Effect of repeated use of an implant handpiece on an output torque: An in-vitro study
PURPOSE: This study aimed to evaluate the effect of repeated use of an implant handpiece under an implant placement torque (35 Ncm) and overloading torque condition (50 Ncm) on an output torque. MATERIALS AND METHODS: Two types of implant handpiece systems (Surgicpro/X-DSG20L [NSK, Kanuma, Japan] an...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Academy of Prosthodontics
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8250193/ https://www.ncbi.nlm.nih.gov/pubmed/34234923 http://dx.doi.org/10.4047/jap.2021.13.3.136 |
_version_ | 1783717035313725440 |
---|---|
author | Son, KeunBaDa Son, Young-Tak Kim, Ji-Young Lee, Jae-Mok Yu, Won-Jae Kim, Jin-Wook Lee, Kyu-Bok |
author_facet | Son, KeunBaDa Son, Young-Tak Kim, Ji-Young Lee, Jae-Mok Yu, Won-Jae Kim, Jin-Wook Lee, Kyu-Bok |
author_sort | Son, KeunBaDa |
collection | PubMed |
description | PURPOSE: This study aimed to evaluate the effect of repeated use of an implant handpiece under an implant placement torque (35 Ncm) and overloading torque condition (50 Ncm) on an output torque. MATERIALS AND METHODS: Two types of implant handpiece systems (Surgicpro/X-DSG20L [NSK, Kanuma, Japan] and SIP20/CRB46LN [SAESHIN, Daegu, South Korea]) were used. The output torque was measured using a digital torque gauge. The height and angle (x, y, and z axes) of the digital torque gauge and implant handpiece were adjusted through a jig for passive connection. The experiment was conducted under the setting torque value of 35 Ncm (implant placement torque) and 50 Ncm (overloading torque condition) and 30 times per set; a total of 5 sets were performed (N = 150). For statistical analysis, the difference between the groups was analyzed using the Mann-Whitney U test and the Friedman test was used to confirm the change in output torque (α=.05). RESULTS: NSK and SAESHIN implant handpieces showed significant differences in output torque results at the setting torques of 35 Ncm and 50 Ncm (P<.001). The type of implant handpiece and repeated use influenced the output torque (P<.001). CONCLUSION: There may be a difference between the setting torque and actual output torque due to repeated use, and the implant handpiece should be managed and repaired during long-term use. In addition, for successful implant results in dental clinics, the output torque of the implant handpiece system should be checked before implant placement. |
format | Online Article Text |
id | pubmed-8250193 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | The Korean Academy of Prosthodontics |
record_format | MEDLINE/PubMed |
spelling | pubmed-82501932021-07-06 Effect of repeated use of an implant handpiece on an output torque: An in-vitro study Son, KeunBaDa Son, Young-Tak Kim, Ji-Young Lee, Jae-Mok Yu, Won-Jae Kim, Jin-Wook Lee, Kyu-Bok J Adv Prosthodont Original Article PURPOSE: This study aimed to evaluate the effect of repeated use of an implant handpiece under an implant placement torque (35 Ncm) and overloading torque condition (50 Ncm) on an output torque. MATERIALS AND METHODS: Two types of implant handpiece systems (Surgicpro/X-DSG20L [NSK, Kanuma, Japan] and SIP20/CRB46LN [SAESHIN, Daegu, South Korea]) were used. The output torque was measured using a digital torque gauge. The height and angle (x, y, and z axes) of the digital torque gauge and implant handpiece were adjusted through a jig for passive connection. The experiment was conducted under the setting torque value of 35 Ncm (implant placement torque) and 50 Ncm (overloading torque condition) and 30 times per set; a total of 5 sets were performed (N = 150). For statistical analysis, the difference between the groups was analyzed using the Mann-Whitney U test and the Friedman test was used to confirm the change in output torque (α=.05). RESULTS: NSK and SAESHIN implant handpieces showed significant differences in output torque results at the setting torques of 35 Ncm and 50 Ncm (P<.001). The type of implant handpiece and repeated use influenced the output torque (P<.001). CONCLUSION: There may be a difference between the setting torque and actual output torque due to repeated use, and the implant handpiece should be managed and repaired during long-term use. In addition, for successful implant results in dental clinics, the output torque of the implant handpiece system should be checked before implant placement. The Korean Academy of Prosthodontics 2021-06 2021-06-25 /pmc/articles/PMC8250193/ /pubmed/34234923 http://dx.doi.org/10.4047/jap.2021.13.3.136 Text en © 2021 The Korean Academy of Prosthodontics https://creativecommons.org/licenses/by-nc/4.0/This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0 (https://creativecommons.org/licenses/by-nc/4.0/) ) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Son, KeunBaDa Son, Young-Tak Kim, Ji-Young Lee, Jae-Mok Yu, Won-Jae Kim, Jin-Wook Lee, Kyu-Bok Effect of repeated use of an implant handpiece on an output torque: An in-vitro study |
title | Effect of repeated use of an implant handpiece on an output torque: An in-vitro study |
title_full | Effect of repeated use of an implant handpiece on an output torque: An in-vitro study |
title_fullStr | Effect of repeated use of an implant handpiece on an output torque: An in-vitro study |
title_full_unstemmed | Effect of repeated use of an implant handpiece on an output torque: An in-vitro study |
title_short | Effect of repeated use of an implant handpiece on an output torque: An in-vitro study |
title_sort | effect of repeated use of an implant handpiece on an output torque: an in-vitro study |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8250193/ https://www.ncbi.nlm.nih.gov/pubmed/34234923 http://dx.doi.org/10.4047/jap.2021.13.3.136 |
work_keys_str_mv | AT sonkeunbada effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy AT sonyoungtak effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy AT kimjiyoung effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy AT leejaemok effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy AT yuwonjae effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy AT kimjinwook effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy AT leekyubok effectofrepeateduseofanimplanthandpieceonanoutputtorqueaninvitrostudy |