Harmonisation of PET/CT contrast recovery performance for brain studies

PURPOSE: In order to achieve comparability of image quality, harmonisation of PET system performance is imperative. In this study, prototype harmonisation criteria for PET brain studies were developed. METHODS: Twelve clinical PET/CT systems (4 GE, 4 Philips, 4 Siemens, including SiPM-based “digital...

Descripción completa

Detalles Bibliográficos
Autores principales: Verwer, E. E., Golla, S. S. V., Kaalep, A., Lubberink, M., van Velden, F. H. P., Bettinardi, V., Yaqub, M., Sera, T., Rijnsdorp, S., Lammertsma, A. A., Boellaard, R.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8263427/
https://www.ncbi.nlm.nih.gov/pubmed/33517517
http://dx.doi.org/10.1007/s00259-021-05201-w
_version_ 1783719388337143808
author Verwer, E. E.
Golla, S. S. V.
Kaalep, A.
Lubberink, M.
van Velden, F. H. P.
Bettinardi, V.
Yaqub, M.
Sera, T.
Rijnsdorp, S.
Lammertsma, A. A.
Boellaard, R.
author_facet Verwer, E. E.
Golla, S. S. V.
Kaalep, A.
Lubberink, M.
van Velden, F. H. P.
Bettinardi, V.
Yaqub, M.
Sera, T.
Rijnsdorp, S.
Lammertsma, A. A.
Boellaard, R.
author_sort Verwer, E. E.
collection PubMed
description PURPOSE: In order to achieve comparability of image quality, harmonisation of PET system performance is imperative. In this study, prototype harmonisation criteria for PET brain studies were developed. METHODS: Twelve clinical PET/CT systems (4 GE, 4 Philips, 4 Siemens, including SiPM-based “digital” systems) were used to acquire 30-min PET scans of a Hoffman 3D Brain phantom filled with ~ 33 kBq·mL(−1) [(18)F]FDG. Scan data were reconstructed using various reconstruction settings. The images were rigidly coregistered to a template (voxel size 1.17 × 1.17 × 2.00 mm(3)) onto which several volumes of interest (VOIs) were defined. Recovery coefficients (RC) and grey matter to white matter ratios (GMWMr) were derived for eroded (denoted in the text by subscript e) and non-eroded grey (GM) and white (WM) matter VOIs as well as a mid-phantom cold spot (VOI(cold)) and VOIs from the Hammers atlas. In addition, left-right hemisphere differences and voxel-by-voxel differences compared to a reference image were assessed. RESULTS: Systematic differences were observed for reconstructions with and without point-spread-function modelling (PSF(ON) and PSF(OFF), respectively). Normalising to image-derived activity, upper and lower limits ensuring image comparability were as follows: for PSF(ON), RC(GMe) = [0.97–1.01] and GMWMr(e) = [3.51–3.91] for eroded VOI and RC(GM) = [0.78–0.83] and GMWMr = [1.77–2.06] for non-eroded VOI, and for PSF(OFF), RC(GMe) = [0.92–0.99] and GMWMr(e) = [3.14–3.68] for eroded VOI and RC(GM) = [0.75–0.81] and GMWMr = [1.72–1.95] for non-eroded VOI. CONCLUSIONS: To achieve inter-scanner comparability, we propose selecting reconstruction settings based on RC(GMe) and GMWMr(e) as specified in “Results”. These proposed standards should be tested prospectively to validate and/or refine the harmonisation criteria. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00259-021-05201-w.
format Online
Article
Text
id pubmed-8263427
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-82634272021-07-20 Harmonisation of PET/CT contrast recovery performance for brain studies Verwer, E. E. Golla, S. S. V. Kaalep, A. Lubberink, M. van Velden, F. H. P. Bettinardi, V. Yaqub, M. Sera, T. Rijnsdorp, S. Lammertsma, A. A. Boellaard, R. Eur J Nucl Med Mol Imaging Original Article PURPOSE: In order to achieve comparability of image quality, harmonisation of PET system performance is imperative. In this study, prototype harmonisation criteria for PET brain studies were developed. METHODS: Twelve clinical PET/CT systems (4 GE, 4 Philips, 4 Siemens, including SiPM-based “digital” systems) were used to acquire 30-min PET scans of a Hoffman 3D Brain phantom filled with ~ 33 kBq·mL(−1) [(18)F]FDG. Scan data were reconstructed using various reconstruction settings. The images were rigidly coregistered to a template (voxel size 1.17 × 1.17 × 2.00 mm(3)) onto which several volumes of interest (VOIs) were defined. Recovery coefficients (RC) and grey matter to white matter ratios (GMWMr) were derived for eroded (denoted in the text by subscript e) and non-eroded grey (GM) and white (WM) matter VOIs as well as a mid-phantom cold spot (VOI(cold)) and VOIs from the Hammers atlas. In addition, left-right hemisphere differences and voxel-by-voxel differences compared to a reference image were assessed. RESULTS: Systematic differences were observed for reconstructions with and without point-spread-function modelling (PSF(ON) and PSF(OFF), respectively). Normalising to image-derived activity, upper and lower limits ensuring image comparability were as follows: for PSF(ON), RC(GMe) = [0.97–1.01] and GMWMr(e) = [3.51–3.91] for eroded VOI and RC(GM) = [0.78–0.83] and GMWMr = [1.77–2.06] for non-eroded VOI, and for PSF(OFF), RC(GMe) = [0.92–0.99] and GMWMr(e) = [3.14–3.68] for eroded VOI and RC(GM) = [0.75–0.81] and GMWMr = [1.72–1.95] for non-eroded VOI. CONCLUSIONS: To achieve inter-scanner comparability, we propose selecting reconstruction settings based on RC(GMe) and GMWMr(e) as specified in “Results”. These proposed standards should be tested prospectively to validate and/or refine the harmonisation criteria. SUPPLEMENTARY INFORMATION: The online version contains supplementary material available at 10.1007/s00259-021-05201-w. Springer Berlin Heidelberg 2021-01-31 2021 /pmc/articles/PMC8263427/ /pubmed/33517517 http://dx.doi.org/10.1007/s00259-021-05201-w Text en © The Author(s) 2021 https://creativecommons.org/licenses/by/4.0/Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/ (https://creativecommons.org/licenses/by/4.0/) .
spellingShingle Original Article
Verwer, E. E.
Golla, S. S. V.
Kaalep, A.
Lubberink, M.
van Velden, F. H. P.
Bettinardi, V.
Yaqub, M.
Sera, T.
Rijnsdorp, S.
Lammertsma, A. A.
Boellaard, R.
Harmonisation of PET/CT contrast recovery performance for brain studies
title Harmonisation of PET/CT contrast recovery performance for brain studies
title_full Harmonisation of PET/CT contrast recovery performance for brain studies
title_fullStr Harmonisation of PET/CT contrast recovery performance for brain studies
title_full_unstemmed Harmonisation of PET/CT contrast recovery performance for brain studies
title_short Harmonisation of PET/CT contrast recovery performance for brain studies
title_sort harmonisation of pet/ct contrast recovery performance for brain studies
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8263427/
https://www.ncbi.nlm.nih.gov/pubmed/33517517
http://dx.doi.org/10.1007/s00259-021-05201-w
work_keys_str_mv AT verweree harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT gollassv harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT kaalepa harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT lubberinkm harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT vanveldenfhp harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT bettinardiv harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT yaqubm harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT serat harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT rijnsdorps harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT lammertsmaaa harmonisationofpetctcontrastrecoveryperformanceforbrainstudies
AT boellaardr harmonisationofpetctcontrastrecoveryperformanceforbrainstudies