bilateral Cochlear Implant Recipients Yields Significant Benefit in Static and Dynamic Listening Conditions
title_short Synchronized Automatic Gain Control in Bilateral Cochlear Implant Recipients Yields Significant Benefit in Static and Dynamic Listening Conditions
title_sort synchronized automatic gain control in bilateral cochlear implant recipients yields significant benefit in static and dynamic listening conditions
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8150445/
https://www.ncbi.nlm.nih.gov/pubmed/34027718
http://dx.doi.org/10.1177/23312165211014139
work_keys_str_mv AT dwyerrobertt synchronizedautomaticgaincontrolinbilateralcochlearimplantrecipientsyieldssignificantbenefitinstaticanddynamiclisteningconditions
AT chenchen synchronizedautomaticgaincontrolinbilateralcochlearimplantrecipientsyieldssignificantbenefitinstaticanddynamiclisteningconditions
AT hehrmannphillipp synchronizedautomaticgaincontrolinbilateralcochlearimplantrecipientsyieldssignificantbenefitinstaticanddynamiclisteningconditions
AT dwyernicholec synchronizedautomaticgaincontrolinbilateralcochlearimplantrecipientsyieldssignificantbenefitinstaticanddynamiclisteningconditions
AT giffordreneh synchronizedautomaticgaincontrolinbilateralcochlearimplantrecipientsyieldssignificantbenefitinstaticanddynamiclisteningconditions