New infusion device for use in acquisition of images during endovascular procedures: an experimental model
BACKGROUND: The contrast power injector (CPI) is the gold standard method for injecting contrast with the pressure and flow needed to generate a satisfactory images during endovascular procedures, but it is an expensive tool, narrowing its wide-scale applications. One alternative is the manual injec...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Sociedade Brasileira de Angiologia e de Cirurgia Vascular (SBACV)
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8210644/ https://www.ncbi.nlm.nih.gov/pubmed/34188669 http://dx.doi.org/10.1590/1677-5449.200191 |
Sumario: | BACKGROUND: The contrast power injector (CPI) is the gold standard method for injecting contrast with the pressure and flow needed to generate a satisfactory images during endovascular procedures, but it is an expensive tool, narrowing its wide-scale applications. One alternative is the manual injection (MI) method, but this does not generate the pressure required for adequate visualization of anatomy. It is therefore imperative to create an alternative low-cost method that is capable of producing high quality images. OBJECTIVES: To compare the injection parameters of a new mechanical device (Hand-Crank) created in a university hospital with the MI method and with the contrast power injector’s ideal values. METHODS: A circulation phantom was constructed to simulate the pressure in the aorto-iliac territory and the injection parameters of the two methods were compared in a laboratory setting. Student’s t test and the Mann-Whitney test were used for statistical analysis. Three vascular surgery residents (the authors) performed the injections (each performed 9 tests using conventional manual injection and 9 tests using the Hand-Crank, totaling 54 injections). RESULTS: There were statistical differences between the two methods (p<0.05) in total volume injected until maximum pressure was attained, pressure variation, maximum pressure, total injection time, and time to reach the maximum pressure. CONCLUSIONS: The Hand-Crank can achieve higher maximum pressure, higher average flow, and lower injection time than the manual method. It is a simple, low-cost, and effective tool for enhancing injection parameters in an experimental setup. It could help to produce higher quality images in a clinical scenario. |
---|