Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation
Purpose To determine the appropriateness of implementing Mobius3D/FX (Varian Medical Systems, Inc., Palo Alto, CA, USA) as not only a pretreatment secondary check but as an alternative to measurement-based patient-specific intensity-modulated radiation therapy (IMRT) quality assurance (QA). Methods...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Cureus
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8184160/ https://www.ncbi.nlm.nih.gov/pubmed/34113520 http://dx.doi.org/10.7759/cureus.14910 |
_version_ | 1783704533102231552 |
---|---|
author | Basavatia, Amar K Fiedler, Derek A Ritter, Johannes Brodin, Patrik Goddard, Lee C Jeong, Kyoungkeun Hsu, Shu-Hui Mynampati, Dinesh K Yaparpalvi, Ravindra Tomé, Wolfgang A |
author_facet | Basavatia, Amar K Fiedler, Derek A Ritter, Johannes Brodin, Patrik Goddard, Lee C Jeong, Kyoungkeun Hsu, Shu-Hui Mynampati, Dinesh K Yaparpalvi, Ravindra Tomé, Wolfgang A |
author_sort | Basavatia, Amar K |
collection | PubMed |
description | Purpose To determine the appropriateness of implementing Mobius3D/FX (Varian Medical Systems, Inc., Palo Alto, CA, USA) as not only a pretreatment secondary check but as an alternative to measurement-based patient-specific intensity-modulated radiation therapy (IMRT) quality assurance (QA). Methods Mobius3D/FX was commissioned and stock beam models were tweaked so that an independent recalculated 3D dose distribution can be obtained. Then, 50 patient-specific treatment plans for various indications were delivered across a 2D ion chamber array, radiochromic film setup, and electronic portal imager and analyzed with MobiusFX and gamma analysis. The concordance of plans scored as passing between MobiusFX and the conventional methods of QA was determined. Results All analyzed treatment plans passed with a gamma passing rate >90% across all conventional QA methods, most commonly using a 3%/3mm gamma criterion except for film measurements where a 5%/3mm criterion was applied. There was good agreement and concordance between MobiusFX and conventional methods when using a 3%/3mm criteria for MobiusFX, whereas a 2%/2mm criteria appeared too stringent as it failed treatment plans deemed clinically acceptable using conventional methods. Conclusions Using a 50-sample subset of clinically delivered treatment plans this non-inferiority-type comparison shows Mobius3D/FX based on log file analysis to be a suitable alternative to conventional QA methods when utilizing the 3%/3mm gamma criterion. Methods based on log file analysis can provide an opportunity for resource sparing, improving the efficiency, and workflow for evaluating IMRT treatment plans. |
format | Online Article Text |
id | pubmed-8184160 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | Cureus |
record_format | MEDLINE/PubMed |
spelling | pubmed-81841602021-06-09 Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation Basavatia, Amar K Fiedler, Derek A Ritter, Johannes Brodin, Patrik Goddard, Lee C Jeong, Kyoungkeun Hsu, Shu-Hui Mynampati, Dinesh K Yaparpalvi, Ravindra Tomé, Wolfgang A Cureus Medical Physics Purpose To determine the appropriateness of implementing Mobius3D/FX (Varian Medical Systems, Inc., Palo Alto, CA, USA) as not only a pretreatment secondary check but as an alternative to measurement-based patient-specific intensity-modulated radiation therapy (IMRT) quality assurance (QA). Methods Mobius3D/FX was commissioned and stock beam models were tweaked so that an independent recalculated 3D dose distribution can be obtained. Then, 50 patient-specific treatment plans for various indications were delivered across a 2D ion chamber array, radiochromic film setup, and electronic portal imager and analyzed with MobiusFX and gamma analysis. The concordance of plans scored as passing between MobiusFX and the conventional methods of QA was determined. Results All analyzed treatment plans passed with a gamma passing rate >90% across all conventional QA methods, most commonly using a 3%/3mm gamma criterion except for film measurements where a 5%/3mm criterion was applied. There was good agreement and concordance between MobiusFX and conventional methods when using a 3%/3mm criteria for MobiusFX, whereas a 2%/2mm criteria appeared too stringent as it failed treatment plans deemed clinically acceptable using conventional methods. Conclusions Using a 50-sample subset of clinically delivered treatment plans this non-inferiority-type comparison shows Mobius3D/FX based on log file analysis to be a suitable alternative to conventional QA methods when utilizing the 3%/3mm gamma criterion. Methods based on log file analysis can provide an opportunity for resource sparing, improving the efficiency, and workflow for evaluating IMRT treatment plans. Cureus 2021-05-08 /pmc/articles/PMC8184160/ /pubmed/34113520 http://dx.doi.org/10.7759/cureus.14910 Text en Copyright © 2021, Basavatia et al. https://creativecommons.org/licenses/by/3.0/This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Medical Physics Basavatia, Amar K Fiedler, Derek A Ritter, Johannes Brodin, Patrik Goddard, Lee C Jeong, Kyoungkeun Hsu, Shu-Hui Mynampati, Dinesh K Yaparpalvi, Ravindra Tomé, Wolfgang A Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation |
title | Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation |
title_full | Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation |
title_fullStr | Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation |
title_full_unstemmed | Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation |
title_short | Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation |
title_sort | comprehensive patient-specific intensity-modulated radiation therapy quality assurance comparing mobius3d/fx to conventional methods of evaluation |
topic | Medical Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8184160/ https://www.ncbi.nlm.nih.gov/pubmed/34113520 http://dx.doi.org/10.7759/cureus.14910 |
work_keys_str_mv | AT basavatiaamark comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation AT fiedlerdereka comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation AT ritterjohannes comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation AT brodinpatrik comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation AT goddardleec comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation AT jeongkyoungkeun comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation AT hsushuhui comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation AT mynampatidineshk comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation AT yaparpalviravindra comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation AT tomewolfganga comprehensivepatientspecificintensitymodulatedradiationtherapyqualityassurancecomparingmobius3dfxtoconventionalmethodsofevaluation |