Cargando…

Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall for VF-waveform guided defibrillation protocols

AIM: In cardiac arrest, ventricular fibrillation (VF) waveform analysis has identified the amplitude spectrum area (AMSA) as a key predictor of defibrillation success and favorable neurologic survival. New resuscitation protocols are under investigation, where prompt defibrillation is restricted to...

Descripción completa

Detalles Bibliográficos
Autores principales: Thannhauser, Jos, Nas, Joris, Vart, Priya, Smeets, Joep L.R.M., de Boer, Menko-Jan, van Royen, Niels, Bonnes, Judith L., Brouwer, Marc A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8244524/
https://www.ncbi.nlm.nih.gov/pubmed/34223374
http://dx.doi.org/10.1016/j.resplu.2021.100114
_version_ 1783715953615306752
author Thannhauser, Jos
Nas, Joris
Vart, Priya
Smeets, Joep L.R.M.
de Boer, Menko-Jan
van Royen, Niels
Bonnes, Judith L.
Brouwer, Marc A.
author_facet Thannhauser, Jos
Nas, Joris
Vart, Priya
Smeets, Joep L.R.M.
de Boer, Menko-Jan
van Royen, Niels
Bonnes, Judith L.
Brouwer, Marc A.
author_sort Thannhauser, Jos
collection PubMed
description AIM: In cardiac arrest, ventricular fibrillation (VF) waveform analysis has identified the amplitude spectrum area (AMSA) as a key predictor of defibrillation success and favorable neurologic survival. New resuscitation protocols are under investigation, where prompt defibrillation is restricted to cases with a high AMSA. Appreciating the variability of in-field pad placement, we aimed to assess the impact of recording direction on AMSA-values, and the inherent defibrillation advice. METHODS: Prospective VF-waveform study on 12-lead surface electrocardiograms (ECGs) obtained during defibrillation testing in ICD-recipients (2010–2017). AMSA-values (mVHz) of simultaneous VF-recordings were calculated and compared between all limb leads, with lead II as reference (proxy for in-field pad position). AMSA-differences between leads I and II were quantified using Bland-Altman analysis. Moreover, we investigated differences between these adjacent leads regarding classification into high (≥15.5), intermediate (6.5–15.5) or low (≤6.5) AMSA-values. RESULTS: In this cohort (n = 243), AMSA-values in lead II (10.2 ± 4.8) differed significantly from the other limb leads (I: 8.0 ± 3.4; III: 12.9 ± 5.6, both p < 0.001). The AMSA-value in lead I was, on average, 2.24 ± 4.3 lower than in lead II. Of the subjects with high AMSA-values in lead II, only 15% were classified as high if based on assessments of lead I. For intermediate and low AMSA-values, concordances were 66% and 72% respectively. CONCLUSIONS: ECG-recording direction markedly affects the result of VF-waveform analysis, with 20–30% lower AMSA-values in lead I than in lead II. Our data suggest that electrode positioning may significantly impact shock guidance by ‘smart defibrillators’, especially affecting the advice for prompt defibrillation.
format Online
Article
Text
id pubmed-8244524
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-82445242021-07-02 Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall for VF-waveform guided defibrillation protocols Thannhauser, Jos Nas, Joris Vart, Priya Smeets, Joep L.R.M. de Boer, Menko-Jan van Royen, Niels Bonnes, Judith L. Brouwer, Marc A. Resusc Plus Clinical Paper AIM: In cardiac arrest, ventricular fibrillation (VF) waveform analysis has identified the amplitude spectrum area (AMSA) as a key predictor of defibrillation success and favorable neurologic survival. New resuscitation protocols are under investigation, where prompt defibrillation is restricted to cases with a high AMSA. Appreciating the variability of in-field pad placement, we aimed to assess the impact of recording direction on AMSA-values, and the inherent defibrillation advice. METHODS: Prospective VF-waveform study on 12-lead surface electrocardiograms (ECGs) obtained during defibrillation testing in