Ultrasound Shear Wave Elastography and Transient Optical Coherence Elastography: Side-by-Side Comparison of Repeatability and Accuracy
Objective: We compare the repeatability and accuracy of ultrasound shear wave elastography (USE) and transient optical coherence elastography (OCE). Methods: Elastic wave speed in gelatin phantoms and chicken breast was measured with USE and OCE and compared with uniaxial mechanical compression test...
Formato: | Online Artículo Texto |
---|---|
Lenguaje: | English |
Publicado: |
IEEE
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8224461/ https://www.ncbi.nlm.nih.gov/pubmed/34179823 http://dx.doi.org/10.1109/OJEMB.2021.3075569 |
_version_ | 1783711894372089856 |
---|---|
collection | PubMed |
description | Objective: We compare the repeatability and accuracy of ultrasound shear wave elastography (USE) and transient optical coherence elastography (OCE). Methods: Elastic wave speed in gelatin phantoms and chicken breast was measured with USE and OCE and compared with uniaxial mechanical compression testing. Intra- and Inter-repeatability were analyzed using Bland-Altman plots and intraclass correlation coefficients (ICC). Results: OCE and USE differed from uniaxial testing by a mean absolute percent error of 8.92% and 16.9%, respectively, across eight phantoms of varying stiffness. Upper and lower limits of agreement for intrasample repeatability for USE and OCE were ±0.075 m/s and −0.14 m/s and 0.13 m/s, respectively. OCE and USE both had ICCs of 0.9991. In chicken breast, ICC for USE was 0.9385 and for OCE was 0.9924. Conclusion: OCE and USE can detect small speed changes and give comparable measurements. These measurements correspond well with uniaxial testing. |
format | Online Article Text |
id | pubmed-8224461 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | IEEE |
record_format | MEDLINE/PubMed |
spelling | pubmed-82244612021-06-24 Ultrasound Shear Wave Elastography and Transient Optical Coherence Elastography: Side-by-Side Comparison of Repeatability and Accuracy IEEE Open J Eng Med Biol Article Objective: We compare the repeatability and accuracy of ultrasound shear wave elastography (USE) and transient optical coherence elastography (OCE). Methods: Elastic wave speed in gelatin phantoms and chicken breast was measured with USE and OCE and compared with uniaxial mechanical compression testing. Intra- and Inter-repeatability were analyzed using Bland-Altman plots and intraclass correlation coefficients (ICC). Results: OCE and USE differed from uniaxial testing by a mean absolute percent error of 8.92% and 16.9%, respectively, across eight phantoms of varying stiffness. Upper and lower limits of agreement for intrasample repeatability for USE and OCE were ±0.075 m/s and −0.14 m/s and 0.13 m/s, respectively. OCE and USE both had ICCs of 0.9991. In chicken breast, ICC for USE was 0.9385 and for OCE was 0.9924. Conclusion: OCE and USE can detect small speed changes and give comparable measurements. These measurements correspond well with uniaxial testing. IEEE 2021-04-27 /pmc/articles/PMC8224461/ /pubmed/34179823 http://dx.doi.org/10.1109/OJEMB.2021.3075569 Text en https://creativecommons.org/licenses/by/4.0/This work is licensed under a Creative Commons Attribution 4.0 License. For more information, see https://creativecommons.org/licenses/by/4.0/ |
spellingShingle | Article Ultrasound Shear Wave Elastography and Transient Optical Coherence Elastography: Side-by-Side Comparison of Repeatability and Accuracy |
title | Ultrasound Shear Wave Elastography and Transient Optical Coherence Elastography: Side-by-Side Comparison of Repeatability and Accuracy |
title_full | Ultrasound Shear Wave Elastography and Transient Optical Coherence Elastography: Side-by-Side Comparison of Repeatability and Accuracy |
title_fullStr | Ultrasound Shear Wave Elastography and Transient Optical Coherence Elastography: Side-by-Side Comparison of Repeatability and Accuracy |
title_full_unstemmed | Ultrasound Shear Wave Elastography and Transient Optical Coherence Elastography: Side-by-Side Comparison of Repeatability and Accuracy |
title_short | Ultrasound Shear Wave Elastography and Transient Optical Coherence Elastography: Side-by-Side Comparison of Repeatability and Accuracy |
title_sort | ultrasound shear wave elastography and transient optical coherence elastography: side-by-side comparison of repeatability and accuracy |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8224461/ https://www.ncbi.nlm.nih.gov/pubmed/34179823 http://dx.doi.org/10.1109/OJEMB.2021.3075569 |
work_keys_str_mv | AT ultrasoundshearwaveelastographyandtransientopticalcoherenceelastographysidebysidecomparisonofrepeatabilityandaccuracy AT ultrasoundshearwaveelastographyandtransientopticalcoherenceelastographysidebysidecomparisonofrepeatabilityandaccuracy AT ultrasoundshearwaveelastographyandtransientopticalcoherenceelastographysidebysidecomparisonofrepeatabilityandaccuracy AT ultrasoundshearwaveelastographyandtransientopticalcoherenceelastographysidebysidecomparisonofrepeatabilityandaccuracy |