Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns

RATIONALE: Mixed‐mode reversed‐phase/anion exchange liquid chromatography is useful for separations of mixtures containing anions (e.g. ionized acids). However, when using this form of liquid chromatography with mass spectrometry detection, the bleed of amine‐containing hydrolysis products from the...

Descripción completa

Detalles Bibliográficos
Autores principales: Walter, Thomas H., Blaze M. T., Melvin, Boissel, Cheryl
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8244038/
https://www.ncbi.nlm.nih.gov/pubmed/33830546
http://dx.doi.org/10.1002/rcm.9098
_version_ 1783715851993612288
author Walter, Thomas H.
Blaze M. T., Melvin
Boissel, Cheryl
author_facet Walter, Thomas H.
Blaze M. T., Melvin
Boissel, Cheryl
author_sort Walter, Thomas H.
collection PubMed
description RATIONALE: Mixed‐mode reversed‐phase/anion exchange liquid chromatography is useful for separations of mixtures containing anions (e.g. ionized acids). However, when using this form of liquid chromatography with mass spectrometry detection, the bleed of amine‐containing hydrolysis products from the columns may cause ion suppression or enhancement. METHODS: Using electrospray ionization tandem quadrupole mass spectrometry detection, we determined the ion suppression or enhancement caused by column bleed for three mixed‐mode reversed‐phase/weak anion‐exchange columns containing stationary phases that differ in chemical structure. Two of the stationary phases are based on silica particles, while the third uses ethylene‐bridged hybrid organic/inorganic particles, which have improved hydrolytic stability. Mixtures of acidic and basic analytes were combined with the chromatography flow postcolumn, both with and without a column, and their mass spectrometry ion signal responses (peak areas) were determined. The ratio of signal response with and without a column is the matrix factor. Positive ion electrospray measurements were carried out using 0.1% formic acid (pH ~ 2.7) as a mobile phase additive, and 10mM ammonium formate (pH ~ 6.4) was used for negative ion electrospray detection. RESULTS: The matrix factors under both positive and negative ionization modes were closest to 1 (0.74–1.16) for the hybrid particle‐based columns, showing minimal ion suppression or enhancement. In contrast, the silica‐based columns gave matrix factors ranging from 0.04 to 1.86, indicating high levels of ion suppression or enhancement. These results may be explained by the differences in the structures of the stationary phases, which affect the relative amounts of hydrolysis products that elute from the columns. CONCLUSIONS: The low levels of mass spectrometry ion suppression or enhancement caused by column bleed from the hybrid particle‐based columns should allow for accurate quantitative mass spectrometric detection combined with mixed‐mode reversed‐phase/weak anion‐exchange chromatography.
format Online
Article
Text
id pubmed-8244038
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-82440382021-07-02 Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns Walter, Thomas H. Blaze M. T., Melvin Boissel, Cheryl Rapid Commun Mass Spectrom Research Articles RATIONALE: Mixed‐mode reversed‐phase/anion exchange liquid chromatography is useful for separations of mixtures containing anions (e.g. ionized acids). However, when using this form of liquid chromatography with mass spectrometry detection, the bleed of amine‐containing hydrolysis products from the columns may cause ion suppression or enhancement. METHODS: Using electrospray ionization tandem quadrupole mass spectrometry detection, we determined the ion suppression or enhancement caused by column bleed for three mixed‐mode reversed‐phase/weak anion‐exchange columns containing stationary phases that differ in chemical structure. Two of the stationary phases are based on silica particles, while the third uses ethylene‐bridged hybrid organic/inorganic particles, which have improved hydrolytic stability. Mixtures of acidic and basic analytes were combined with the chromatography flow postcolumn, both with and without a column, and their mass spectrometry ion signal responses (peak areas) were determined. The ratio of signal response with and without a column is the matrix factor. Positive ion electrospray measurements were carried out using 0.1% formic acid (pH ~ 2.7) as a mobile phase additive, and 10mM ammonium formate (pH ~ 6.4) was used for negative ion electrospray detection. RESULTS: The matrix factors under both positive and negative ionization modes were closest to 1 (0.74–1.16) for the hybrid particle‐based columns, showing minimal ion suppression or enhancement. In contrast, the silica‐based columns gave matrix factors ranging from 0.04 to 1.86, indicating high levels of ion suppression or enhancement. These results may be explained by the differences in the structures of the stationary phases, which affect the relative amounts of hydrolysis products that elute from the columns. CONCLUSIONS: The low levels of mass spectrometry ion suppression or enhancement caused by column bleed from the hybrid particle‐based columns should allow for accurate quantitative mass spectrometric detection combined with mixed‐mode reversed‐phase/weak anion‐exchange chromatography. John Wiley and Sons Inc. 2021-05-04 2021-06-30 /pmc/articles/PMC8244038/ /pubmed/33830546 http://dx.doi.org/10.1002/rcm.9098 Text en © 2021 The Authors. Rapid Communications in Mass Spectrometry published by John Wiley & Sons Ltd. https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Articles
Walter, Thomas H.
Blaze M. T., Melvin
Boissel, Cheryl
Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns
title Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns
title_full Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns
title_fullStr Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns
title_full_unstemmed Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns
title_short Electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns
title_sort electrospray ionization mass spectrometry ion suppression/enhancement caused by column bleed for three mixed‐mode reversed‐phase/anion‐exchange high‐performance liquid chromatography columns
topic Research Articles
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8244038/
https://www.ncbi.nlm.nih.gov/pubmed/33830546
http://dx.doi.org/10.1002/rcm.9098
work_keys_str_mv AT walterthomash electrosprayionizationmassspectrometryionsuppressionenhancementcausedbycolumnbleedforthreemixedmodereversedphaseanionexchangehighperformanceliquidchromatographycolumns
AT blazemtmelvin electrosprayionizationmassspectrometryionsuppressionenhancementcausedbycolumnbleedforthreemixedmodereversedphaseanionexchangehighperformanceliquidchromatographycolumns
AT boisselcheryl electrosprayionizationmassspectrometryionsuppressionenhancementcausedbycolumnbleedforthreemixedmodereversedphaseanionexchangehighperformanceliquidchromatographycolumns