Cargando…

Impact of a Smart-Ring-Based Feedback System on the Quality of Chest Compressions in Adult Cardiac Arrest: A Randomized Preliminary Study

This study aimed to assess the effectiveness of a novel chest compression (CC) smart-ring-based feedback system in a manikin simulation. In this randomized, crossover, controlled study, we evaluated the effect of smart-ring CC feedback on cardiopulmonary resuscitation (CPR). The learnability and usa...

Descripción completa

Detalles Bibliográficos
Autores principales: Ahn, Chiwon, Lee, Seungjae, Lee, Jongshill, Oh, Jaehoon, Song, Yeongtak, Kim, In Young, Kang, Hyunggoo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8158714/
https://www.ncbi.nlm.nih.gov/pubmed/34069369
http://dx.doi.org/10.3390/ijerph18105408
_version_ 1783699922469519360
author Ahn, Chiwon
Lee, Seungjae
Lee, Jongshill
Oh, Jaehoon
Song, Yeongtak
Kim, In Young
Kang, Hyunggoo
author_facet Ahn, Chiwon
Lee, Seungjae
Lee, Jongshill
Oh, Jaehoon
Song, Yeongtak
Kim, In Young
Kang, Hyunggoo
author_sort Ahn, Chiwon
collection PubMed
description This study aimed to assess the effectiveness of a novel chest compression (CC) smart-ring-based feedback system in a manikin simulation. In this randomized, crossover, controlled study, we evaluated the effect of smart-ring CC feedback on cardiopulmonary resuscitation (CPR). The learnability and usability of the tool were evaluated with the System Usability Scale (SUS). Participants were divided into two groups and each performed CCs with and without feedback 2 weeks apart, using different orders. The primary outcome was compression depth; the proportion of accurate-depth (5–6 cm) CCs, CC rate, and the proportion of complete CCs (≤1 cm of residual leaning) were assessed additionally. The feedback group and the non-feedback group showed significant differences in compression depth (52.1 (46.3–54.8) vs. 47.1 (40.5–49.9) mm, p = 0.021). The proportion of accurate-depth CCs was significantly higher in the interventional than in the control condition (88.7 (30.0–99.1) vs. 22.6 (0.0–58.5%), p = 0.033). The mean SUS score was 83.9 ± 8.7 points. The acceptability ranges were ‘acceptable’, and the adjective rating was ‘excellent’. CCs with smart-ring feedback could help achieve the ideal range of depth during CPR. The smart-ring may be a valuable source of CPR feedback.
format Online
Article
Text
id pubmed-8158714
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-81587142021-05-28 Impact of a Smart-Ring-Based Feedback System on the Quality of Chest Compressions in Adult Cardiac Arrest: A Randomized Preliminary Study Ahn, Chiwon Lee, Seungjae Lee, Jongshill Oh, Jaehoon Song, Yeongtak Kim, In Young Kang, Hyunggoo Int J Environ Res Public Health Article This study aimed to assess the effectiveness of a novel chest compression (CC) smart-ring-based feedback system in a manikin simulation. In this randomized, crossover, controlled study, we evaluated the effect of smart-ring CC feedback on cardiopulmonary resuscitation (CPR). The learnability and usability of the tool were evaluated with the System Usability Scale (SUS). Participants were divided into two groups and each performed CCs with and without feedback 2 weeks apart, using different orders. The primary outcome was compression depth; the proportion of accurate-depth (5–6 cm) CCs, CC rate, and the proportion of complete CCs (≤1 cm of residual leaning) were assessed additionally. The feedback group and the non-feedback group showed significant differences in compression depth (52.1 (46.3–54.8) vs. 47.1 (40.5–49.9) mm, p = 0.021). The proportion of accurate-depth CCs was significantly higher in the interventional than in the control condition (88.7 (30.0–99.1) vs. 22.6 (0.0–58.5%), p = 0.033). The mean SUS score was 83.9 ± 8.7 points. The acceptability ranges were ‘acceptable’, and the adjective rating was ‘excellent’. CCs with smart-ring feedback could help achieve the ideal range of depth during CPR. The smart-ring may be a valuable source of CPR feedback. MDPI 2021-05-19 /pmc/articles/PMC8158714/ /pubmed/34069369 http://dx.doi.org/10.3390/ijerph18105408 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Ahn, Chiwon
Lee, Seungjae
Lee, Jongshill
Oh, Jaehoon
Song, Yeongtak
Kim, In Young
Kang, Hyunggoo
Impact of a Smart-Ring-Based Feedback System on the Quality of Chest Compressions in Adult Cardiac Arrest: A Randomized Preliminary Study
title Impact of a Smart-Ring-Based Feedback System on the Quality of Chest Compressions in Adult Cardiac Arrest: A Randomized Preliminary Study
title_full Impact of a Smart-Ring-Based Feedback System on the Quality of Chest Compressions in Adult Cardiac Arrest: A Randomized Preliminary Study
title_fullStr Impact of a Smart-Ring-Based Feedback System on the Quality of Chest Compressions in Adult Cardiac Arrest: A Randomized Preliminary Study
title_full_unstemmed Impact of a Smart-Ring-Based Feedback System on the Quality of Chest Compressions in Adult Cardiac Arrest: A Randomized Preliminary Study
title_short Impact of a Smart-Ring-Based Feedback System on the Quality of Chest Compressions in Adult Cardiac Arrest: A Randomized Preliminary Study
title_sort impact of a smart-ring-based feedback system on the quality of chest compressions in adult cardiac arrest: a randomized preliminary study
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8158714/
https://www.ncbi.nlm.nih.gov/pubmed/34069369
http://dx.doi.org/10.3390/ijerph18105408
work_keys_str_mv AT ahnchiwon impactofasmartringbasedfeedbacksystemonthequalityofchestcompressionsinadultcardiacarrestarandomizedpreliminarystudy
AT leeseungjae impactofasmartringbasedfeedbacksystemonthequalityofchestcompressionsinadultcardiacarrestarandomizedpreliminarystudy
AT leejongshill impactofasmartringbasedfeedbacksystemonthequalityofchestcompressionsinadultcardiacarrestarandomizedpreliminarystudy
AT ohjaehoon impactofasmartringbasedfeedbacksystemonthequalityofchestcompressionsinadultcardiacarrestarandomizedpreliminarystudy
AT songyeongtak impactofasmartringbasedfeedbacksystemonthequalityofchestcompressionsinadultcardiacarrestarandomizedpreliminarystudy
AT kiminyoung impactofasmartringbasedfeedbacksystemonthequalityofchestcompressionsinadultcardiacarrestarandomizedpreliminarystudy
AT kanghyunggoo impactofasmartringbasedfeedbacksystemonthequalityofchestcompressionsinadultcardiacarrestarandomizedpreliminarystudy