Cargando…
Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures
BACKGROUND: We investigated wearable components of the Dexcom G6 continuous glucose monitoring (CGM) System in simulated therapeutic and diagnostic radiologic procedures. METHODS: G6 transmitters were loaded with simulated glucose data and attached to sensors. Sets of sensor/transmitter pairs were e...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
SAGE Publications
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8252147/ https://www.ncbi.nlm.nih.gov/pubmed/32319318 http://dx.doi.org/10.1177/1932296820920948 |
_version_ | 1783717239876222976 |
---|---|
author | Thomas, Christopher Welsh, John B. Lu, Spencer Gray, J. Michael |
author_facet | Thomas, Christopher Welsh, John B. Lu, Spencer Gray, J. Michael |
author_sort | Thomas, Christopher |
collection | PubMed |
description | BACKGROUND: We investigated wearable components of the Dexcom G6 continuous glucose monitoring (CGM) System in simulated therapeutic and diagnostic radiologic procedures. METHODS: G6 transmitters were loaded with simulated glucose data and attached to sensors. Sets of sensor/transmitter pairs were exposed to x-rays to simulate a radiotherapeutic procedure and to radiofrequency (RF) and magnetic fields to simulate diagnostic magnetic resonance imaging (MRI). The x-ray simulation provided a cumulative dose of 80 Gy. The MRI simulation used RF fields oscillating at 64 or 128 MHz and magnetic fields of 1.5 or 3 T. During the MRI simulation, displacement force, induced heating, and induced currents were measured. After the simulations, bench tests were used to assess data integrity on the transmitters and responsiveness of sensors to various concentrations of aqueous glucose. RESULTS: Glucose concentrations reported by sensor/transmitter pairs after undergoing x-irradiation or a simulated MRI exam were similar to those from control (unexposed) devices. During the 3 T MRI simulation, the devices experienced a displacement force of 306 g, which was insufficient to dislodge the sensor/transmitter from the substrate, RF-induced heating of <2°C, and an induced current of <16 pA. Data stored on the transmitters prior to the MRI simulation remained intact. CONCLUSION: Wearable components of the G6 CGM System retain basic functionality and data integrity after exposure to simulated therapeutic and diagnostic radiologic procedures. The devices are unlikely to be affected by x-irradiation used in typical imaging studies. Simulated MRI procedures create displacement force, minimal heating, and current in sensor/transmitter pairs. |
format | Online Article Text |
id | pubmed-8252147 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | SAGE Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-82521472021-07-13 Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures Thomas, Christopher Welsh, John B. Lu, Spencer Gray, J. Michael J Diabetes Sci Technol Original Articles BACKGROUND: We investigated wearable components of the Dexcom G6 continuous glucose monitoring (CGM) System in simulated therapeutic and diagnostic radiologic procedures. METHODS: G6 transmitters were loaded with simulated glucose data and attached to sensors. Sets of sensor/transmitter pairs were exposed to x-rays to simulate a radiotherapeutic procedure and to radiofrequency (RF) and magnetic fields to simulate diagnostic magnetic resonance imaging (MRI). The x-ray simulation provided a cumulative dose of 80 Gy. The MRI simulation used RF fields oscillating at 64 or 128 MHz and magnetic fields of 1.5 or 3 T. During the MRI simulation, displacement force, induced heating, and induced currents were measured. After the simulations, bench tests were used to assess data integrity on the transmitters and responsiveness of sensors to various concentrations of aqueous glucose. RESULTS: Glucose concentrations reported by sensor/transmitter pairs after undergoing x-irradiation or a simulated MRI exam were similar to those from control (unexposed) devices. During the 3 T MRI simulation, the devices experienced a displacement force of 306 g, which was insufficient to dislodge the sensor/transmitter from the substrate, RF-induced heating of <2°C, and an induced current of <16 pA. Data stored on the transmitters prior to the MRI simulation remained intact. CONCLUSION: Wearable components of the G6 CGM System retain basic functionality and data integrity after exposure to simulated therapeutic and diagnostic radiologic procedures. The devices are unlikely to be affected by x-irradiation used in typical imaging studies. Simulated MRI procedures create displacement force, minimal heating, and current in sensor/transmitter pairs. SAGE Publications 2020-04-22 /pmc/articles/PMC8252147/ /pubmed/32319318 http://dx.doi.org/10.1177/1932296820920948 Text en © 2020 Diabetes Technology Society https://creativecommons.org/licenses/by-nc/4.0/This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 License (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage). |
spellingShingle | Original Articles Thomas, Christopher Welsh, John B. Lu, Spencer Gray, J. Michael Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures |
title | Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures |
title_full | Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures |
title_fullStr | Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures |
title_full_unstemmed | Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures |
title_short | Safety and Functional Integrity of Continuous Glucose Monitoring Components After Simulated Radiologic Procedures |
title_sort | safety and functional integrity of continuous glucose monitoring components after simulated radiologic procedures |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8252147/ https://www.ncbi.nlm.nih.gov/pubmed/32319318 http://dx.doi.org/10.1177/1932296820920948 |
work_keys_str_mv | AT thomaschristopher safetyandfunctionalintegrityofcontinuousglucosemonitoringcomponentsaftersimulatedradiologicprocedures AT welshjohnb safetyandfunctionalintegrityofcontinuousglucosemonitoringcomponentsaftersimulatedradiologicprocedures AT luspencer safetyandfunctionalintegrityofcontinuousglucosemonitoringcomponentsaftersimulatedradiologicprocedures AT grayjmichael safetyandfunctionalintegrityofcontinuousglucosemonitoringcomponentsaftersimulatedradiologicprocedures |