Directed graph mapping exceeds phase mapping in discriminating true and false rotors detected with a basket catheter in a complex in-silico excitation pattern

Atrial fibrillation (AF) is the most frequently encountered arrhythmia in clinical practise. One of the major problems in the management of AF is the difficulty in identifying the arrhythmia sources from clinical recordings. That difficulty occurs because it is currently impossible to verify algorit...

Descripción completa

Detalles Bibliográficos
Autores principales: Van Nieuwenhuyse, Enid, Martinez-Mateu, Laura, Saiz, Javier, Panfilov, Alexander V., Vandersickel, Nele
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8204274/
https://www.ncbi.nlm.nih.gov/pubmed/33901713
http://dx.doi.org/10.1016/j.compbiomed.2021.104381
_version_ 1783708319868780544
author Van Nieuwenhuyse, Enid
Martinez-Mateu, Laura
Saiz, Javier
Panfilov, Alexander V.
Vandersickel, Nele
author_facet Van Nieuwenhuyse, Enid
Martinez-Mateu, Laura
Saiz, Javier
Panfilov, Alexander V.
Vandersickel, Nele
author_sort Van Nieuwenhuyse, Enid
collection PubMed
description Atrial fibrillation (AF) is the most frequently encountered arrhythmia in clinical practise. One of the major problems in the management of AF is the difficulty in identifying the arrhythmia sources from clinical recordings. That difficulty occurs because it is currently impossible to verify algorithms which determine these sources in clinical data, as high resolution true excitation patterns cannot be recorded in patients. Therefore, alternative approaches, like computer modelling are of great interest. In a recent published study such an approach was applied for the verification of one of the most commonly used algorithms, phase mapping (PM). A meandering rotor was simulated in the right atrium and a basket catheter was placed at 3 different locations: at the Superior Vena Cava (SVC), the Crista Terminalis (CT) and at the Coronary Sinus (CS). It was shown that although PM can identify the true source, it also finds several false sources due to the far-field effects and interpolation errors in all three positions. In addition, the detection efficiency strongly depended on the basket location. Recently, a novel tool was developed to analyse any arrhythmia called Directed Graph Mapping (DGM). DGM is based on network theory and creates a directed graph of the excitation pattern, from which the location and the source of the arrhythmia can be detected. Therefore, the objective of the current study was to compare the efficiency of DGM with PM on the basket dataset of this meandering rotor. The DGM-tool was applied for a wide variety of conduction velocities (minimal and maximal), which are input parameters of DGM. Overall we found that DGM was able to distinguish between the true rotor and false rotors for both the SVC and CT basket positions. For example, for the SVC position with a [Formula: see text] , DGM detected the true core with a prevalence of 82% versus 94% for PM. Three false rotors where detected for 39.16% (DGM) versus 100% (PM); 22.64% (DGM) versus 100% (PM); and 0% (DGM) versus 57% (PM). Increasing [Formula: see text] to [Formula: see text] had a stronger effect on the false rotors than on the true rotor. This led to a detection rate of [Formula: see text] for the true rotor, while all the other false rotors disappeared. A similar trend was observed for the CT position. For the CS position, DGM already had a low performance for the true rotor for [Formula: see text] (14.7%). For [Formula: see text] the false and the true rotors could therefore not be distinguished. We can conclude that DGM can overcome some of the limitations of PM by varying one of its input parameters ([Formula: see text]). The true rotor is less dependent on this parameter than the false rotors, which disappear at a [Formula: see text]. In order to increase to detection rate of the true rotor, one can decrease [Formula: see text] and discard the new rotors which also appear at lower values of [Formula: see text].
format Online
Article
Text
id pubmed-8204274
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-82042742021-06-21 Directed graph mapping exceeds phase mapping in discriminating true and false rotors detected with a basket catheter in a complex in-silico excitation pattern Van Nieuwenhuyse, Enid Martinez-Mateu, Laura Saiz, Javier Panfilov, Alexander V. Vandersickel, Nele Comput Biol Med Article Atrial fibrillation (AF) is the most frequently encountered arrhythmia in clinical practise. One of the major problems in the management of AF is the difficulty in identifying the arrhythmia sources from clinical recordings. That difficulty occurs because it is currently impossible to verify algorithms which determine these sources in clinical data, as high resolution true excitation patterns cannot be recorded in patients. Therefore, alternative approaches, like