Repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating Scheimpflug camera/Placido disc tomography, and swept source anterior segment optical coherence tomography

PURPOSE: To assess the agreement and repeatability of horizontal visible iris diameter (HVID) or white-to-white (WTW) measurements between four imaging modalities; combination slit scanning elevation/Placido tomography, infrared biometry, dual rotating scheimpflug camera/Placido tomography, and swep...

Descripción completa

Detalles Bibliográficos
Autores principales: Buckenham Boyle, Alexander, Namkung, Soobin, Shew, William, Gokul, Akilesh, McGhee, Charles N. J., Ziaei, Mohammed
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8284630/
https://www.ncbi.nlm.nih.gov/pubmed/34270605
http://dx.doi.org/10.1371/journal.pone.0254832
_version_ 1783723424091209728
author Buckenham Boyle, Alexander
Namkung, Soobin
Shew, William
Gokul, Akilesh
McGhee, Charles N. J.
Ziaei, Mohammed
author_facet Buckenham Boyle, Alexander
Namkung, Soobin
Shew, William
Gokul, Akilesh
McGhee, Charles N. J.
Ziaei, Mohammed
author_sort Buckenham Boyle, Alexander
collection PubMed
description PURPOSE: To assess the agreement and repeatability of horizontal visible iris diameter (HVID) or white-to-white (WTW) measurements between four imaging modalities; combination slit scanning elevation/Placido tomography, infrared biometry, dual rotating scheimpflug camera/Placido tomography, and swept source anterior segment optical coherence tomography (AS-OCT). METHODS: A prospective study of 35 right eyes of healthy volunteers were evaluated using the Orbscan IIz, IOL Master 700, Galilei G2, and DRI Triton OCT devices. The inter-device agreement and repeatability of HVID/WTW measurements for each device were analysed. RESULTS: Mean HVID/WTW values obtained by the Orbscan IIz, IOL Master 700, Galilei G2 and DRI Triton OCT were 11.77 ± 0.40 mm, 12.40 ± 0.43 mm, 12.25 ± 0.42 mm, and 12.42 ± 0.47 mm, respectively. All pairwise comparisons revealed statistically significant differences in mean HVID/WTW measurements (p = <0.01) except for the IOL Master 700—DRI OCT Triton pair (p = 0.56). Mean differences showed that the DRI Triton OCT produced the highest HVID/WTW values, followed by the IOL Master 700, Galilei G2 and Orbscan IIz, respectively. The limits of agreement were large on all device pairs. There was high repeatability for all devices (ICC ≥ 0.980). The highest repeatability was seen in the Galilei G2 (ICC = 0.995) and lowest in the Orbscan IIz (ICC = 0.980). CONCLUSIONS: The four devices exhibit high repeatability, but should not be used interchangeably for HVID/WTW measurements in clinical practice.
format Online
Article
Text
id pubmed-8284630
institution National Center for Biotechnology Information
language English
publishDate 2021
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-82846302021-07-28 Repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating Scheimpflug camera/Placido disc tomography, and swept source anterior segment optical coherence tomography Buckenham Boyle, Alexander Namkung, Soobin Shew, William Gokul, Akilesh McGhee, Charles N. J. Ziaei, Mohammed PLoS One Research Article PURPOSE: To assess the agreement and repeatability of horizontal visible iris diameter (HVID) or white-to-white (WTW) measurements between four imaging modalities; combination slit scanning elevation/Placido tomography, infrared biometry, dual rotating scheimpflug camera/Placido tomography, and swept source anterior segment optical coherence tomography (AS-OCT). METHODS: A prospective study of 35 right eyes of healthy volunteers were evaluated using the Orbscan IIz, IOL Master 700, Galilei G2, and DRI Triton OCT devices. The inter-device agreement and repeatability of HVID/WTW measurements for each device were analysed. RESULTS: Mean HVID/WTW values obtained by the Orbscan IIz, IOL Master 700, Galilei G2 and DRI Triton OCT were 11.77 ± 0.40 mm, 12.40 ± 0.43 mm, 