Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins

OBJECTIVES: The purpose of this study was to compare data obtained from the reticulocyte channel (RET channel) heated to 41 °C with those obtained from impedance channel (I-Channel) at room temperature in the samples with the mean corpuscular hemoglobin concentration (MCHC) < 370 g/L and in sampl...

Descripción completa

Detalles Bibliográficos
Autores principales: Roccaforte, Vincenzo, Sciarini, Flavia, Proserpio, Vanessa, Buonocore, Ruggero, Zavaroni, Emanuela Marina, Burati, Silvia, Bussetti, Marco, Liuzzi, Giammaria, Russo, Rosalba Monica, Porreca, Wanda Patrizia, Angelis, Maria Luisa De, Perno, Carlo Federico, Bonato, Claudio, Pastori, Stefano
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Sociedade Brasileira de Hematologia e Hemoterapia 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8211624/
https://www.ncbi.nlm.nih.gov/pubmed/32199923
http://dx.doi.org/10.1016/j.htct.2020.01.004
_version_ 1783709504186089472
author Roccaforte, Vincenzo
Sciarini, Flavia
Proserpio, Vanessa
Buonocore, Ruggero
Zavaroni, Emanuela Marina
Burati, Silvia
Bussetti, Marco
Liuzzi, Giammaria
Russo, Rosalba Monica
Porreca, Wanda Patrizia
Angelis, Maria Luisa De
Perno, Carlo Federico
Bonato, Claudio
Pastori, Stefano
author_facet Roccaforte, Vincenzo
Sciarini, Flavia
Proserpio, Vanessa
Buonocore, Ruggero
Zavaroni, Emanuela Marina
Burati, Silvia
Bussetti, Marco
Liuzzi, Giammaria
Russo, Rosalba Monica
Porreca, Wanda Patrizia
Angelis, Maria Luisa De
Perno, Carlo Federico
Bonato, Claudio
Pastori, Stefano
author_sort Roccaforte, Vincenzo
collection PubMed
description OBJECTIVES: The purpose of this study was to compare data obtained from the reticulocyte channel (RET channel) heated to 41 °C with those obtained from impedance channel (I-Channel) at room temperature in the samples with the mean corpuscular hemoglobin concentration (MCHC) < 370 g/L and in samples with the MCHC > 370 g/L, in the presence of cold agglutinins. METHODS: In this study, 60 blood samples (group 1) with the MCHC < 370 g/L (without cold agglutinins) and 78 blood samples (group 2) with the MCHC > 370 g/L (with cold agglutinins) were used to compare the two analytical channels of the XN-9000 analyzer in different preanalytical conditions. The parameters evaluated in both groups were the following: red blood cell (RBC), hemoglobin (HGB), hematocrit (HCT), mean cell volume (MCV), RBC-most frequent volume (R-MFV), mean hemoglobin concentration (MCH) and mean cellular hemoglobin concentration (MCHC). RESULTS: The results of this study showed an excellent correlation with both channels of the XN-9000 analyzer in samples with and without cold agglutinins, except for the MCHC. The bias between the values obtained in the I-channel and those obtained in the RET channel of both groups was insignificant and remained within the limits of acceptability, as reported by Ricos et al. for all considered parameters, except for MCHC. CONCLUSIONS: The presence of cold agglutinins in blood samples can be detected by a spurious lowering of the RBC count and by a spurious increase in the MCHC. The RET channel represents a great opportunity to correct the RBC count in a rapid manner without preheating. However, neither methodology can completely solve the residual presence of cold agglutinins in all samples, despite the MCHC values being < 370 g/L.
format Online
Article
Text
id pubmed-8211624
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Sociedade Brasileira de Hematologia e Hemoterapia
record_format MEDLINE/PubMed
spelling pubmed-82116242021-06-25 Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins Roccaforte, Vincenzo Sciarini, Flavia Proserpio, Vanessa Buonocore, Ruggero Zavaroni, Emanuela Marina Burati, Silvia Bussetti, Marco Liuzzi, Giammaria Russo, Rosalba Monica Porreca, Wanda Patrizia Angelis, Maria Luisa De Perno, Carlo Federico Bonato, Claudio Pastori, Stefano Hematol Transfus Cell Ther Original Article OBJECTIVES: The purpose of this study was to compare data obtained from the reticulocyte channel (RET channel) heated to 41 °C with those obtained from impedance channel (I-Channel) at room temperature in the samples with the mean corpuscular hemoglobin concentration (MCHC) < 370 g/L and in samples with the MCHC > 370 g/L, in the presence of cold agglutinins. METHODS: In this study, 60 blood samples (group 1) with the MCHC < 370 g/L (without cold agglutinins) and 78 blood samples (group 2) with the MCHC > 370 g/L (with cold agglutinins) were used to compare the two analytical channels of the XN-9000 analyzer in different preanalytical conditions. The parameters evaluated in both groups were the following: red blood cell (RBC), hemoglobin (HGB), hematocrit (HCT), mean cell volume (MCV), RBC-most frequent volume (R-MFV), mean hemoglobin concentration (MCH) and mean cellular hemoglobin concentration (MCHC). RESULTS: The results of this study showed an excellent correlation with both channels of the XN-9000 analyzer in samples with and without cold agglutinins, except for the MCHC. The bias between the values obtained in the I-channel and those obtained in the RET channel of both groups was insignificant and remained within the limits of acceptability, as reported by Ricos et al. for all considered parameters, except for MCHC. CONCLUSIONS: The presence of cold agglutinins in blood samples can be detected by a spurious lowering of the RBC count and by a spurious increase in the MCHC. The RET channel represents a great opportunity to correct the RBC count in a rapid manner without preheating. However, neither methodology can completely solve the residual presence of cold agglutinins in all samples, despite the MCHC values being < 370 g/L. Sociedade Brasileira de Hematologia e Hemoterapia 2021 2020-03-06 /pmc/articles/PMC8211624/ /pubmed/32199923 http://dx.doi.org/10.1016/j.htct.2020.01.004 Text en © 2020 Associação Brasileira de Hematologia, Hemoterapia e Terapia Celular. Published by Elsevier Editora Ltda. https://creativecommons.org/licenses/by-nc-nd/4.0/This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Original Article
Roccaforte, Vincenzo
Sciarini, Flavia
Proserpio, Vanessa
Buonocore, Ruggero
Zavaroni, Emanuela Marina
Burati, Silvia
Bussetti, Marco
Liuzzi, Giammaria
Russo, Rosalba Monica
Porreca, Wanda Patrizia
Angelis, Maria Luisa De
Perno, Carlo Federico
Bonato, Claudio
Pastori, Stefano
Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_full Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_fullStr Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_full_unstemmed Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_short Use of the reticulocyte channel warmed to 41 °C of the XN-9000 analyzer in samples with the presence of cold agglutinins
title_sort use of the reticulocyte channel warmed to 41 °c of the xn-9000 analyzer in samples with the presence of cold agglutinins
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8211624/
https://www.ncbi.nlm.nih.gov/pubmed/32199923
http://dx.doi.org/10.1016/j.htct.2020.01.004
work_keys_str_mv AT roccafortevincenzo useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT sciariniflavia useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT proserpiovanessa useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT buonocoreruggero useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT zavaroniemanuelamarina useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT buratisilvia useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT bussettimarco useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT liuzzigiammaria useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT russorosalbamonica useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT porrecawandapatrizia useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT angelismarialuisade useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT pernocarlofederico useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT bonatoclaudio useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins
AT pastoristefano useofthereticulocytechannelwarmedto41cofthexn9000analyzerinsampleswiththepresenceofcoldagglutinins