Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy
PURPOSE: The development of collimating technologies has become a recent focus in pencil beam scanning (PBS) proton therapy to improve the target conformity and healthy tissue sparing through field-specific or energy-layer–specific collimation. Given the growing popularity of collimators for low-ene...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Particle Therapy Co-operative Group
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8270095/ https://www.ncbi.nlm.nih.gov/pubmed/34285937 http://dx.doi.org/10.14338/IJPT-20-00039.1 |
_version_ | 1783720728420417536 |
---|---|
author | Hyer, Daniel E. Bennett, Laura C. Geoghegan, Theodore J. Bues, Martin Smith, Blake R. |
author_facet | Hyer, Daniel E. Bennett, Laura C. Geoghegan, Theodore J. Bues, Martin Smith, Blake R. |
author_sort | Hyer, Daniel E. |
collection | PubMed |
description | PURPOSE: The development of collimating technologies has become a recent focus in pencil beam scanning (PBS) proton therapy to improve the target conformity and healthy tissue sparing through field-specific or energy-layer–specific collimation. Given the growing popularity of collimators for low-energy treatments, the purpose of this work was to summarize the recent literature that has focused on the efficacy of collimators for PBS and highlight the development of clinical and preclinical collimators. MATERIALS AND METHODS: The collimators presented in this work were organized into 3 categories: per-field apertures, multileaf collimators (MLCs), and sliding-bar collimators. For each case, the system design and planning methodologies are summarized and intercompared from their existing literature. Energy-specific collimation is still a new paradigm in PBS and the 2 specific collimators tailored toward PBS are presented including the dynamic collimation system (DCS) and the Mevion Adaptive Aperture. RESULTS: Collimation during PBS can improve the target conformity and associated healthy tissue and critical structure avoidance. Between energy-specific collimators and static apertures, static apertures have the poorest dose conformity owing to collimating only the largest projection of a target in the beam's eye view but still provide an improvement over uncollimated treatments. While an external collimator increases secondary neutron production, the benefit of collimating the primary beam appears to outweigh the risk. The greatest benefit has been observed for low- energy treatment sites. CONCLUSION: The consensus from current literature supports the use of external collimators in PBS under certain conditions, namely low-energy treatments or where the nominal spot size is large. While many recent studies paint a supportive picture, it is also important to understand the limitations of collimation in PBS that are specific to each collimator type. The emergence and paradigm of energy-specific collimation holds many promises for PBS proton therapy. |
format | Online Article Text |
id | pubmed-8270095 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | The Particle Therapy Co-operative Group |
record_format | MEDLINE/PubMed |
spelling | pubmed-82700952021-07-19 Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy Hyer, Daniel E. Bennett, Laura C. Geoghegan, Theodore J. Bues, Martin Smith, Blake R. Int J Part Ther Physics PURPOSE: The development of collimating technologies has become a recent focus in pencil beam scanning (PBS) proton therapy to improve the target conformity and healthy tissue sparing through field-specific or energy-layer–specific collimation. Given the growing popularity of collimators for low-energy treatments, the purpose of this work was to summarize the recent literature that has focused on the efficacy of collimators for PBS and highlight the development of clinical and preclinical collimators. MATERIALS AND METHODS: The collimators presented in this work were organized into 3 categories: per-field apertures, multileaf collimators (MLCs), and sliding-bar collimators. For each case, the system design and planning methodologies are summarized and intercompared from their existing literature. Energy-specific collimation is still a new paradigm in PBS and the 2 specific collimators tailored toward PBS are presented including the dynamic collimation system (DCS) and the Mevion Adaptive Aperture. RESULTS: Collimation during PBS can improve the target conformity and associated healthy tissue and critical structure avoidance. Between energy-specific collimators and static apertures, static apertures have the poorest dose conformity owing to collimating only the largest projection of a target in the beam's eye view but still provide an improvement over uncollimated treatments. While an external collimator increases secondary neutron production, the benefit of collimating the primary beam appears to outweigh the risk. The greatest benefit has been observed for low- energy treatment sites. CONCLUSION: The consensus from current literature supports the use of external collimators in PBS under certain conditions, namely low-energy treatments or where the nominal spot size is large. While many recent studies paint a supportive picture, it is also important to understand the limitations of collimation in PBS that are specific to each collimator type. The emergence and paradigm of energy-specific collimation holds many promises for PBS proton therapy. The Particle Therapy Co-operative Group 2021-06-25 /pmc/articles/PMC8270095/ /pubmed/34285937 http://dx.doi.org/10.14338/IJPT-20-00039.1 Text en ©Copyright 2021 The Author(s) https://creativecommons.org/licenses/by/4.0/Distributed under Creative Commons CC-BY (https://creativecommons.org/licenses/by/4.0/) |
spellingShingle | Physics Hyer, Daniel E. Bennett, Laura C. Geoghegan, Theodore J. Bues, Martin Smith, Blake R. Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy |
title | Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy |
title_full | Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy |
title_fullStr | Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy |
title_full_unstemmed | Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy |
title_short | Innovations and the Use of Collimators in the Delivery of Pencil Beam Scanning Proton Therapy |
title_sort | innovations and the use of collimators in the delivery of pencil beam scanning proton therapy |
topic | Physics |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8270095/ https://www.ncbi.nlm.nih.gov/pubmed/34285937 http://dx.doi.org/10.14338/IJPT-20-00039.1 |
work_keys_str_mv | AT hyerdaniele innovationsandtheuseofcollimatorsinthedeliveryofpencilbeamscanningprotontherapy AT bennettlaurac innovationsandtheuseofcollimatorsinthedeliveryofpencilbeamscanningprotontherapy AT geoghegantheodorej innovationsandtheuseofcollimatorsinthedeliveryofpencilbeamscanningprotontherapy AT buesmartin innovationsandtheuseofcollimatorsinthedeliveryofpencilbeamscanningprotontherapy AT smithblaker innovationsandtheuseofcollimatorsinthedeliveryofpencilbeamscanningprotontherapy |