computer modelling are of great interest. In a recent published study such an approach was applied for the verification of one of the most commonly used algorithms, phase mapping (PM). A meandering rotor was simulated in the right atrium and a basket catheter was placed at 3 different locations: at the Superior Vena Cava (SVC), the Crista Terminalis (CT) and at the Coronary Sinus (CS). It was shown that although PM can identify the true source, it also finds several false sources due to the far-field effects and interpolation errors in all three positions. In addition, the detection efficiency strongly depended on the basket location. Recently, a novel tool was developed to analyse any arrhythmia called Directed Graph Mapping (DGM). DGM is based on network theory and creates a directed graph of the excitation pattern, from which the location and the source of the arrhythmia can be detected. Therefore, the objective of the current study was to compare the efficiency of DGM with PM on the basket dataset of this meandering rotor. The DGM-tool was applied for a wide variety of conduction velocities (minimal and maximal), which are input parameters of DGM. Overall we found that DGM was able to distinguish between the true rotor and false rotors for both the SVC and CT basket positions. For example, for the SVC position with a [Formula: see text] , DGM detected the true core with a prevalence of 82% versus 94% for PM. Three false rotors where detected for 39.16% (DGM) versus 100% (PM); 22.64% (DGM) versus 100% (PM); and 0% (DGM) versus 57% (PM). Increasing [Formula: see text] to [Formula: see text] had a stronger effect on the false rotors than on the true rotor. This led to a detection rate of [Formula: see text] for the true rotor, while all the other false rotors disappeared. A similar trend was observed for the CT position. For the CS position, DGM already had a low performance for the true rotor for [Formula: see text] (14.7%). For [Formula: see text] the false and the true rotors could therefore not be distinguished. We can conclude that DGM can overcome some of the limitations of PM by varying one of its input parameters ([Formula: see text]). The true rotor is less dependent on this parameter than the false rotors, which disappear at a [Formula: see text]. In order to increase to detection rate of the true rotor, one can decrease [Formula: see text] and discard the new rotors which also appear at lower values of [Formula: see text]. Elsevier 2021-06 /pmc/articles/PMC8204274/ /pubmed/33901713 http://dx.doi.org/10.1016/j.compbiomed.2021.104381 Text en © 2021 The Author(s) https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Article
Van Nieuwenhuyse, Enid
Martinez-Mateu, Laura
Saiz, Javier
Panfilov, Alexander V.
Vandersickel, Nele
Directed graph mapping exceeds phase mapping in discriminating true and false rotors detected with a basket catheter in a complex in-silico excitation pattern
title Directed graph mapping exceeds phase mapping in discriminating true and false rotors detected with a basket catheter in a complex in-silico excitation pattern
title_full Directed graph mapping exceeds phase mapping in discriminating true and false rotors detected with a basket catheter in a complex in-silico excitation pattern
title_fullStr Directed graph mapping exceeds phase mapping in discriminating true and false rotors detected with a basket catheter in a complex in-silico excitation pattern
title_full_unstemmed Directed graph mapping exceeds phase mapping in discriminating true and false rotors detected with a basket catheter in a complex in-silico excitation pattern
title_short Directed graph mapping exceeds phase mapping in discriminating true and false rotors detected with a basket catheter in a complex in-silico excitation pattern
title_sort directed graph mapping exceeds phase mapping in discriminating true and false rotors detected with a basket catheter in a complex in-silico excitation pattern
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8204274/
https://www.ncbi.nlm.nih.gov/pubmed/33901713
http://dx.doi.org/10.1016/j.compbiomed.2021.104381
work_keys_str_mv AT vannieuwenhuyseenid directedgraphmappingexceedsphasemappingindiscriminatingtrueandfalserotorsdetectedwithabasketcatheterinacomplexinsilicoexcitationpattern
AT martinezmateulaura directedgraphmappingexceedsphasemappingindiscriminatingtrueandfalserotorsdetectedwithabasketcatheterinacomplexinsilicoexcitationpattern
AT saizjavier directedgraphmappingexceedsphasemappingindiscriminatingtrueandfalserotorsdetectedwithabasketcatheterinacomplexinsilicoexcitationpattern
AT panfilovalexanderv directedgraphmappingexceedsphasemappingindiscriminatingtrueandfalserotorsdetectedwithabasketcatheterinacomplexinsilicoexcitationpattern
AT vandersickelnele directedgraphmappingexceedsphasemappingindiscriminatingtrueandfalserotorsdetectedwithabasketcatheterinacomplexinsilicoexcitationpattern