Comparison of four routinely used vitamin D automated immunoassays

BACKGROUND: To compare four automated immunoassays for the measurement of 25(OH)-vitamin D (25-OHD) and to assess the impact on the results obtained from a healthy population. METHODS: We analysed 100 serum samples on Unicel DxI 800 (Beckman Coulter), Architect i1000 (Abbott), Cobas e411 (Roche) and...

Descripción completa

Detalles Bibliográficos
Autores principales: Windrichova, Jindra, Broz, Pavel, Fuchsova, Radka, Topolcan, Ondrej, Pecen, Ladislav, Mayer, Otto, Kucera, Radek
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Society of Medical Biochemists of Serbia, Belgrade 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8199437/
https://www.ncbi.nlm.nih.gov/pubmed/34177372
http://dx.doi.org/10.5937/jomb0-27531
_version_ 1783707376686202880
author Windrichova, Jindra
Broz, Pavel
Fuchsova, Radka
Topolcan, Ondrej
Pecen, Ladislav
Mayer, Otto
Kucera, Radek
author_facet Windrichova, Jindra
Broz, Pavel
Fuchsova, Radka
Topolcan, Ondrej
Pecen, Ladislav
Mayer, Otto
Kucera, Radek
author_sort Windrichova, Jindra
collection PubMed
description BACKGROUND: To compare four automated immunoassays for the measurement of 25(OH)-vitamin D (25-OHD) and to assess the impact on the results obtained from a healthy population. METHODS: We analysed 100 serum samples on Unicel DxI 800 (Beckman Coulter), Architect i1000 (Abbott), Cobas e411 (Roche) and Liaison XL (DiaSorin). Passing-Bablok regression and Bland-Altman plots were used for method comparison. In order to categorise the obtained values, results were categorised into the following groups: 0-25 nmol/L, 25-50 nmol/L, 50-75 nmol/L and above 75 nmol/L and compared. The percentage of samples below 75 nmol/L, and below 50 nmol/L was then calculated for every method. RESULTS: According to paired comparisons, each method differs from others (p<0.0001) except Cobas vs Architect, which do not show a statistically significant difference (p=0.39). The strongest correlation was found between Liaison and Architect (ρ=0.94, p<0.0001). The percentage of samples below the recommended value of 75 nmol/L were: 70% (Architect), 92% (Liaison), 71% (Cobas) and 89% (Unicel). The percentage of samples below the value of 50 nmol/L were: 17% (Architect), 55% (Liaison), 28% (Cobas) and 47% (Unicel). CONCLUSIONS: The observed differences stem from the use of different analytical systems for 25-OHD concentration analysis and can result in different outcomes. The recommended values should be established for each assay in accordance with the data provided by the manufacturer or in the laboratory, in accordance with proper standardisation.
format Online
Article
Text
id pubmed-8199437
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Society of Medical Biochemists of Serbia, Belgrade
record_format MEDLINE/PubMed
spelling pubmed-81994372021-06-24 Comparison of four routinely used vitamin D automated immunoassays Windrichova, Jindra Broz, Pavel Fuchsova, Radka Topolcan, Ondrej Pecen, Ladislav Mayer, Otto Kucera, Radek J Med Biochem Original Paper BACKGROUND: To compare four automated immunoassays for the measurement of 25(OH)-vitamin D (25-OHD) and to assess the impact on the results obtained from a healthy population. METHODS: We analysed 100 serum samples on Unicel DxI 800 (Beckman Coulter), Architect i1000 (Abbott), Cobas e411 (Roche) and Liaison XL (DiaSorin). Passing-Bablok regression and Bland-Altman plots were used for method comparison. In order to categorise the obtained values, results were categorised into the following groups: 0-25 nmol/L, 25-50 nmol/L, 50-75 nmol/L and above 75 nmol/L and compared. The percentage of samples below 75 nmol/L, and below 50 nmol/L was then calculated for every method. RESULTS: According to paired comparisons, each method differs from others (p<0.0001) except Cobas vs Architect, which do not show a statistically significant difference (p=0.39). The strongest correlation was found between Liaison and Architect (ρ=0.94, p<0.0001). The percentage of samples below the recommended value of 75 nmol/L were: 70% (Architect), 92% (Liaison), 71% (Cobas) and 89% (Unicel). The percentage of samples below the value of 50 nmol/L were: 17% (Architect), 55% (Liaison), 28% (Cobas) and 47% (Unicel). CONCLUSIONS: The observed differences stem from the use of different analytical systems for 25-OHD concentration analysis and can result in different outcomes. The recommended values should be established for each assay in accordance with the data provided by the manufacturer or in the laboratory, in accordance with proper standardisation. Society of Medical Biochemists of Serbia, Belgrade 2021-06-05 2021-06-05 /pmc/articles/PMC8199437/ /pubmed/34177372 http://dx.doi.org/10.5937/jomb0-27531 Text en 2021 Jindra Windrichova, Pavel Broz, Radka Fuchsova, Ondrej Topolcan, Ladislav Pecen, Otto Mayer, Radek Kucera, published by CEON/CEES https://creativecommons.org/licenses/by/4.0/This work is licensed under the Creative Commons Attribution 4.0 License.
spellingShingle Original Paper
Windrichova, Jindra
Broz, Pavel
Fuchsova, Radka
Topolcan, Ondrej
Pecen, Ladislav
Mayer, Otto
Kucera, Radek
Comparison of four routinely used vitamin D automated immunoassays
title Comparison of four routinely used vitamin D automated immunoassays
title_full Comparison of four routinely used vitamin D automated immunoassays
title_fullStr Comparison of four routinely used vitamin D automated immunoassays
title_full_unstemmed Comparison of four routinely used vitamin D automated immunoassays
title_short Comparison of four routinely used vitamin D automated immunoassays
title_sort comparison of four routinely used vitamin d automated immunoassays
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8199437/
https://www.ncbi.nlm.nih.gov/pubmed/34177372
http://dx.doi.org/10.5937/jomb0-27531
work_keys_str_mv AT windrichovajindra comparisonoffourroutinelyusedvitamindautomatedimmunoassays
AT brozpavel comparisonoffourroutinelyusedvitamindautomatedimmunoassays
AT fuchsovaradka comparisonoffourroutinelyusedvitamindautomatedimmunoassays
AT topolcanondrej comparisonoffourroutinelyusedvitamindautomatedimmunoassays
AT pecenladislav comparisonoffourroutinelyusedvitamindautomatedimmunoassays
AT mayerotto comparisonoffourroutinelyusedvitamindautomatedimmunoassays
AT kuceraradek comparisonoffourroutinelyusedvitamindautomatedimmunoassays