ICD-recipients (2010–2017). AMSA-values (mVHz) of simultaneous VF-recordings were calculated and compared between all limb leads, with lead II as reference (proxy for in-field pad position). AMSA-differences between leads I and II were quantified using Bland-Altman analysis. Moreover, we investigated differences between these adjacent leads regarding classification into high (≥15.5), intermediate (6.5–15.5) or low (≤6.5) AMSA-values. RESULTS: In this cohort (n = 243), AMSA-values in lead II (10.2 ± 4.8) differed significantly from the other limb leads (I: 8.0 ± 3.4; III: 12.9 ± 5.6, both p < 0.001). The AMSA-value in lead I was, on average, 2.24 ± 4.3 lower than in lead II. Of the subjects with high AMSA-values in lead II, only 15% were classified as high if based on assessments of lead I. For intermediate and low AMSA-values, concordances were 66% and 72% respectively. CONCLUSIONS: ECG-recording direction markedly affects the result of VF-waveform analysis, with 20–30% lower AMSA-values in lead I than in lead II. Our data suggest that electrode positioning may significantly impact shock guidance by ‘smart defibrillators’, especially affecting the advice for prompt defibrillation. Elsevier 2021-04-02 /pmc/articles/PMC8244524/ /pubmed/34223374 http://dx.doi.org/10.1016/j.resplu.2021.100114 Text en © 2021 The Authors https://creativecommons.org/licenses/by/4.0/This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Clinical Paper
Thannhauser, Jos
Nas, Joris
Vart, Priya
Smeets, Joep L.R.M.
de Boer, Menko-Jan
van Royen, Niels
Bonnes, Judith L.
Brouwer, Marc A.
Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall for VF-waveform guided defibrillation protocols
title Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall for VF-waveform guided defibrillation protocols
title_full Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall for VF-waveform guided defibrillation protocols
title_fullStr Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall for VF-waveform guided defibrillation protocols
title_full_unstemmed Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall for VF-waveform guided defibrillation protocols
title_short Electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: A potential pitfall for VF-waveform guided defibrillation protocols
title_sort electrocardiographic recording direction impacts ventricular fibrillation waveform measurements: a potential pitfall for vf-waveform guided defibrillation protocols
topic Clinical Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8244524/
https://www.ncbi.nlm.nih.gov/pubmed/34223374
http://dx.doi.org/10.1016/j.resplu.2021.100114
work_keys_str_mv AT thannhauserjos electrocardiographicrecordingdirectionimpactsventricularfibrillationwaveformmeasurementsapotentialpitfallforvfwaveformguideddefibrillationprotocols
AT nasjoris electrocardiographicrecordingdirectionimpactsventricularfibrillationwaveformmeasurementsapotentialpitfallforvfwaveformguideddefibrillationprotocols
AT vartpriya electrocardiographicrecordingdirectionimpactsventricularfibrillationwaveformmeasurementsapotentialpitfallforvfwaveformguideddefibrillationprotocols
AT smeetsjoeplrm electrocardiographicrecordingdirectionimpactsventricularfibrillationwaveformmeasurementsapotentialpitfallforvfwaveformguideddefibrillationprotocols
AT deboermenkojan electrocardiographicrecordingdirectionimpactsventricularfibrillationwaveformmeasurementsapotentialpitfallforvfwaveformguideddefibrillationprotocols
AT vanroyenniels electrocardiographicrecordingdirectionimpactsventricularfibrillationwaveformmeasurementsapotentialpitfallforvfwaveformguideddefibrillationprotocols
AT bonnesjudithl electrocardiographicrecordingdirectionimpactsventricularfibrillationwaveformmeasurementsapotentialpitfallforvfwaveformguideddefibrillationprotocols
AT brouwermarca electrocardiographicrecordingdirectionimpactsventricularfibrillationwaveformmeasurementsapotentialpitfallforvfwaveformguideddefibrillationprotocols