Development and Evaluation of a Point-of-Care Test in a Low-Resource Setting with High Rates of Chlamydia trachomatis Urogenital Infections in Fiji
Rapid and precise detection of Chlamydia trachomatis, the leading global cause of sexually transmitted infections (STI), at the point of care (POC) is required for treatment decisions to prevent transmission and sequelae, including pelvic inflammatory disease, ectopic pregnancy, tubal factor inferti...
Autores principales: | , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
American Society for Microbiology
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8218753/ https://www.ncbi.nlm.nih.gov/pubmed/33910964 http://dx.doi.org/10.1128/JCM.00182-21 |
Sumario: | Rapid and precise detection of Chlamydia trachomatis, the leading global cause of sexually transmitted infections (STI), at the point of care (POC) is required for treatment decisions to prevent transmission and sequelae, including pelvic inflammatory disease, ectopic pregnancy, tubal factor infertility, and preterm birth. We developed a rapid POC test (POCT), termed LH-POCT, which uses loop-mediated amplification (LAMP) of nucleic acids. We performed a head-to-head comparison with the Cepheid Xpert CT/NG assay using clinician-collected, deidentified paired vaginal samples from a parent study that consecutively enrolled symptomatic and asymptomatic females over 18 years of age from the Ministry of Health and Medical Services Health Centers in Fiji. Samples were processed by the Xpert CT/NG assay and LH-POCT, blinded to the comparator. Discrepant samples were resolved by quantitative PCR. Deidentified clinical data and tests for Trichomonas vaginalis, Candida, and bacterial vaginosis (BV) were provided. There were a total of 353 samples from 327 females. C. trachomatis positivity was 16.7% (59/353), while the prevalence was 16.82% (55/327) after discrepant resolution. Seven discrepant samples resolved to four false negatives, two false positives, and one true positive for the LH-POCT. The sensitivity of the LH-POCT was 93.65% (95% confidence interval [CI], 84.53% to 98.24%), and specificity was 99.31% (95% CI, 97.53% to 99.92%). Discrepant samples clustered among women with vaginal discharge and/or BV. The prototype LH-POCT workflow has excellent performance, meeting many World Health Organization ASSURED criteria for POC tests, including a sample-to-result time of 35 min. Our LH-POCT holds promise for improving clinical practice to prevent and control C. trachomatis STIs in diverse health care settings globally. |
---|