In Vitro and In Vivo Comparative Evaluation of a Shellac-Ammonium Paclitaxel-Coated Balloon versus a Benchmark Device
OBJECTIVES: The present study was designed to compare the characteristics and performance regarding drug delivery of a novel drug-coated balloon (DCB) to a benchmark device (Restore® versus SeQuent® Please) in an in vitro and in vivo model. BACKGROUND: Although Restore® and SeQuent® are both paclita...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8172300/ https://www.ncbi.nlm.nih.gov/pubmed/34121951 http://dx.doi.org/10.1155/2021/9962313 |
_version_ | 1783702515017056256 |
---|---|
author | Xia, Congying Jiang, Yunhan Li, Shuangshuang Xiong, Dan Chen, Xiaojie Chen, Yufang |
author_facet | Xia, Congying Jiang, Yunhan Li, Shuangshuang Xiong, Dan Chen, Xiaojie Chen, Yufang |
author_sort | Xia, Congying |
collection | PubMed |
description | OBJECTIVES: The present study was designed to compare the characteristics and performance regarding drug delivery of a novel drug-coated balloon (DCB) to a benchmark device (Restore® versus SeQuent® Please) in an in vitro and in vivo model. BACKGROUND: Although Restore® and SeQuent® are both paclitaxel-coated, they use different coating excipient, shellac-ammonium salt and iopromide, respectively. Preclinical study comparing these two different commercial DCBs regarding their characteristics and effects on early vascular response is sparse. METHODS: Restore® and SeQuent® DCBs were scanned with electron microscopy for surface characteristic assessment. Both DCBs were transported in an in vitro vessel model for the evaluation of drug wash-off rate and particulate formation. Eighteen coronary angioplasties with either Restore® or SeQuent® DCBs were conducted in 6 swine (three coronary vessels each). Histopathological images of each vessel were evaluated for vessel injury. RESULTS: The surface of Restore® DCB was smooth and evenly distributed with hardly visible crystal, while SeQuent® DCB showed a rougher surface with relatively larger apparent crystals. Restore® DCB had a lower drug wash-off rate and fewer large visible particles, compared to the SeQuent® DCB. No significant difference in mean injure score was found between Restore® and SeQuent® group. CONCLUSION: Our results suggest that Restore® is better in preclinical performance regarding less release of particles and lower drug wash-off rate as compared to SeQuent® Please. The Restore® DCB, using stable amorphous coating and shellac-ammonium salt as an excipient, appears to provide an advantage in drug delivery efficacy; however, further clinical studies are warranted. |
format | Online Article Text |
id | pubmed-8172300 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-81723002021-06-11 In Vitro and In Vivo Comparative Evaluation of a Shellac-Ammonium Paclitaxel-Coated Balloon versus a Benchmark Device Xia, Congying Jiang, Yunhan Li, Shuangshuang Xiong, Dan Chen, Xiaojie Chen, Yufang J Interv Cardiol Research Article OBJECTIVES: The present study was designed to compare the characteristics and performance regarding drug delivery of a novel drug-coated balloon (DCB) to a benchmark device (Restore® versus SeQuent® Please) in an in vitro and in vivo model. BACKGROUND: Although Restore® and SeQuent® are both paclitaxel-coated, they use different coating excipient, shellac-ammonium salt and iopromide, respectively. Preclinical study comparing these two different commercial DCBs regarding their characteristics and effects on early vascular response is sparse. METHODS: Restore® and SeQuent® DCBs were scanned with electron microscopy for surface characteristic assessment. Both DCBs were transported in an in vitro vessel model for the evaluation of drug wash-off rate and particulate formation. Eighteen coronary angioplasties with either Restore® or SeQuent® DCBs were conducted in 6 swine (three coronary vessels each). Histopathological images of each vessel were evaluated for vessel injury. RESULTS: The surface of Restore® DCB was smooth and evenly distributed with hardly visible crystal, while SeQuent® DCB showed a rougher surface with relatively larger apparent crystals. Restore® DCB had a lower drug wash-off rate and fewer large visible particles, compared to the SeQuent® DCB. No significant difference in mean injure score was found between Restore® and SeQuent® group. CONCLUSION: Our results suggest that Restore® is better in preclinical performance regarding less release of particles and lower drug wash-off rate as compared to SeQuent® Please. The Restore® DCB, using stable amorphous coating and shellac-ammonium salt as an excipient, appears to provide an advantage in drug delivery efficacy; however, further clinical studies are warranted. Hindawi 2021-05-25 /pmc/articles/PMC8172300/ /pubmed/34121951 http://dx.doi.org/10.1155/2021/9962313 Text en Copyright © 2021 Congying Xia et al. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Xia, Congying Jiang, Yunhan Li, Shuangshuang Xiong, Dan Chen, Xiaojie Chen, Yufang In Vitro and In Vivo Comparative Evaluation of a Shellac-Ammonium Paclitaxel-Coated Balloon versus a Benchmark Device |
title | In Vitro and In Vivo Comparative Evaluation of a Shellac-Ammonium Paclitaxel-Coated Balloon versus a Benchmark Device |
title_full | In Vitro and In Vivo Comparative Evaluation of a Shellac-Ammonium Paclitaxel-Coated Balloon versus a Benchmark Device |
title_fullStr | In Vitro and In Vivo Comparative Evaluation of a Shellac-Ammonium Paclitaxel-Coated Balloon versus a Benchmark Device |
title_full_unstemmed | In Vitro and In Vivo Comparative Evaluation of a Shellac-Ammonium Paclitaxel-Coated Balloon versus a Benchmark Device |
title_short | In Vitro and In Vivo Comparative Evaluation of a Shellac-Ammonium Paclitaxel-Coated Balloon versus a Benchmark Device |
title_sort | in vitro and in vivo comparative evaluation of a shellac-ammonium paclitaxel-coated balloon versus a benchmark device |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8172300/ https://www.ncbi.nlm.nih.gov/pubmed/34121951 http://dx.doi.org/10.1155/2021/9962313 |
work_keys_str_mv | AT xiacongying invitroandinvivocomparativeevaluationofashellacammoniumpaclitaxelcoatedballoonversusabenchmarkdevice AT jiangyunhan invitroandinvivocomparativeevaluationofashellacammoniumpaclitaxelcoatedballoonversusabenchmarkdevice AT lishuangshuang invitroandinvivocomparativeevaluationofashellacammoniumpaclitaxelcoatedballoonversusabenchmarkdevice AT xiongdan invitroandinvivocomparativeevaluationofashellacammoniumpaclitaxelcoatedballoonversusabenchmarkdevice AT chenxiaojie invitroandinvivocomparativeevaluationofashellacammoniumpaclitaxelcoatedballoonversusabenchmarkdevice AT chenyufang invitroandinvivocomparativeevaluationofashellacammoniumpaclitaxelcoatedballoonversusabenchmarkdevice |