3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT

Detector saturation in cone-beam computed tomography occurs when an object of highly varying shape and material composition is imaged using an automatic exposure control (AEC) system. When imaging a subject’s knees, high beam energy ensures the visibility of internal structures but leads to overexpo...

Descripción completa

Detalles Bibliográficos
Autores principales: MAIER, JENNIFER, MAIER, ANDREAS, ESKOFIER, BJOERN, FAHRIG, REBECCA, CHOI, JANG-HWAN
Formato: Online Artículo Texto
Lenguaje:English
Publicado: 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8208599/
https://www.ncbi.nlm.nih.gov/pubmed/34141516
http://dx.doi.org/10.1109/access.2021.3079368
_version_ 1783708956673179648
author MAIER, JENNIFER
MAIER, ANDREAS
ESKOFIER, BJOERN
FAHRIG, REBECCA
CHOI, JANG-HWAN
author_facet MAIER, JENNIFER
MAIER, ANDREAS
ESKOFIER, BJOERN
FAHRIG, REBECCA
CHOI, JANG-HWAN
author_sort MAIER, JENNIFER
collection PubMed
description Detector saturation in cone-beam computed tomography occurs when an object of highly varying shape and material composition is imaged using an automatic exposure control (AEC) system. When imaging a subject’s knees, high beam energy ensures the visibility of internal structures but leads to overexposure in less dense border regions. In this work, we propose to use an additional low-dose scan to correct the saturation artifacts of AEC scans. Overexposed pixels are identified in the projection images of the AEC scan using histogram-based thresholding. The saturation-free pixels from the AEC scan are combined with the skin border pixels of the low-dose scan prior to volumetric reconstruction. To compensate for patient motion between the two scans, a 3D non-rigid alignment of the projection images in a backward-forward-projection process based on fiducial marker positions is proposed. On numerical simulations, the projection combination improved the structural similarity index measure from 0.883 to 0.999. Further evaluations were performed on two in vivo subject knee acquisitions, one without and one with motion between the AEC and low-dose scans. Saturation-free reference images were acquired using a beam attenuator. The proposed method could qualitatively restore the information of peripheral tissue structures. Applying the 3D non-rigid alignment made it possible to use the projection images with inter-scan subject motion for projection image combination. The increase in radiation exposure due to the additional low-dose scan was found to be negligibly low. The presented methods allow simple but effective correction of saturation artifacts.
format Online
Article
Text
id pubmed-8208599
institution National Center for Biotechnology Information
language English
publishDate 2021
record_format MEDLINE/PubMed
spelling pubmed-82085992021-06-16 3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT MAIER, JENNIFER MAIER, ANDREAS ESKOFIER, BJOERN FAHRIG, REBECCA CHOI, JANG-HWAN IEEE Access Article Detector saturation in cone-beam computed tomography occurs when an object of highly varying shape and material composition is imaged using an automatic exposure control (AEC) system. When imaging a subject’s knees, high beam energy ensures the visibility of internal structures but leads to overexposure in less dense border regions. In this work, we propose to use an additional low-dose scan to correct the saturation artifacts of AEC scans. Overexposed pixels are identified in the projection images of the AEC scan using histogram-based thresholding. The saturation-free pixels from the AEC scan are combined with the skin border pixels of the low-dose scan prior to volumetric reconstruction. To compensate for patient motion between the two scans, a 3D non-rigid alignment of the projection images in a backward-forward-projection process based on fiducial marker positions is proposed. On numerical simulations, the projection combination improved the structural similarity index measure from 0.883 to 0.999. Further evaluations were performed on two in vivo subject knee acquisitions, one without and one with motion between the AEC and low-dose scans. Saturation-free reference images were acquired using a beam attenuator. The proposed method could qualitatively restore the information of peripheral tissue structures. Applying the 3D non-rigid alignment made it possible to use the projection images with inter-scan subject motion for projection image combination. The increase in radiation exposure due to the additional low-dose scan was found to be negligibly low. The presented methods allow simple but effective correction of saturation artifacts. 2021-05-11 2021 /pmc/articles/PMC8208599/ /pubmed/34141516 http://dx.doi.org/10.1109/access.2021.3079368 Text en https://creativecommons.org/licenses/by/4.0/This work is licensed under a Creative Commons Attribution 4.0 License. For more information, see https://creativecommons.org/licenses/by/4.0/
spellingShingle Article
MAIER, JENNIFER
MAIER, ANDREAS
ESKOFIER, BJOERN
FAHRIG, REBECCA
CHOI, JANG-HWAN
3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT
title 3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT
title_full 3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT
title_fullStr 3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT
title_full_unstemmed 3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT
title_short 3D Non-Rigid Alignment of Low-Dose Scans Allows to Correct for Saturation in Lower Extremity Cone-Beam CT
title_sort 3d non-rigid alignment of low-dose scans allows to correct for saturation in lower extremity cone-beam ct
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8208599/
https://www.ncbi.nlm.nih.gov/pubmed/34141516
http://dx.doi.org/10.1109/access.2021.3079368
work_keys_str_mv AT maierjennifer 3dnonrigidalignmentoflowdosescansallowstocorrectforsaturationinlowerextremityconebeamct
AT maierandreas 3dnonrigidalignmentoflowdosescansallowstocorrectforsaturationinlowerextremityconebeamct
AT eskofierbjoern 3dnonrigidalignmentoflowdosescansallowstocorrectforsaturationinlowerextremityconebeamct
AT fahrigrebecca 3dnonrigidalignmentoflowdosescansallowstocorrectforsaturationinlowerextremityconebeamct
AT choijanghwan 3dnonrigidalignmentoflowdosescansallowstocorrectforsaturationinlowerextremityconebeamct