Validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle

Point-of-care (POC) devices that veterinary practitioners can use to easily and rapidly measure blood ionized calcium (iCa) levels in cows immediately after withdrawing a blood sample on the dairy farm are needed. Aims of present studies was to compare the commercially available ion-selective electr...

Descripción completa

Detalles Bibliográficos
Autores principales: SUZUKI, Kazuyuki, KONDO, Nao, TAKAGI, Kaede, NISHIKAWA, Akitoyo, MURAKAMI, Yoshiki, OTSUKA, Marina, TSUKANO, Kenji, IKEDA, Keiko, FUNAKURA, Hisashi, YASUTOMI, Ichiro, KAWAMOTO, Satoshi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Japanese Society of Veterinary Science 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8182322/
https://www.ncbi.nlm.nih.gov/pubmed/33775988
http://dx.doi.org/10.1292/jvms.21-0001
_version_ 1783704185771917312
author SUZUKI, Kazuyuki
KONDO, Nao
TAKAGI, Kaede
NISHIKAWA, Akitoyo
MURAKAMI, Yoshiki
OTSUKA, Marina
TSUKANO, Kenji
IKEDA, Keiko
FUNAKURA, Hisashi
YASUTOMI, Ichiro
KAWAMOTO, Satoshi
author_facet SUZUKI, Kazuyuki
KONDO, Nao
TAKAGI, Kaede
NISHIKAWA, Akitoyo
MURAKAMI, Yoshiki
OTSUKA, Marina
TSUKANO, Kenji
IKEDA, Keiko
FUNAKURA, Hisashi
YASUTOMI, Ichiro
KAWAMOTO, Satoshi
author_sort SUZUKI, Kazuyuki
collection PubMed
description Point-of-care (POC) devices that veterinary practitioners can use to easily and rapidly measure blood ionized calcium (iCa) levels in cows immediately after withdrawing a blood sample on the dairy farm are needed. Aims of present studies was to compare the commercially available ion-selective electrode handheld iCa meter (bovine blood iCa checker) with the benchtop blood gas analyzer GEM premier 3500 and handheld analyzer i-STAT 1. Sixty-two paired-point whole blood samples were obtained from three cows with hypocalcemia experimentally induced by Na(2)-EDTA infusion. Whole blood samples were also obtained from the 36 cows kept on a farm in field conditions. The results using the bovine blood iCa checker correlated with those using the GEM premier 3500 and i-STAT 1. Bovine blood iCa checker was “compatible” with the GEM premier 3500 and i-STAT 1 because the frequency of differences between the measurements within ± 20% of the mean were 100% (65/65, >75%) and 90.8% (59/65, >75%), respectively. In the field trial, the blood iCa concentration measured by the bovine blood Ca checker was significantly positively correlated with that measured by the i-STAT 1 portable analyzer. Bovine blood iCa checker was “compatible” with the i-STAT 1 because the frequency of differences between the measurements within ± 20% of the mean was 100% (36/36, >75%). Results from these findings, the bovine blood iCa checker may be applied as a simplified system to measure the iCa concentration in bovine whole blood.
format Online
Article
Text
id pubmed-8182322
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher The Japanese Society of Veterinary Science
record_format MEDLINE/PubMed
spelling pubmed-81823222021-06-09 Validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle SUZUKI, Kazuyuki KONDO, Nao TAKAGI, Kaede NISHIKAWA, Akitoyo MURAKAMI, Yoshiki OTSUKA, Marina TSUKANO, Kenji IKEDA, Keiko FUNAKURA, Hisashi YASUTOMI, Ichiro KAWAMOTO, Satoshi J Vet Med Sci Clinical Pathology Point-of-care (POC) devices that veterinary practitioners can use to easily and rapidly measure blood ionized calcium (iCa) levels in cows immediately after withdrawing a blood sample on the dairy farm are needed. Aims of present studies was to compare the commercially available ion-selective electrode handheld iCa meter (bovine blood iCa checker) with the benchtop blood gas analyzer GEM premier 3500 and handheld analyzer i-STAT 1. Sixty-two paired-point whole blood samples were obtained from three cows with hypocalcemia experimentally induced by Na(2)-EDTA infusion. Whole blood samples were also obtained from the 36 cows kept on a farm in field conditions. The results using the bovine blood iCa checker correlated with those using the GEM premier 3500 and i-STAT 1. Bovine blood iCa checker was “compatible” with the GEM premier 3500 and i-STAT 1 because the frequency of differences between the measurements within ± 20% of the mean were 100% (65/65, >75%) and 90.8% (59/65, >75%), respectively. In the field trial, the blood iCa concentration measured by the bovine blood Ca checker was significantly positively correlated with that measured by the i-STAT 1 portable analyzer. Bovine blood iCa checker was “compatible” with the i-STAT 1 because the frequency of differences between the measurements within ± 20% of the mean was 100% (36/36, >75%). Results from these findings, the bovine blood iCa checker may be applied as a simplified system to measure the iCa concentration in bovine whole blood. The Japanese Society of Veterinary Science 2021-03-26 2021-05 /pmc/articles/PMC8182322/ /pubmed/33775988 http://dx.doi.org/10.1292/jvms.21-0001 Text en ©2021 The Japanese Society of Veterinary Science https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial No Derivatives (by-nc-nd) License. (CC-BY-NC-ND 4.0: https://creativecommons.org/licenses/by-nc-nd/4.0/)
spellingShingle Clinical Pathology
SUZUKI, Kazuyuki
KONDO, Nao
TAKAGI, Kaede
NISHIKAWA, Akitoyo
MURAKAMI, Yoshiki
OTSUKA, Marina
TSUKANO, Kenji
IKEDA, Keiko
FUNAKURA, Hisashi
YASUTOMI, Ichiro
KAWAMOTO, Satoshi
Validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle
title Validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle
title_full Validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle
title_fullStr Validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle
title_full_unstemmed Validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle
title_short Validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle
title_sort validation of the bovine blood calcium checker as a rapid and simple measuring tool for the ionized calcium concentration in cattle
topic Clinical Pathology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8182322/
https://www.ncbi.nlm.nih.gov/pubmed/33775988
http://dx.doi.org/10.1292/jvms.21-0001
work_keys_str_mv AT suzukikazuyuki validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT kondonao validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT takagikaede validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT nishikawaakitoyo validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT murakamiyoshiki validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT otsukamarina validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT tsukanokenji validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT ikedakeiko validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT funakurahisashi validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT yasutomiichiro validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle
AT kawamotosatoshi validationofthebovinebloodcalciumcheckerasarapidandsimplemeasuringtoolfortheionizedcalciumconcentrationincattle