Method comparison of Particle Enhanced Immunoturbidimetry (PEIT) with High Performance Liquid Chromatography (HPLC) for glycated hemoglobin (HbA1c) analysis
BACKGROUND AND OBJECTIVE: High Performance Liquid Chromatography (HPLC) technique is considered as a gold standard for HbA1c analysis however all laboratories cannot adopt it due to certain limitations. Our aim was to compare Particle Enhanced Immunoturbidimetry (PEIT) method with High Performance L...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8201666/ https://www.ncbi.nlm.nih.gov/pubmed/34120656 http://dx.doi.org/10.1186/s40842-021-00123-w |
_version_ | 1783707847370997760 |
---|---|
author | Dildar, Shabnam Imran, Sheharbano Naz, Farah |
author_facet | Dildar, Shabnam Imran, Sheharbano Naz, Farah |
author_sort | Dildar, Shabnam |
collection | PubMed |
description | BACKGROUND AND OBJECTIVE: High Performance Liquid Chromatography (HPLC) technique is considered as a gold standard for HbA1c analysis however all laboratories cannot adopt it due to certain limitations. Our aim was to compare Particle Enhanced Immunoturbidimetry (PEIT) method with High Performance Liquid Chromatography (HPLC) for HbA1c analysis. METHOD: All blood samples were analyzed by HPLC assay on a Bio-Rad D-10 analyzer and PEIT on an Erba XL-200 analyzer. Precision studies were undertaken and Coefficient of Variation (%CV) calculated. Systemic Error (SE), Random Error (RE) and Total Error (TE(calc)) were obtained. The Total Allowable Error (TEa) set by the National Glycohemoglobin Standardization Program (NGSP) for HbA1c is 6%.The acceptable evaluation method is where TE(calc) is less than TE(a.) RESULTS: The Precision studies were satisfactory with Coefficient of Variation (%CV) being less than 4% for both techniques. Mean HbA1c levels were slightly higher from HPLC than PEIT 9.07 ± 2.23% and 8.93 ± 2.10% respectively, although the difference was minimal. RE was 1.41%, TE(calc) was 1.55%, which was less than TEa set by the NGSP. Both methods strongly correlated with the correlation coefficient (r) 0.9716, p < 0.0001. CONCLUSION: Our study showed HbA1c analysis by PEIT technique is precise, accurate, rapid and convenient and can be employed as an alternative to HPLC technique in countries where cost is a major problem for diagnosis and treatment. |
format | Online Article Text |
id | pubmed-8201666 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-82016662021-06-15 Method comparison of Particle Enhanced Immunoturbidimetry (PEIT) with High Performance Liquid Chromatography (HPLC) for glycated hemoglobin (HbA1c) analysis Dildar, Shabnam Imran, Sheharbano Naz, Farah Clin Diabetes Endocrinol Research Article BACKGROUND AND OBJECTIVE: High Performance Liquid Chromatography (HPLC) technique is considered as a gold standard for HbA1c analysis however all laboratories cannot adopt it due to certain limitations. Our aim was to compare Particle Enhanced Immunoturbidimetry (PEIT) method with High Performance Liquid Chromatography (HPLC) for HbA1c analysis. METHOD: All blood samples were analyzed by HPLC assay on a Bio-Rad D-10 analyzer and PEIT on an Erba XL-200 analyzer. Precision studies were undertaken and Coefficient of Variation (%CV) calculated. Systemic Error (SE), Random Error (RE) and Total Error (TE(calc)) were obtained. The Total Allowable Error (TEa) set by the National Glycohemoglobin Standardization Program (NGSP) for HbA1c is 6%.The acceptable evaluation method is where TE(calc) is less than TE(a.) RESULTS: The Precision studies were satisfactory with Coefficient of Variation (%CV) being less than 4% for both techniques. Mean HbA1c levels were slightly higher from HPLC than PEIT 9.07 ± 2.23% and 8.93 ± 2.10% respectively, although the difference was minimal. RE was 1.41%, TE(calc) was 1.55%, which was less than TEa set by the NGSP. Both methods strongly correlated with the correlation coefficient (r) 0.9716, p < 0.0001. CONCLUSION: Our study showed HbA1c analysis by PEIT technique is precise, accurate, rapid and convenient and can be employed as an alternative to HPLC technique in countries where cost is a major problem for diagnosis and treatment. BioMed Central 2021-06-13 /pmc/articles/PMC8201666/ /pubmed/34120656 http://dx.doi.org/10.1186/s40842-021-00123-w Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) . The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/ (https://creativecommons.org/publicdomain/zero/1.0/) ) applies to the data made available in this article, unless otherwise stated in a credit line to the data. |
spellingShingle | Research Article Dildar, Shabnam Imran, Sheharbano Naz, Farah Method comparison of Particle Enhanced Immunoturbidimetry (PEIT) with High Performance Liquid Chromatography (HPLC) for glycated hemoglobin (HbA1c) analysis |
title | Method comparison of Particle Enhanced Immunoturbidimetry (PEIT) with High Performance Liquid Chromatography (HPLC) for glycated hemoglobin (HbA1c) analysis |
title_full | Method comparison of Particle Enhanced Immunoturbidimetry (PEIT) with High Performance Liquid Chromatography (HPLC) for glycated hemoglobin (HbA1c) analysis |
title_fullStr | Method comparison of Particle Enhanced Immunoturbidimetry (PEIT) with High Performance Liquid Chromatography (HPLC) for glycated hemoglobin (HbA1c) analysis |
title_full_unstemmed | Method comparison of Particle Enhanced Immunoturbidimetry (PEIT) with High Performance Liquid Chromatography (HPLC) for glycated hemoglobin (HbA1c) analysis |
title_short | Method comparison of Particle Enhanced Immunoturbidimetry (PEIT) with High Performance Liquid Chromatography (HPLC) for glycated hemoglobin (HbA1c) analysis |
title_sort | method comparison of particle enhanced immunoturbidimetry (peit) with high performance liquid chromatography (hplc) for glycated hemoglobin (hba1c) analysis |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8201666/ https://www.ncbi.nlm.nih.gov/pubmed/34120656 http://dx.doi.org/10.1186/s40842-021-00123-w |
work_keys_str_mv | AT dildarshabnam methodcomparisonofparticleenhancedimmunoturbidimetrypeitwithhighperformanceliquidchromatographyhplcforglycatedhemoglobinhba1canalysis AT imransheharbano methodcomparisonofparticleenhancedimmunoturbidimetrypeitwithhighperformanceliquidchromatographyhplcforglycatedhemoglobinhba1canalysis AT nazfarah methodcomparisonofparticleenhancedimmunoturbidimetrypeitwithhighperformanceliquidchromatographyhplcforglycatedhemoglobinhba1canalysis |