Reliability of hip muscle strength measured in principal and intermediate planes of movement
BACKGROUND: Muscle strength testing is widely used in clinical and athletic populations. Commercially available dynamometers are designed to assess strength in three principal planes (sagittal, transverse, frontal). However, the anatomy of the hip suggests muscles may only be recruited submaximally...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
PeerJ Inc.
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8179216/ https://www.ncbi.nlm.nih.gov/pubmed/34141482 http://dx.doi.org/10.7717/peerj.11521 |
_version_ | 1783703731518308352 |
---|---|
author | Goncalves, Basilio A.M. Saxby, David J. Kositsky, Adam Barrett, Rod S. Diamond, Laura E. |
author_facet | Goncalves, Basilio A.M. Saxby, David J. Kositsky, Adam Barrett, Rod S. Diamond, Laura E. |
author_sort | Goncalves, Basilio A.M. |
collection | PubMed |
description | BACKGROUND: Muscle strength testing is widely used in clinical and athletic populations. Commercially available dynamometers are designed to assess strength in three principal planes (sagittal, transverse, frontal). However, the anatomy of the hip suggests muscles may only be recruited submaximally during tasks performed in these principal planes. OBJECTIVE: To evaluate the inter-session reliability of maximal isometric hip strength in the principal planes and three intermediate planes. METHODS: Twenty participants (26.1 ± 2.7 years, 50% female) attended two testing sessions 6.2 ± 1.8 days apart. Participants completed 3-5 maximal voluntary isometric contractions for hip abduction, adduction, flexion, extension, and internal and external rotation measured using a fixed uniaxial load cell (custom rig) and commercial dynamometer (Biodex). Three intermediate hip actions were also tested using the custom rig: extension with abduction, extension with external rotation, and extension with both abduction and external rotation. RESULTS: Moderate-to-excellent intraclass correlation coefficients were observed for all principal and intermediate muscle actions using the custom rig (0.72–0.95) and the Biodex (0.85–0.95). The minimum detectable change was also similar between devices (custom rig = 11–31%; Biodex = 9–20%). Bland-Altman analysis revealed poor agreement between devices (range between upper and lower limits of agreement = 77–131%). CONCLUSIONS: Although the custom rig and Biodex showed similar reliability, both devices may lack the sensitivity to detect small changes in hip strength commonly observed following intervention. |
format | Online Article Text |
id | pubmed-8179216 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | PeerJ Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-81792162021-06-16 Reliability of hip muscle strength measured in principal and intermediate planes of movement Goncalves, Basilio A.M. Saxby, David J. Kositsky, Adam Barrett, Rod S. Diamond, Laura E. PeerJ Biophysics BACKGROUND: Muscle strength testing is widely used in clinical and athletic populations. Commercially available dynamometers are designed to assess strength in three principal planes (sagittal, transverse, frontal). However, the anatomy of the hip suggests muscles may only be recruited submaximally during tasks performed in these principal planes. OBJECTIVE: To evaluate the inter-session reliability of maximal isometric hip strength in the principal planes and three intermediate planes. METHODS: Twenty participants (26.1 ± 2.7 years, 50% female) attended two testing sessions 6.2 ± 1.8 days apart. Participants completed 3-5 maximal voluntary isometric contractions for hip abduction, adduction, flexion, extension, and internal and external rotation measured using a fixed uniaxial load cell (custom rig) and commercial dynamometer (Biodex). Three intermediate hip actions were also tested using the custom rig: extension with abduction, extension with external rotation, and extension with both abduction and external rotation. RESULTS: Moderate-to-excellent intraclass correlation coefficients were observed for all principal and intermediate muscle actions using the custom rig (0.72–0.95) and the Biodex (0.85–0.95). The minimum detectable change was also similar between devices (custom rig = 11–31%; Biodex = 9–20%). Bland-Altman analysis revealed poor agreement between devices (range between upper and lower limits of agreement = 77–131%). CONCLUSIONS: Although the custom rig and Biodex showed similar reliability, both devices may lack the sensitivity to detect small changes in hip strength commonly observed following intervention. PeerJ Inc. 2021-06-02 /pmc/articles/PMC8179216/ /pubmed/34141482 http://dx.doi.org/10.7717/peerj.11521 Text en ©2021 Goncalves et al. https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, reproduction and adaptation in any medium and for any purpose provided that it is properly attributed. For attribution, the original author(s), title, publication source (PeerJ) and either DOI or URL of the article must be cited. |
spellingShingle | Biophysics Goncalves, Basilio A.M. Saxby, David J. Kositsky, Adam Barrett, Rod S. Diamond, Laura E. Reliability of hip muscle strength measured in principal and intermediate planes of movement |
title | Reliability of hip muscle strength measured in principal and intermediate planes of movement |
title_full | Reliability of hip muscle strength measured in principal and intermediate planes of movement |
title_fullStr | Reliability of hip muscle strength measured in principal and intermediate planes of movement |
title_full_unstemmed | Reliability of hip muscle strength measured in principal and intermediate planes of movement |
title_short | Reliability of hip muscle strength measured in principal and intermediate planes of movement |
title_sort | reliability of hip muscle strength measured in principal and intermediate planes of movement |
topic | Biophysics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8179216/ https://www.ncbi.nlm.nih.gov/pubmed/34141482 http://dx.doi.org/10.7717/peerj.11521 |
work_keys_str_mv | AT goncalvesbasilioam reliabilityofhipmusclestrengthmeasuredinprincipalandintermediateplanesofmovement AT saxbydavidj reliabilityofhipmusclestrengthmeasuredinprincipalandintermediateplanesofmovement AT kositskyadam reliabilityofhipmusclestrengthmeasuredinprincipalandintermediateplanesofmovement AT barrettrods reliabilityofhipmusclestrengthmeasuredinprincipalandintermediateplanesofmovement AT diamondlaurae reliabilityofhipmusclestrengthmeasuredinprincipalandintermediateplanesofmovement |