12.25 ± 0.42 mm, and 12.42 ± 0.47 mm, respectively. All pairwise comparisons revealed statistically significant differences in mean HVID/WTW measurements (p = <0.01) except for the IOL Master 700—DRI OCT Triton pair (p = 0.56). Mean differences showed that the DRI Triton OCT produced the highest HVID/WTW values, followed by the IOL Master 700, Galilei G2 and Orbscan IIz, respectively. The limits of agreement were large on all device pairs. There was high repeatability for all devices (ICC ≥ 0.980). The highest repeatability was seen in the Galilei G2 (ICC = 0.995) and lowest in the Orbscan IIz (ICC = 0.980). CONCLUSIONS: The four devices exhibit high repeatability, but should not be used interchangeably for HVID/WTW measurements in clinical practice. Public Library of Science 2021-07-16 /pmc/articles/PMC8284630/ /pubmed/34270605 http://dx.doi.org/10.1371/journal.pone.0254832 Text en © 2021 Buckenham Boyle et al https://creativecommons.org/licenses/by/4.0/This is an open access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Buckenham Boyle, Alexander
Namkung, Soobin
Shew, William
Gokul, Akilesh
McGhee, Charles N. J.
Ziaei, Mohammed
Repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating Scheimpflug camera/Placido disc tomography, and swept source anterior segment optical coherence tomography
title Repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating Scheimpflug camera/Placido disc tomography, and swept source anterior segment optical coherence tomography
title_full Repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating Scheimpflug camera/Placido disc tomography, and swept source anterior segment optical coherence tomography
title_fullStr Repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating Scheimpflug camera/Placido disc tomography, and swept source anterior segment optical coherence tomography
title_full_unstemmed Repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating Scheimpflug camera/Placido disc tomography, and swept source anterior segment optical coherence tomography
title_short Repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating Scheimpflug camera/Placido disc tomography, and swept source anterior segment optical coherence tomography
title_sort repeatability and agreement of white-to-white measurements between slit-scanning tomography, infrared biometry, dual rotating scheimpflug camera/placido disc tomography, and swept source anterior segment optical coherence tomography
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8284630/
https://www.ncbi.nlm.nih.gov/pubmed/34270605
http://dx.doi.org/10.1371/journal.pone.0254832
work_keys_str_mv AT buckenhamboylealexander repeatabilityandagreementofwhitetowhitemeasurementsbetweenslitscanningtomographyinfraredbiometrydualrotatingscheimpflugcameraplacidodisctomographyandsweptsourceanteriorsegmentopticalcoherencetomography
AT namkungsoobin repeatabilityandagreementofwhitetowhitemeasurementsbetweenslitscanningtomographyinfraredbiometrydualrotatingscheimpflugcameraplacidodisctomographyandsweptsourceanteriorsegmentopticalcoherencetomography
AT shewwilliam repeatabilityandagreementofwhitetowhitemeasurementsbetweenslitscanningtomographyinfraredbiometrydualrotatingscheimpflugcameraplacidodisctomographyandsweptsourceanteriorsegmentopticalcoherencetomography
AT gokulakilesh repeatabilityandagreementofwhitetowhitemeasurementsbetweenslitscanningtomographyinfraredbiometrydualrotatingscheimpflugcameraplacidodisctomographyandsweptsourceanteriorsegmentopticalcoherencetomography
AT mcgheecharlesnj repeatabilityandagreementofwhitetowhitemeasurementsbetweenslitscanningtomographyinfraredbiometrydualrotatingscheimpflugcameraplacidodisctomographyandsweptsourceanteriorsegmentopticalcoherencetomography
AT ziaeimohammed repeatabilityandagreementofwhitetowhitemeasurementsbetweenslitscanningtomographyinfraredbiometrydualrotatingscheimpflugcameraplacidodisctomographyandsweptsourceanteriorsegmentopticalcoherencetomography