Validation of a Low-Cost Electrocardiography (ECG) System for Psychophysiological Research
Background and Objective: The reliability of low-cost mobile systems for recording Electrocardiographic (ECG) data is mostly unknown, posing questions regarding the quality of the recorded data and the validity of the extracted physiological parameters. The present study compared the BITalino toolki...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2021
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8271611/ https://www.ncbi.nlm.nih.gov/pubmed/34209063 http://dx.doi.org/10.3390/s21134485 |
_version_ | 1783721040358146048 |
---|---|
author | Wagner, Ruth Erna Plácido da Silva, Hugo Gramann, Klaus |
author_facet | Wagner, Ruth Erna Plácido da Silva, Hugo Gramann, Klaus |
author_sort | Wagner, Ruth Erna |
collection | PubMed |
description | Background and Objective: The reliability of low-cost mobile systems for recording Electrocardiographic (ECG) data is mostly unknown, posing questions regarding the quality of the recorded data and the validity of the extracted physiological parameters. The present study compared the BITalino toolkit with an established medical-grade ECG system (BrainAmp-ExG). Methods: Participants underwent simultaneous ECG recordings with the two instruments while watching pleasant and unpleasant pictures of the “International Affective Picture System” (IAPS). Common ECG parameters were extracted and compared between the two systems. The Intraclass Correlation Coefficients (ICCs) and the Bland–Altman Limits of Agreement (LoA) method served as criteria for measurement agreement. Results: All but one parameter showed an excellent agreement (>80%) between both devices in the ICC analysis. No criteria for Bland–Altman LoA and bias were found in the literature regarding ECG parameters. Conclusion: The results of the ICC and Bland–Altman methods demonstrate that the BITalino system can be considered as an equivalent recording device for stationary ECG recordings in psychophysiological experiments. |
format | Online Article Text |
id | pubmed-8271611 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2021 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-82716112021-07-11 Validation of a Low-Cost Electrocardiography (ECG) System for Psychophysiological Research Wagner, Ruth Erna Plácido da Silva, Hugo Gramann, Klaus Sensors (Basel) Article Background and Objective: The reliability of low-cost mobile systems for recording Electrocardiographic (ECG) data is mostly unknown, posing questions regarding the quality of the recorded data and the validity of the extracted physiological parameters. The present study compared the BITalino toolkit with an established medical-grade ECG system (BrainAmp-ExG). Methods: Participants underwent simultaneous ECG recordings with the two instruments while watching pleasant and unpleasant pictures of the “International Affective Picture System” (IAPS). Common ECG parameters were extracted and compared between the two systems. The Intraclass Correlation Coefficients (ICCs) and the Bland–Altman Limits of Agreement (LoA) method served as criteria for measurement agreement. Results: All but one parameter showed an excellent agreement (>80%) between both devices in the ICC analysis. No criteria for Bland–Altman LoA and bias were found in the literature regarding ECG parameters. Conclusion: The results of the ICC and Bland–Altman methods demonstrate that the BITalino system can be considered as an equivalent recording device for stationary ECG recordings in psychophysiological experiments. MDPI 2021-06-30 /pmc/articles/PMC8271611/ /pubmed/34209063 http://dx.doi.org/10.3390/s21134485 Text en © 2021 by the authors. https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Wagner, Ruth Erna Plácido da Silva, Hugo Gramann, Klaus Validation of a Low-Cost Electrocardiography (ECG) System for Psychophysiological Research |
title | Validation of a Low-Cost Electrocardiography (ECG) System for Psychophysiological Research |
title_full | Validation of a Low-Cost Electrocardiography (ECG) System for Psychophysiological Research |
title_fullStr | Validation of a Low-Cost Electrocardiography (ECG) System for Psychophysiological Research |
title_full_unstemmed | Validation of a Low-Cost Electrocardiography (ECG) System for Psychophysiological Research |
title_short | Validation of a Low-Cost Electrocardiography (ECG) System for Psychophysiological Research |
title_sort | validation of a low-cost electrocardiography (ecg) system for psychophysiological research |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8271611/ https://www.ncbi.nlm.nih.gov/pubmed/34209063 http://dx.doi.org/10.3390/s21134485 |
work_keys_str_mv | AT wagnerrutherna validationofalowcostelectrocardiographyecgsystemforpsychophysiologicalresearch AT placidodasilvahugo validationofalowcostelectrocardiographyecgsystemforpsychophysiologicalresearch AT gramannklaus validationofalowcostelectrocardiographyecgsystemforpsychophysiologicalresearch |