Cargando…

Injection time related to intraocular pressure using a CO(2) driven preloaded injector: An experimental laboratory study

PURPOSE: Experimental study to measure the intraocular lens (IOL) injection time and injection speed at different intraocular pressure (IOP) settings when using the AutonoMe® injector. METHODS: In this experimental study, following phacoemulsification in porcine cadaver eyes, a trocar was inserted a...

Descripción completa

Detalles Bibliográficos
Autores principales: Weindler, Jan N., Naujokaitis, Tadas, Schickhardt, Sonja K., Khoramnia, Ramin, Auffarth, Gerd U.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2021
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC8289067/
https://www.ncbi.nlm.nih.gov/pubmed/34280247
http://dx.doi.org/10.1371/journal.pone.0254901
Descripción
Sumario:PURPOSE: Experimental study to measure the intraocular lens (IOL) injection time and injection speed at different intraocular pressure (IOP) settings when using the AutonoMe® injector. METHODS: In this experimental study, following phacoemulsification in porcine cadaver eyes, a trocar was inserted at pars plana with a connected infusion and IOPs of 20, 50 and 80 mmHg were generated by altering the infusion height. Twelve CO(2) gas-driven injectors were used to implant an IOL via a corneal incision of 2.2 mm. For each IOP setting, the duration of the IOL injection and the injection speed was measured by analyzing a video recording of the procedure. RESULTS: The mean ±SD injection time (seconds) was 4.47±0.50 at 20 mmHg, 4.98±0.55 at 50 mmHg and 5.47±0.20 at 80 mmHg. The mean ±SD injection speed (millimeters per seconds) was 1.36±0.15 at 20 mmHg, 1.22±0.14 at 50 mmHg and 1.10±0.04 at 80 mmHg. There was a significant (p<0.05) difference between the 20 and 80 mmHg groups in mean injection duration and injection speed. CONCLUSION: The CO(2) gas driven injector allows a safe IOL injection even at elevated IOP. Although the implantation time is slightly extended at higher IOPs, this does not seem to be clinically relevant. No IOL damage was observed at these pressure settings.