Repeatability and Reproducibility of Macular Hole Size Measurements Using Optical Coherence Tomography
The purpose of this study was to assess the repeatability and reproducibility of measuring the minimum linear diameter (MLD) of macular holes (MHs) using horizontal linear and radial scan modes in optical coherence tomography (OCT). Patients with concurrent sets of radial and horizontal linear OCT v...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8268292/ https://www.ncbi.nlm.nih.gov/pubmed/34209752 http://dx.doi.org/10.3390/jcm10132899 |
_version_ | 1783720324002480128 |
---|---|
author | Baumann, Carmen Almarzooqi, Ahmed Blobner, Katharina Zapp, Daniel Kirchmair, Katharina Schwer, Lydia S. Lohmann, Chris P. Kaye, Stephen B. |
author_facet | Baumann, Carmen Almarzooqi, Ahmed Blobner, Katharina Zapp, Daniel Kirchmair, Katharina Schwer, Lydia S. Lohmann, Chris P. Kaye, Stephen B. |
author_sort | Baumann, Carmen |
collection | PubMed |
description | The purpose of this study was to assess the repeatability and reproducibility of measuring the minimum linear diameter (MLD) of macular holes (MHs) using horizontal linear and radial scan modes in optical coherence tomography (OCT). Patients with concurrent sets of radial and horizontal linear OCT volume scans were included. The MLD was measured twice in both scan modes by six raters of three different experience levels (groups). Outcome measures were the reliability and repeatability of MLD measurements. Fifty patients were included. Mean MLD was 317.21(±170.63) µm in horizontal linear and 364.52 (±161.71) µm in radial mode, a difference of 47.31 (±26.48) µm (p < 0.001). In the radial scan mode, MLD was identified within 15° of the horizontal meridian in 27% and within 15° of the vertical meridian in 26.7%, with the remainder (46.3%) in oblique meridians. The intra-group coefficients of repeatability (CR) for horizontal linear mode were 23 µm, 33 µm and 45 µm, and for radial mode 25 µm, 44 µm and 57 µm for groups 1, 2 and 3, respectively. The inter-group CR, taking group 1 as reference standard for groups 2 and 3, were 74 µm and 71 µm for the linear mode, and 62 µm and 78 µm for radial mode. The radial mode provides good repeatability and reliability for measurement of MLD. In a majority of cases the MLD does not lie in the horizontal meridian and would be underestimated using a horizontal OCT mode. |
format | Online Article Text |
id | pubmed-8268292 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-82682922021-07-10 Repeatability and Reproducibility of Macular Hole Size Measurements Using Optical Coherence Tomography Baumann, Carmen Almarzooqi, Ahmed Blobner, Katharina Zapp, Daniel Kirchmair, Katharina Schwer, Lydia S. Lohmann, Chris P. Kaye, Stephen B. J Clin Med Article The purpose of this study was to assess the repeatability and reproducibility of measuring the minimum linear diameter (MLD) of macular holes (MHs) using horizontal linear and radial scan modes in optical coherence tomography (OCT). Patients with concurrent sets of radial and horizontal linear OCT volume scans were included. The MLD was measured twice in both scan modes by six raters of three different experience levels (groups). Outcome measures were the reliability and repeatability of MLD measurements. Fifty patients were included. Mean MLD was 317.21(±170.63) µm in horizontal linear and 364.52 (±161.71) µm in radial mode, a difference of 47.31 (±26.48) µm (p < 0.001). In the radial scan mode, MLD was identified within 15° of the horizontal meridian in 27% and within 15° of the vertical meridian in 26.7%, with the remainder (46.3%) in oblique meridians. The intra-group coefficients of repeatability (CR) for horizontal linear mode were 23 µm, 33 µm and 45 µm, and for radial mode 25 µm, 44 µm and 57 µm for groups 1, 2 and 3, respectively. The inter-group CR, taking group 1 as reference standard for groups 2 and 3, were 74 µm and 71 µm for the linear mode, and 62 µm and 78 µm for radial mode. The radial mode provides good repeatability and reliability for measurement of MLD. In a majority of cases the MLD does not lie in the horizontal meridian and would be underestimated using a horizontal OCT mode. MDPI 2021-06-29 /pmc/articles/PMC8268292/ /pubmed/34209752 http://dx.doi.org/10.3390/jcm10132899 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Baumann, Carmen Almarzooqi, Ahmed Blobner, Katharina Zapp, Daniel Kirchmair, Katharina Schwer, Lydia S. Lohmann, Chris P. Kaye, Stephen B. Repeatability and Reproducibility of Macular Hole Size Measurements Using Optical Coherence Tomography |
title | Repeatability and Reproducibility of Macular Hole Size Measurements Using Optical Coherence Tomography |
title_full | Repeatability and Reproducibility of Macular Hole Size Measurements Using Optical Coherence Tomography |
title_fullStr | Repeatability and Reproducibility of Macular Hole Size Measurements Using Optical Coherence Tomography |
title_full_unstemmed | Repeatability and Reproducibility of Macular Hole Size Measurements Using Optical Coherence Tomography |
title_short | Repeatability and Reproducibility of Macular Hole Size Measurements Using Optical Coherence Tomography |
title_sort | repeatability and reproducibility of macular hole size measurements using optical coherence tomography |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8268292/ https://www.ncbi.nlm.nih.gov/pubmed/34209752 http://dx.doi.org/10.3390/jcm10132899 |
work_keys_str_mv | AT baumanncarmen repeatabilityandreproducibilityofmacularholesizemeasurementsusingopticalcoherencetomography AT almarzooqiahmed repeatabilityandreproducibilityofmacularholesizemeasurementsusingopticalcoherencetomography AT blobnerkatharina repeatabilityandreproducibilityofmacularholesizemeasurementsusingopticalcoherencetomography AT zappdaniel repeatabilityandreproducibilityofmacularholesizemeasurementsusingopticalcoherencetomography AT kirchmairkatharina repeatabilityandreproducibilityofmacularholesizemeasurementsusingopticalcoherencetomography AT schwerlydias repeatabilityandreproducibilityofmacularholesizemeasurementsusingopticalcoherencetomography AT lohmannchrisp repeatabilityandreproducibilityofmacularholesizemeasurementsusingopticalcoherencetomography AT kayestephenb repeatabilityandreproducibilityofmacularholesizemeasurementsusingopticalcoherencetomography |