Performance of the LIAISON(®) SARS-CoV-2 Antigen Assay vs. SARS-CoV-2-RT-PCR
We aimed to evaluate the LIAISON(®) SARS-CoV-2 antigen assay (DiaSorin), comparing its performance to real-time polymerase chain reaction (RT-PCR) for the detection of SARS-CoV-2 RNA. 182 (110 PCR-positive and 72 PCR-negative) nasopharyngeal swab samples were taken for the detection of SARS-CoV-2. R...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8228578/ https://www.ncbi.nlm.nih.gov/pubmed/34073618 http://dx.doi.org/10.3390/pathogens10060658 |
Sumario: | We aimed to evaluate the LIAISON(®) SARS-CoV-2 antigen assay (DiaSorin), comparing its performance to real-time polymerase chain reaction (RT-PCR) for the detection of SARS-CoV-2 RNA. 182 (110 PCR-positive and 72 PCR-negative) nasopharyngeal swab samples were taken for the detection of SARS-CoV-2. RT-PCR and antigen assay were performed using the same material. The sensitivity and specificity of the antigen assay were calculated for different cut-offs, with RT-PCR serving as the reference method. Stored clinical samples that were positive for other respiratory viruses were tested to evaluate cross-reactivity. One third (33/110, 30%) were falsely classified as negative, while no false positives were found using the 200 TCID(50)/mL cut-off for the SARS-CoV-2 antigen as proposed by the manufacturer. This corresponded to a sensitivity of 70% (60–78%) and a specificity of 100% (94–100%). Lowering the cut-off for positivity of the antigen assay to 22.79 or 57.68 TCID(50)/mL increased the sensitivity of the method, reaching a sensitivity of 92% (85–96%) vs. 79% (70–86%) and a specificity of 81% (69–89%) vs. 99% (91–100%), respectively. The antigen assay reliably detected samples with high SARS-CoV-2 viral loads (≥10(6) copies SARS-CoV-2/mL), while it cannot differentiate between negative and low positive samples. Cross-reactivity toward other respiratory viruses was not detected. |
---|