Performance of the Emprint and Amica Microwave Ablation Systems in ex vivo Porcine Livers: Sphericity and Reproducibility Versus Size

PURPOSE: To investigate the performance of two microwave ablation (MWA) systems regarding ablation volume, ablation shape and variability. MATERIALS AND METHODS: In this ex vivo study, the Emprint and Amica MWA systems were used to ablate porcine livers at 4 different settings of time and power (3 a...

Descripción completa

Detalles Bibliográficos
Autores principales: Hendriks, P., Berkhout, W. E. M., Kaanen, C. I., Sluijter, J. H., Visser, I. J., van den Dobbelsteen, J. J., de Geus-Oei, L. F., Webb, A. G., Burgmans, M. C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer US 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8172387/
https://www.ncbi.nlm.nih.gov/pubmed/33462682
http://dx.doi.org/10.1007/s00270-020-02742-9
Descripción
Sumario:PURPOSE: To investigate the performance of two microwave ablation (MWA) systems regarding ablation volume, ablation shape and variability. MATERIALS AND METHODS: In this ex vivo study, the Emprint and Amica MWA systems were used to ablate porcine livers at 4 different settings of time and power (3 and 5 minutes at 60 and 80 Watt). In total, 48 ablations were analysed for ablation size and shape using Vitrea Advanced Visualization software after acquisition of a 7T MRI scan. RESULTS: Emprint ablations were smaller (11,1 vs. 21,1 mL p < 0.001), more spherical (sphericity index of 0.89 vs. 0.59 p < 0.001) and showed less variability than Amica ablations. In both systems, longer ablation time and higher power resulted in significantly larger ablation volumes. CONCLUSION: Emprint ablations were more spherical, and the results showed a lower variability than those of Amica ablations. This comes at the price of smaller ablation volumes. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at (10.1007/s00270-020-02742-9).