Geometrical model for calculating the effect of surface morphology on total x‐ray output of medical x‐ray tubes

PURPOSE: Correlation of characteristic surface appearance and surface roughness with measured air kerma (kinetic energy released in air) reduction of tungsten‐rhenium (WRe) stationary anode surfaces. METHODS: A stationary anode test system was developed and used to alter nine initially ground sample...

Descripción completa

Detalles Bibliográficos
Autores principales: Siller, Maximilian, Minkkinen, Mika, Bogust, Pamela, Jelinek, Alexander, Schatte, Jürgen, Bostrom, Neil, Greenland, Kasey, Knabl, Wolfram, Clemens, Helmut, Pippan, Reinhard, Maier‐Kiener, Verena
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8248437/
https://www.ncbi.nlm.nih.gov/pubmed/33296505
http://dx.doi.org/10.1002/mp.14649
_version_ 1783716724438204416
author Siller, Maximilian
Minkkinen, Mika
Bogust, Pamela
Jelinek, Alexander
Schatte, Jürgen
Bostrom, Neil
Greenland, Kasey
Knabl, Wolfram
Clemens, Helmut
Pippan, Reinhard
Maier‐Kiener, Verena
author_facet Siller, Maximilian
Minkkinen, Mika
Bogust, Pamela
Jelinek, Alexander
Schatte, Jürgen
Bostrom, Neil
Greenland, Kasey
Knabl, Wolfram
Clemens, Helmut
Pippan, Reinhard
Maier‐Kiener, Verena
author_sort Siller, Maximilian
collection PubMed
description PURPOSE: Correlation of characteristic surface appearance and surface roughness with measured air kerma (kinetic energy released in air) reduction of tungsten‐rhenium (WRe) stationary anode surfaces. METHODS: A stationary anode test system was developed and used to alter nine initially ground sample surfaces through thermal cycling at high temperatures. A geometrical model based on high resolution surface data was implemented to correlate the measured reduction of the air kerma rate with the changing surface appearance of the samples. In addition to the nine thermally cycled samples, three samples received synthetic surface structuring to prove the applicability of the model to nonconventional surface alterations. Representative surface data and surface roughness values were acquired by laser scanning confocal microscopy. RESULTS: After thermal cycling in the stationary anode test system, the samples showed surface features comparable to rotating anodes after long‐time operation. The established model enables the appearance of characteristic surface features like crack networks, pitting, and local melting to be linked to the local x‐ray output at 100 kV tube voltage ,10° anode take off angle and 2 mm of added Al filtration. The results from the conducted air kerma measurements were compared to the predicted total x‐ray output reduction from the geometrical model and show, on average, less than 10 % error within the 12 tested samples. In certain boundaries, the calculated surface roughness R(a) showed a linear correlation with the measured air kerma reduction when samples were having comparable damaging characteristics and similar operation parameters. The orientation of the surface features had a strong impact on the measured air kerma rate which was shown by testing synthetically structured surfaces. CONCLUSIONS: The geometrical model used herein considers and describes the effect of individual surface features on the x‐ray output. In close boundaries arithmetic surface roughness R(a) was found to be a useful characteristic value on estimating the effect of surface damage on total x‐ray output.
format Online
Article
Text
id pubmed-8248437
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-82484372021-07-06 Geometrical model for calculating the effect of surface morphology on total x‐ray output of medical x‐ray tubes Siller, Maximilian Minkkinen, Mika Bogust, Pamela Jelinek, Alexander Schatte, Jürgen Bostrom, Neil Greenland, Kasey Knabl, Wolfram Clemens, Helmut Pippan, Reinhard Maier‐Kiener, Verena Med Phys DIAGNOSTIC IMAGING (IONIZING AND NON‐IONIZING) PURPOSE: Correlation of characteristic surface appearance and surface roughness with measured air kerma (kinetic energy released in air) reduction of tungsten‐rhenium (WRe) stationary anode surfaces. METHODS: A stationary anode test system was developed and used to alter nine initially ground sample surfaces through thermal cycling at high temperatures. A geometrical model based on high resolution surface data was implemented to correlate the measured reduction of the air kerma rate with the changing surface appearance of the samples. In addition to the nine thermally cycled samples, three samples received synthetic surface structuring to prove the applicability of the model to nonconventional surface alterations. Representative surface data and surface roughness values were acquired by laser scanning confocal microscopy. RESULTS: After thermal cycling in the stationary anode test system, the samples showed surface features comparable to rotating anodes after long‐time operation. The established model enables the appearance of characteristic surface features like crack networks, pitting, and local melting to be linked to the local x‐ray output at 100 kV tube voltage ,10° anode take off angle and 2 mm of added Al filtration. The results from the conducted air kerma measurements were compared to the predicted total x‐ray output reduction from the geometrical model and show, on average, less than 10 % error within the 12 tested samples. In certain boundaries, the calculated surface roughness R(a) showed a linear correlation with the measured air kerma reduction when samples were having comparable damaging characteristics and similar operation parameters. The orientation of the surface features had a strong impact on the measured air kerma rate which was shown by testing synthetically structured surfaces. CONCLUSIONS: The geometrical model used herein considers and describes the effect of individual surface features on the x‐ray output. In close boundaries arithmetic surface roughness R(a) was found to be a useful characteristic value on estimating the effect of surface damage on total x‐ray output. John Wiley and Sons Inc. 2021-02-17 2021-04 /pmc/articles/PMC8248437/ /pubmed/33296505 http://dx.doi.org/10.1002/mp.14649 Text en © 2020 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine https://creativecommons.org/licenses/by/4.0/This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle DIAGNOSTIC IMAGING (IONIZING AND NON‐IONIZING)
Siller, Maximilian
Minkkinen, Mika
Bogust, Pamela
Jelinek, Alexander
Schatte, Jürgen
Bostrom, Neil
Greenland, Kasey
Knabl, Wolfram
Clemens, Helmut
Pippan, Reinhard
Maier‐Kiener, Verena
Geometrical model for calculating the effect of surface morphology on total x‐ray output of medical x‐ray tubes
title Geometrical model for calculating the effect of surface morphology on total x‐ray output of medical x‐ray tubes
title_full Geometrical model for calculating the effect of surface morphology on total x‐ray output of medical x‐ray tubes
title_fullStr Geometrical model for calculating the effect of surface morphology on total x‐ray output of medical x‐ray tubes
title_full_unstemmed Geometrical model for calculating the effect of surface morphology on total x‐ray output of medical x‐ray tubes
title_short Geometrical model for calculating the effect of surface morphology on total x‐ray output of medical x‐ray tubes
title_sort geometrical model for calculating the effect of surface morphology on total x‐ray output of medical x‐ray tubes
topic DIAGNOSTIC IMAGING (IONIZING AND NON‐IONIZING)
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8248437/
https://www.ncbi.nlm.nih.gov/pubmed/33296505
http://dx.doi.org/10.1002/mp.14649
work_keys_str_mv AT sillermaximilian geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT minkkinenmika geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT bogustpamela geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT jelinekalexander geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT schattejurgen geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT bostromneil geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT greenlandkasey geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT knablwolfram geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT clemenshelmut geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT pippanreinhard geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes
AT maierkienerverena geometricalmodelforcalculatingtheeffectofsurfacemorphologyontotalxrayoutputofmedicalxraytubes