Effect of repeated use of an implant handpiece on an output torque: An in-vitro study

PURPOSE: This study aimed to evaluate the effect of repeated use of an implant handpiece under an implant placement torque (35 Ncm) and overloading torque condition (50 Ncm) on an output torque. MATERIALS AND METHODS: Two types of implant handpiece systems (Surgicpro/X-DSG20L [NSK, Kanuma, Japan] an...

Descripción completa

Detalles Bibliográficos
Autores principales: Son, KeunBaDa, Son, Young-Tak, Kim, Ji-Young, Lee, Jae-Mok, Yu, Won-Jae, Kim, Jin-Wook, Lee, Kyu-Bok
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Prosthodontics 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8250193/
https://www.ncbi.nlm.nih.gov/pubmed/34234923
http://dx.doi.org/10.4047/jap.2021.13.3.136
_version_ 1783717035313725440
author Son, KeunBaDa
Son, Young-Tak
Kim, Ji-Young
Lee, Jae-Mok
Yu, Won-Jae
Kim, Jin-Wook
Lee, Kyu-Bok
author_facet Son, KeunBaDa
Son, Young-Tak
Kim, Ji-Young
Lee, Jae-Mok
Yu, Won-Jae
Kim, Jin-Wook
Lee, Kyu-Bok
author_sort Son, KeunBaDa
collection PubMed
description PURPOSE: This study aimed to evaluate the effect of repeated use of an implant handpiece under an implant placement torque (35 Ncm) and overloading torque condition (50 Ncm) on an output torque. MATERIALS AND METHODS: Two types of implant handpiece systems (Surgicpro/X-DSG20L [NSK, Kanuma, Japan] and SIP20/CRB46LN [SAESHIN, Daegu, South Korea]) were used. The output torque was measured using a digital torque gauge. The height and angle (x, y, and z axes) of the digital torque gauge and implant handpiece were adjusted through a jig for passive connection. The experiment was conducted under the setting torque value of 35 Ncm (implant placement torque) and 50 Ncm (overloading torque condition) and 30 times per set; a total of 5 sets were performed (N = 150). For statistical analysis, the difference between the groups was analyzed using the Mann-Whitney U test and the Friedman test was used to confirm the change in output torque (α=.05). RESULTS: NSK and SAESHIN implant handpieces showed significant differences in output torque results at the setting torques of 35 Ncm and 50 Ncm (P<.001). The type of implant handpiece and repeated use influenced the output torque (P<.001). CONCLUSION: There may be a difference between the setting torque and actual output torque due to repeated use, and the implant handpiece should be managed and repaired during long-term use. In addition, for successful implant results in dental clinics, the output torque of the implant handpiece system should be checked before implant placement.
format Online
Article
Text
id pubmed-8250193
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher The Korean Academy of Prosthodontics
record_format MEDLINE/PubMed
spelling pubmed-82501932021-07-06 Effect of repeated use of an implant handpiece on an output torque: An in-vitro study Son, KeunBaDa Son, Young-Tak Kim, Ji-Young Lee, Jae-Mok Yu, Won-Jae Kim, Jin-Wook Lee, Kyu-Bok J Adv Prosthodont Original Article PURPOSE: This study aimed to evaluate the effect of repeated use of an implant handpiece under an implant placement torque (35 Ncm) and overloading torque condition (50 Ncm) on an output torque. MATERIALS AND METHODS: Two types of implant handpiece systems (Surgicpro/X-DSG20L [NSK, Kanuma, Japan] and SIP20/CRB46LN [SAESHIN, Daegu, South Korea]) were used. The output torque was measured using a digital torque gauge. The height and angle (x, y, and z axes) of the digital torque gauge and implant handpiece were adjusted through a jig for passive connection. The experiment was conducted under the setting torque value of 35 Ncm (implant placement torque) and 50 Ncm (overloading torque condition) and 30 times per set; a total of 5 sets were performed (N = 150). For statistical analysis, the difference between the groups was analyzed using the Mann-Whitney U test and the Friedman test was used to confirm the change in output torque (α=.05). RESULTS: NSK and SAESHIN implant handpieces showed significant differences in output torque results at the setting torques of 35 Ncm and 50 Ncm (P<.001). The type of implant handpiece and repeated use influenced the output torque (P<.001). CONCLUSION: There may be a difference between the setting torque and actual output torque due to repeated use, and the implant handpiece should be managed and repaired during long-term use. In addition, for successful implant results in dental clinics, the output torque of the implant handpiece system should be checked before implant placement. The Korean Academy of Prosthodontics 2021-06 2021-06-25 /pmc/articles/PMC8250193/ /pubmed/34234923 http://dx.doi.org/10.4047/jap.2021.13.3.136 Text en © 2021 The Korean Academy of Prosthodontics https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0 (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Son, KeunBaDa
Son, Young-Tak
Kim, Ji-Young
Lee, Jae-Mok
Yu, Won-Jae
Kim, Jin-Wook
Lee, Kyu-Bok
Effect of repeated use of an implant handpiece on an output torque: An in-vitro study
title Effect of repeated use of an implant handpiece on an output torque: An in-vitro study
title_full Effect of repeated use of an implant handpiece on an output torque: An in-vitro study
title_fullStr Effect of repeated use of an implant handpiece on an output torque: An in-vitro study
title_full_unstemmed Effect of repeated use of an implant handpiece on an output torque: An in-vitro study
title_short Effect of repeated use of an implant handpiece on an output torque: An in-vitro study
title_sort effect of repeated use of an implant handpiece on an output torque: an in-vitro study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8250193/
https://www.ncbi.nlm.nih.gov/pubmed/34234923
http://dx.doi.org/10.4047/jap.2021.13.3.136
work_keys_str_mv AT sonkeunbada effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy
AT sonyoungtak effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy
AT kimjiyoung effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy
AT leejaemok effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy
AT yuwonjae effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy
AT kimjinwook effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy
AT leekyubok effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy