Cargando…

Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases

OBJECTIVES: The aim of this work was to evaluate the operation of the 1600SRS detector and to develop a calibration procedure for verifying the dose delivered by a single isocenter stereotactic radiosurgery (SRS) treatment of small multiple brain metastases (BM). METHODS: 14 clinical treatment cases...

Descripción completa

Detalles Bibliográficos
Autores principales: Decabooter, Esther, Swinnen, Ans CC, Öllers, Michel C, Göpfert, Fabian, Verhaegen, Frank
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Institute of Radiology. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8248213/
https://www.ncbi.nlm.nih.gov/pubmed/34111967
http://dx.doi.org/10.1259/bjr.20210473
_version_ 1783716678684639232
author Decabooter, Esther
Swinnen, Ans CC
Öllers, Michel C
Göpfert, Fabian
Verhaegen, Frank
author_facet Decabooter, Esther
Swinnen, Ans CC
Öllers, Michel C
Göpfert, Fabian
Verhaegen, Frank
author_sort Decabooter, Esther
collection PubMed
description OBJECTIVES: The aim of this work was to evaluate the operation of the 1600SRS detector and to develop a calibration procedure for verifying the dose delivered by a single isocenter stereotactic radiosurgery (SRS) treatment of small multiple brain metastases (BM). METHODS: 14 clinical treatment cases were selected with the number of BM ranging from 2 to 11. The dosimetric agreement was investigated between the calculated and the measured dose by an OCTAVIUS 1600SRS array detector in an OCTAVIUS 4D phantom equipped with dedicated SRS top. The cross-calibration procedure deviated from the manufacturer’s as it applied field sizes and dose rates corresponding to the volumetric modulated arc therapy segments in each plan. RESULTS: Measurements with a plan specific cross-calibration showed mean ± standard deviation (SD) agreement scores for cut-off values 50%, 80%, 95%, of 98.6 ± 1.7%, 96.5 ± 4.6%, 97.3 ± 4.4% for the 6 MV plans respectively, and 98.6 ± 1.5%, 96.6 ± 4.0% 96.4 ± 6.3%, for the 6 MV flattening filter free (FFF) plans respectively. Using the default calibration procedure instead of the plan specific calibration could lead to a combined systematic dose offset of 4.1% for our treatment plans. CONCLUSION: The 1600SRS detector array with the 4D phantom offers an accurate solution to perform routine quality assurance measurements of single isocenter SRS treatments of multiple BM. This work points out the necessity of an adapted cross-calibration procedure. ADVANCES IN KNOWLEDGE: A dedicated calibration procedure enables accurate dosimetry with the 1600SRS detector for small field single isocenter SRS treatment of multiple brain metastases for a large amount of BM.
format Online
Article
Text
id pubmed-8248213
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher The British Institute of Radiology.
record_format MEDLINE/PubMed
spelling pubmed-82482132022-07-01 Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases Decabooter, Esther Swinnen, Ans CC Öllers, Michel C Göpfert, Fabian Verhaegen, Frank Br J Radiol Full Paper OBJECTIVES: The aim of this work was to evaluate the operation of the 1600SRS detector and to develop a calibration procedure for verifying the dose delivered by a single isocenter stereotactic radiosurgery (SRS) treatment of small multiple brain metastases (BM). METHODS: 14 clinical treatment cases were selected with the number of BM ranging from 2 to 11. The dosimetric agreement was investigated between the calculated and the measured dose by an OCTAVIUS 1600SRS array detector in an OCTAVIUS 4D phantom equipped with dedicated SRS top. The cross-calibration procedure deviated from the manufacturer’s as it applied field sizes and dose rates corresponding to the volumetric modulated arc therapy segments in each plan. RESULTS: Measurements with a plan specific cross-calibration showed mean ± standard deviation (SD) agreement scores for cut-off values 50%, 80%, 95%, of 98.6 ± 1.7%, 96.5 ± 4.6%, 97.3 ± 4.4% for the 6 MV plans respectively, and 98.6 ± 1.5%, 96.6 ± 4.0% 96.4 ± 6.3%, for the 6 MV flattening filter free (FFF) plans respectively. Using the default calibration procedure instead of the plan specific calibration could lead to a combined systematic dose offset of 4.1% for our treatment plans. CONCLUSION: The 1600SRS detector array with the 4D phantom offers an accurate solution to perform routine quality assurance measurements of single isocenter SRS treatments of multiple BM. This work points out the necessity of an adapted cross-calibration procedure. ADVANCES IN KNOWLEDGE: A dedicated calibration procedure enables accurate dosimetry with the 1600SRS detector for small field single isocenter SRS treatment of multiple brain metastases for a large amount of BM. The British Institute of Radiology. 2021-07-01 2021-06-11 /pmc/articles/PMC8248213/ /pubmed/34111967 http://dx.doi.org/10.1259/bjr.20210473 Text en © 2021 The Authors. Published by the British Institute of Radiology https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 Unported License http://creativecommons.org/licenses/by-nc/4.0/ (https://creativecommons.org/licenses/by-nc/4.0/) , which permits unrestricted non-commercial reuse, provided the original author and source are credited.
spellingShingle Full Paper
Decabooter, Esther
Swinnen, Ans CC
Öllers, Michel C
Göpfert, Fabian
Verhaegen, Frank
Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases
title Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases
title_full Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases
title_fullStr Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases
title_full_unstemmed Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases
title_short Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases
title_sort operation and calibration of the novel ptw 1600srs detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases
topic Full Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8248213/
https://www.ncbi.nlm.nih.gov/pubmed/34111967
http://dx.doi.org/10.1259/bjr.20210473
work_keys_str_mv AT decabooteresther operationandcalibrationofthenovelptw1600srsdetectorfortheverificationofsingleisocenterstereotacticradiosurgerytreatmentsofmultiplesmallbrainmetastases
AT swinnenanscc operationandcalibrationofthenovelptw1600srsdetectorfortheverificationofsingleisocenterstereotacticradiosurgerytreatmentsofmultiplesmallbrainmetastases
AT ollersmichelc operationandcalibrationofthenovelptw1600srsdetectorfortheverificationofsingleisocenterstereotacticradiosurgerytreatmentsofmultiplesmallbrainmetastases
AT gopfertfabian operationandcalibrationofthenovelptw1600srsdetectorfortheverificationofsingleisocenterstereotacticradiosurgerytreatmentsofmultiplesmallbrainmetastases
AT verhaegenfrank operationandcalibrationofthenovelptw1600srsdetectorfortheverificationofsingleisocenterstereotacticradiosurgerytreatmentsofmultiplesmallbrainmetastases